The study investigates the necessity of 3 dimensional dose distribution evaluation instead of point dose and 2 dimensional dose distribution evaluation. Treatment plans were generated on the RANDO phantom to measure the precise dose distribution of the treatment site 0.5, 1, 1.5, 2, 2.5, 3 cm with the prescribed dose; 1,200 cGy, 5 fractions. Gamma analysis (3%/3 mm, 2%/2 mm) of dose distribution was evaluated with gafchromic EBT2 film and ArcCHECK phantom. The average error of absolute dose was measured at 0.76±0.59% and 1.37±0.76% in cheese phantom and ArcCHECK phantom respectively. The average passing ratio for 3%/3 mm were 97.72±0.02% and 99.26±0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. The average passing ratio for 2%/2 mm were 94.21±0.02% and 93.02±0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. There was a more accurate dose distribution of 3D volume phantom than cheese phantom in patients DQA using tomotherapy. Therefor it should be evaluated simultaneously 3 dimensional dose evaluation on target and peripheral area in rotational radiotherapy such as tomotherapy.

Abstract

The study investigates the necessity of 3 dimensional dose distribution evaluation instead of point dose and 2 dimensional dose distribution evaluation. Treatment plans were generated on the RANDO phantom to measure the precise dose distribution of the treatment site 0.5, 1, 1.5, 2, 2.5, 3 cm with the prescribed dose; 1,200 cGy, 5 fractions. Gamma analysis (3%/3 mm, 2%/2 mm) of dose distribution was evaluated with gafchromic EBT2 film and ArcCHECK phantom. The average error of absolute dose was measured at 0.76±0.59% and 1.37±0.76% in cheese phantom and ArcCHECK phantom respectively. The average passing ratio for 3%/3 mm were 97.72±0.02% and 99.26±0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. The average passing ratio for 2%/2 mm were 94.21±0.02% and 93.02±0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. There was a more accurate dose distribution of 3D volume phantom than cheese phantom in patients DQA using tomotherapy. Therefor it should be evaluated simultaneously 3 dimensional dose evaluation on target and peripheral area in rotational radiotherapy such as tomotherapy.

참고문헌 (15)

  1. Mack A., "Precision dosimetry for narrow photon beams used in radiosurgery Determination of Gamma Knife output factors", Medical Physics, 29(9), pp.2080~2089 (2002)
  2. 최태진, "TM31010 계열의 공동이온전리함과 TG-51을 이용한 선형가속기광자선의 5년간 출력선량 평가", 의학물리, 22(2), pp.92~98 (2011), 한국의학물리학회
  3. Welsh J.S., "Clinical implementation of adaptive helical tomotherapy: a unique approach to image-guided intensity modulated radiotherapy", Technology in cancer research & treatment, 5(5), pp.465~480 (2006)
  4. Cadman P., "Validation of physics improvements for IMRT with a commercial treatment-planning system", Journal of Applied Clinical Medical Physics, 6(2), pp.74~86 (2005)
  5. Mu G., "Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer", Physics in medicine & biology, 53(1), pp.77~88 (2008)
  6. Bayouth J. E., "MLC quality assurance techniques for IMRT applications", Medical Physics, 30(5), pp.743~750 (2003)
  7. LoSasso T., "Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode", Medical Physics, 28(11), pp.2209~2219 (2001)
  8. Ju S. G., "Film dosimetry for intensity modulated radiation therapy: dosimetric evaluation", Medical Physics, 29(3), pp.351~355 (2002)
  9. Herzen J., "Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine", Physics in medicine and biology, 52(4), pp.1197~1208 (2007)
  10. Daniel L, "Evaluation of a 2D diode array for IMRT assurance", Radiation &Oncology, 70(1), pp.199~206 (2004)
  11. 오영택, "중합체 겔과 자기공명영상을 이용한 3차원 선량분포 측정", Radiation oncology journal, 20(3), pp.10~273 (2002), 대한방사선종양학회
  12. Bedford J.L., "Warrington A.P.: Evaluation of the Delta4 phantom for IMRT and VMAT verification", Physics in medicine and biology, 54(9), pp.167~176 (2009)
  13. Van Esch A., "On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom", Medical Physics, 34(10), pp.3825~3837 (2007)
  14. H. Fakir, "Development of a novel ArcCHECK insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities", Medical Physics., 39(7), pp.4203~4208 (2012)
  15. Welsh J.S., "Clinical implementation of adaptive helical tomotherapy:a unique approach to image-guided intensity modulated radiotherapy", in cancer research & treatment, 5(5), pp.465~480 (2006)