Objectives: This study compared the cyclic fatigue resistance of nickel–titanium (NiTi) files obtained in a conventional test using a simulated canal with a newly developed method that allows the application of constant fatigue load conditions. Materials and Methods: ProFile and K3 files of #25/.06, #30/.06, and #40/.04 were selected. Two types of testing devices were built to test their fatigue performance. The first (conventional) device prescribed curvature inside a simulated canal (C-test),the second new device exerted a constant load (L-test) whilst allowing any resulting curvature. Ten new instruments of each size and brand were tested with each device. The files were rotated until fracture and the number of cycles to failure (NCF) was determined. The NCF were subjected to one-way ANOVA and Duncan’s post-hoc test for each method. Spearman’s rank correlation coefficient was computed to examine any association between methods. Results: Spearman’s rank correlation coefficient (ρ = -0.905) showed a significant negative correlation between methods. Groups with significant difference after the L-test divided into 4 clusters, whilst the C-test gave just 2 clusters. From the L-test, considering the negative correlation of NCF, K3gave a significantly lower fatigue resistance than ProFile as in the C-test. K3 #30/.06showed a lower fatigue resistance than K3 #25/.06, which was not found by the C-test. Variation in fatigue test methodology resulted in different cyclic fatigue resistance rankings for various NiTi files. Conclusions: The new methodology standardized the load during fatigue testing, allowing determination fatigue behavior under constant load conditions. (Restor Dent Endod 2013;38(1):31-35)

Abstract

Objectives: This study compared the cyclic fatigue resistance of nickel–titanium (NiTi) files obtained in a conventional test using a simulated canal with a newly developed method that allows the application of constant fatigue load conditions. Materials and Methods: ProFile and K3 files of #25/.06, #30/.06, and #40/.04 were selected. Two types of testing devices were built to test their fatigue performance. The first (conventional) device prescribed curvature inside a simulated canal (C-test),the second new device exerted a constant load (L-test) whilst allowing any resulting curvature. Ten new instruments of each size and brand were tested with each device. The files were rotated until fracture and the number of cycles to failure (NCF) was determined. The NCF were subjected to one-way ANOVA and Duncan’s post-hoc test for each method. Spearman’s rank correlation coefficient was computed to examine any association between methods. Results: Spearman’s rank correlation coefficient (ρ = -0.905) showed a significant negative correlation between methods. Groups with significant difference after the L-test divided into 4 clusters, whilst the C-test gave just 2 clusters. From the L-test, considering the negative correlation of NCF, K3gave a significantly lower fatigue resistance than ProFile as in the C-test. K3 #30/.06showed a lower fatigue resistance than K3 #25/.06, which was not found by the C-test. Variation in fatigue test methodology resulted in different cyclic fatigue resistance rankings for various NiTi files. Conclusions: The new methodology standardized the load during fatigue testing, allowing determination fatigue behavior under constant load conditions. (Restor Dent Endod 2013;38(1):31-35)

참고문헌 (17)

  1. Glosson CR, "A comparison of root canal preparations using Ni–Ti hand, Ni–Ti engine-driven, and K-Flex endodontic instruments", J Endod, 21, pp.146~151 (1995)
  2. Sattapan B, "Defects in rotary nickel–titanium files after clinical use", J Endod, 26, pp.161~165 (2000)
  3. Cheung GSP, "Instrument fracture: mechanisms, removal of fragments, and clinical outcomes", Endod Topics, 16, pp.1~26 (2009)
  4. Pruett JP, "Cyclic fatigue testing of nickel–titanium endodontic instruments", J Endod, 23, pp.77~85 (1997)
  5. Kim HC, "Mechanical response of nickel-titanium instruments with different cross-sectional designs during shaping of simulated curved canals", Int Endod J, 42, pp.593~602 (2009)
  6. Kim TO, "Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysis", Int Endod J, 42, pp.14~21 (2009)
  7. Kim HC, "Cyclic fatigue and fracture characteristics of ground and twisted nickeltitanium rotary files", J Endod, 36, pp.147~152 (2010)
  8. Bahia MG, "Decrease in the fatigue resistance of nickel–titanium rotary instruments after clinical use in curved root canals", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 100, pp.249~255 (2005)
  9. Haïkel Y, "Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments", J Endod, 25, pp.434~440 (1999)
  10. Gambarini G, "Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors", J Endod, 27, pp.772~774 (2001)
  11. Cheung GS, "Fatigue testing of a NiTi rotary instrument. Part 1: strain-life relationship", Int Endod J, 40, pp.612~618 (2007)
  12. Plotino G, "A review of cyclic fatigue testing of nickel-titanium rotary instruments", J Endod, 35, pp.1469~1476 (2009)
  13. Hilt BR, "Torsional properties of stainless steel and nickel–titanium files after multiple autoclave sterilizations", J Endod, 26, pp.76~80 (2000)
  14. You SY, "Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals", J Endod, 36, pp.1991~1994 (2010)
  15. 김현철, "근관치료 기구의 기계 형태적 특성과 이에 따른 임상적 영향 고찰", Restorative Dentistry and Endodontics, 36(1), pp.1~11 (2011), 대한치과보존학회
  16. Xu X, "Comparative study of torsional and bending properties for six models of nickel–titanium root canal instruments with different cross-sections", J Endod, 32, pp.372~375 (2006)
  17. Park SY, "Dynamic torsional resistance of nickel-titanium rotary instruments", J Endod, 36, pp.1200~1204 (2010)

관련 논문AI추천

EFFECT OF CROSS-SECTIONAL AREA OF 6 NICKEL-TITANIUM ROTARY INSTRUMENTS ON THE FATIGUE FRACTURE UNDER CYCLIC FLEXURAL STRESS: A FRACTOGRAPHIC ANALYSIS이윤, 금기연, 오소람, 임상민, 황수연대한치과보존학회 · Restorative Dentistry and Endodontics2009 · 제34권 제5호Effect of adaptive motion on cyclic fatigue resistance of a nickel titanium instrument designed for retreatmentKoray Yılmaz and Gülşah Uslu, Taha Özyürek대한치과보존학회 · Restorative Dentistry and Endodontics2017 · 제42권 제1호Cyclic fatigue resistance of M-Pro and RaCe Ni-Ti rotary endodontic instruments in artificial curved canals: a comparative in vitro studyHadeer Mostafa El Feky, Khalid Mohammed Ezzat, Marwa Mahmoud Ali Bedier대한치과보존학회 · Restorative Dentistry and Endodontics2019 · 제44권 제4호Comparison of the cyclic fatigue resistance of VDW.ROTATE, TruNatomy, 2Shape, and HyFlex CM nickel-titanium rotary files at body temperatureGündoğar Mustafa, Özyürek Taha, Plotino Gianluca, Uslu Gülşah대한치과보존학회 · Restorative Dentistry and Endodontics2020 · 제45권 제3호Cyclic fatigue resistance, torsional resistance, and metallurgical characteristics of M3 Rotary and M3 Pro Gold NiTi files장석우, Elena Bruno, Ernesto Rapisarda, Eugenio Pedullà, Fabio Lo Savio, Gabriele Miccoli, Gianluca Gambarini, Giusy Rita Maria La Rosa, Guido La Rosa, Luca Testarelli, Silvia Rapisarda대한치과보존학회 · Restorative Dentistry and Endodontics2018 · 제43권 제2호Effect of surface treatment on the mechanical properties of nickel-titanium files with a similar cross-section곽상원, 구혜진, 김현철, 이주영대한치과보존학회 · Restorative Dentistry and Endodontics2017 · 제42권 제3호