토모테라피에서 통계적공정관리를 이용한 EBT 필름 기반의 선량품질보증의 치료계획 가이드라인
Treatment Planning Guideline of EBT Film-based Delivery Quality Assurance Using Statistical Process Control in Helical Tomotherapy
2
조회
0
다운로드
4
인용
0
참고문헌
키워드
초록
The purpose of this study was to analyze the results from statistical process control (SPC) to recommend upper and lower control limits for planning parameters based on delivery quality assurance (DQA) results and establish our institutional guidelines regarding planning parameters for helical tomotherapy (HT). A total of 53 brain, 41 head and neck (H & N), and 51 pelvis cases who had passing or failing DQA measurements were selected. The absolute point dose difference (DD) and the global gamma passing rate (GPR) for all patients were analyzed. Control charts were used to evaluate upper and lower control limits (UCL and LCL) for all assessed treatment planning parameters. Treatment planning parameters were analyzed to provide its range for DQA pass cases. We confirmed that the probability of DQA failure was higher when the proportion of leaf open time (LOT) below 100 ms was greater than 30%. LOT and gantry period (GP) were significant predictor for DQA failure using the SPC method. We investigated the availability of the SPC statistic method to establish the local planning guideline based on DQA results for HT system. The guideline of each planning parameter in HT may assist in the prediction of DQA failure using the SPC statistic method in the future.
Abstract
The purpose of this study was to analyze the results from statistical process control (SPC) to recommend upper and lower control limits for planning parameters based on delivery quality assurance (DQA) results and establish our institutional guidelines regarding planning parameters for helical tomotherapy (HT). A total of 53 brain, 41 head and neck (H & N), and 51 pelvis cases who had passing or failing DQA measurements were selected. The absolute point dose difference (DD) and the global gamma passing rate (GPR) for all patients were analyzed. Control charts were used to evaluate upper and lower control limits (UCL and LCL) for all assessed treatment planning parameters. Treatment planning parameters were analyzed to provide its range for DQA pass cases. We confirmed that the probability of DQA failure was higher when the proportion of leaf open time (LOT) below 100 ms was greater than 30%. LOT and gantry period (GP) were significant predictor for DQA failure using the SPC method. We investigated the availability of the SPC statistic method to establish the local planning guideline based on DQA results for HT system. The guideline of each planning parameter in HT may assist in the prediction of DQA failure using the SPC statistic method in the future.
참고문헌 (29)
- Kyung Hwan Chang, "Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations", 의학물리, 27(3), pp.111~116 (2016), 한국의학물리학회
- 함일식, "Deep Inspiration Breath Holding을 적용한 유방암 세기변조방사선치료 시 위치잡이오차 분석을 통한 선량 평가", 방사선기술과학, 42(2), pp.137~143 (2019), 대한방사선과학회
- 김대건, "토모테라피에서 쌀 볼루스 팬텀의 유용성 평가: 팬텀연구", 방사선기술과학, 44(6), pp.663~669 (2021), 대한방사선과학회
- Chang KH, "Development of a 3D optical scanner for evaluating patient-specific dose distributions", Physica Medica, 31(5), pp.553~559 (2015)
- Thiyagarajan R, "Analyzing the performance of ArcCHECK diode array detector for VMAT plan", Reports of Practical Oncology &Radiotherapy, 21(1), pp.50~56 (2016)
- Dong P, "4π noncoplanar stereotactic body radiation therapy for centrally located or larger lung tumors", Int J Radiat Oncol Biol Phys, 86(3), pp.407~413 (2013)
- Guckenberger M, "Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy(IGRT)", Radiother Oncol, 84(1), pp.56~63 (2007)
- Montgomery DC, "Statistical quality control" (2009), Wiley
- SEMATECH e-Handbook of Statistical Methods, "NIST", pp.2007, Prins J
- Breen SL, "Statistical process control for IMRT dosimetric verification", Med Phys, 35(10), pp.4417~4425 (2008)
- Chen M, "Theoretical analysis of the thread effect in helical TomoTherapy", Medical Physics, 38(11), pp.5945~5960 (2011)
- Shimizu H, "Analysis of modulation factor to shorten the delivery time in helical tomotherapy", J Appl Clin Med Phys, 18(3), pp.83~87 (2017)
- Skorska M, "Optimization of treatment planning parameters used in tomotherapy for prostate cancer patients", Phys Med, 29(3), pp.273~285 (2013)
- Thomas SJ, "Implications of leaf fluence opening factors on transfer of plans between matched helical tomotherapy machines", Biomedical Physics & Engineering Express, 4(1), pp.017001~ (2017)
- Efficient planning for efficient treatments on tomotherapy, "Vaden R", ASTRO, ~ (2016)
- Boyd R, "Determining efficient helical IMRT modulation factor from the MLC leaf-open time distribution on precision treatment planning system", J Appl Clin Med Phys, 20(5), pp.64~74 (2019)
- Binny D, "Statistical process control and verifying positional accuracy of a cobra motion couch using step-wedge quality assurance tool", J Appl Clin Med Phys, 18(5), pp.70~79 (2017)
- Binny D, "A multi-institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control", J Appl Clin Med Phys, 20(3), pp.71~80 (2019)
- Binny D, "Tomotherapy treatment site specific planning using statistical process control", Phys Med, 53, pp.32~39 (2018)
- Mezzenga E, "Preliminary Retrospective Analysis of Daily Tomotherapy Output Constancy Checks Using Statistical Process Control", PLoS One, 11(2), pp.e0147936~ (2016)
- Meyers SM, "Evaluation of Elekta Agility multi-leaf collimator performance using statistical process control tools", J Appl Clin Med Phys, 20(7), pp.100~108 (2019)
- Sanghangthum T, "Statistical process control analysis for patient-specific IMRT and VMAT QA", Journal of Radiation Research, 54(3), pp.546~552 (2013)
- Kyung Hwan Chang, "Dosimetric Comparison of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy", Journal of the Korean Physical Society, 76(3), pp.257~263 (2020), 한국물리학회[DOI]
- Chang KH, "Statistical Analysis of Treatment Planning Parameters for Prediction of Delivery Quality Assurance Failure for Helical Tomotherapy", Technology in Cancer Research & Treatment, 19, pp.1~9 (2020)
- 고윤석, "CART 알고리즘 기반의 의사결정트리 기법을 이용한 규칙기반 전문가 시스템 구축 방법론", 한국전자통신학회 논문지, 6(6), pp.849~854 (2011), 한국전자통신학회
- "TomoTherapy® Treatment System StatRT™ Guide", TomoTherapy Incorporated, ~ (2010)
- Westerly DC, "Treatment planning to improve delivery accuracy and patient throughput in helical tomotherapy", Int J Radiat Oncol Biol Phys, 74(4), pp.1290~1297 (2009)
- Martisikova M, "Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams", Phys Med Biol, 53(24), pp.7013~7027 (2008)
- Zhao YL, "Monte Carlo calculation of helical tomotherapy dose delivery", Med Phys, 35(8), pp.3491~3500 (2008)