본문 바로가기
KERD
논문KCI 등재

Feasibility on Statistical Process Control Analysis of Delivery Quality Assurance in Helical Tomotherapy

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The purpose of this study was to retrospectively investigate the upper and lower control limits of treatment planning parameters using EBT film based delivery quality assurance (DQA) results and to analyze the results of statistical process control (SPC) in helical tomotherapy (HT). A total of 152 patients who passed or failed DQA results were retrospectively included in this study. Prostate (n = 66), rectal (n = 51), and large-field cancer patients, including lymph nodes (n = 35), were randomly selected. The absolute point dose difference (DD) and global gamma passing rate (GPR) were analyzed for all patients. Control charts were used to evaluate the upper and lower control limits (UCL and LCL) for all the assessed treatment planning parameters. Treatment planning parameters such as gantry period, leaf open time (LOT), pitch, field width, actual and planning modulation factor, treatment time, couch speed, and couch travel were analyzed to provide the optimal range using the DQA results. The classification and regression tree (CART) was used to predict the relative importance of variables in the DQA results from various treatment planning parameters. We confirmed that the proportion of patients with an LOT below 100 ms in the failure group was relatively higher than that in the passing group. SPC can detect QA failure prior to over dosimetric QA tolerance levels. The acceptable tolerance range of each planning parameter may assist in the prediction of DQA failures using the SPC tool in the future.

Abstract

The purpose of this study was to retrospectively investigate the upper and lower control limits of treatment planning parameters using EBT film based delivery quality assurance (DQA) results and to analyze the results of statistical process control (SPC) in helical tomotherapy (HT). A total of 152 patients who passed or failed DQA results were retrospectively included in this study. Prostate (n = 66), rectal (n = 51), and large-field cancer patients, including lymph nodes (n = 35), were randomly selected. The absolute point dose difference (DD) and global gamma passing rate (GPR) were analyzed for all patients. Control charts were used to evaluate the upper and lower control limits (UCL and LCL) for all the assessed treatment planning parameters. Treatment planning parameters such as gantry period, leaf open time (LOT), pitch, field width, actual and planning modulation factor, treatment time, couch speed, and couch travel were analyzed to provide the optimal range using the DQA results. The classification and regression tree (CART) was used to predict the relative importance of variables in the DQA results from various treatment planning parameters. We confirmed that the proportion of patients with an LOT below 100 ms in the failure group was relatively higher than that in the passing group. SPC can detect QA failure prior to over dosimetric QA tolerance levels. The acceptable tolerance range of each planning parameter may assist in the prediction of DQA failures using the SPC tool in the future.

참고문헌 (24)

  1. Miften M, "Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218", Med Phys, 45(4), pp.e53~e83 (2018)
  2. Intensity Modulated Radiation Therapy Collaborative Working Group, "Intensity-modulated radiotherapy : Current status and issues of interest", Int J Radiat Oncol Biol Phys, 51(4), pp.880914~ (2001)
  3. 조병철, "Intensity-modulated radiation therapy: a review with a physics perspective", Radiation oncology journal, 36(1), pp.1~10 (2018), 대한방사선종양학회
  4. Chang KH, "Development of a 3D optical scanner for evaluating patient-specific dose distributions", Phys Med, 31(5), pp.553~559 (2015)
  5. Thiyagarajan R, "Analyzing the performance of ArcCHECK diode array detector for VMAT plan", Rep Pract Oncol Radiother, 21(1), pp.50~56 (2016)
  6. 장경환, "토모테라피에서 통계적공정관리를 이용한 EBT 필름 기반의 선량품질보증의 치료계획 가이드라인", 방사선기술과학, 45(5), pp.439~448 (2022), 대한방사선과학회
  7. TomoTherapy Incorporated, "TomoTherapy® Treatment System StatRT™ Guide", ~ (2010)
  8. 정광호, "방사선치료에서의 품질보증을 위한 통계적공정관리의 활용", 의학물리, 26(2), pp.059~071 (2015), 한국의학물리학회[DOI]
  9. Wheeler DJ, "Understanding statistical process control" (1992), SPC Press
  10. Montgomery DC, "Statistical quality control" (2009), Wiley
  11. SEMATECH e-Handbook of Statistical Methods, "NIST", pp.2007, Prins J
  12. Breen SL, "Statistical process control for IMRT dosimetric verification", Med Phys, 35(10), pp.4417~4425 (2008)
  13. Binny D, "A multi-institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control", J Appl Clin Med Phys, 20(3), pp.71~80 (2019)
  14. Binny D, "Tomotherapy treatment site specific planning using statistical process control", Phys Med, 53, pp.32~39 (2018)
  15. Mezzenga E, "Preliminary Retrospective Analysis of Daily Tomotherapy Output Constancy Checks Using Statistical Process Control", PLoS One, 11(2), pp.e0147936~ (2016)
  16. Meyers SM, "Evaluation of Elekta Agility multi-leaf collimator performance using statistical process control tools", J Appl Clin Med Phys, 20(7), pp.100~108 (2019)
  17. Sanghangthum T, "Statistical process control analysis for patient-specific IMRT and VMAT QA", J Radiat Res, 54(3), pp.546~552 (2013)
  18. Chang KH, "Statistical Analysis of Treatment Planning Parameters for Prediction of Delivery Quality Assurance Failure for Helical Tomotherapy", Technol Cancer Res Treat, 19, pp.1~9 (2020)
  19. Kyung Hwan Chang, "Dosimetric Comparison of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy", Journal of the Korean Physical Society, 76(3), pp.257~263 (2020), 한국물리학회[DOI]
  20. Razali NM, "Power comparison of Shapiro-Wilk, Kolmogorov-Smirnov, Lilliefors and Anderson-Darling tests", J Stat Mod Analyt, 2(1), pp.21~33 (2011)
  21. Thomas SJ, "Implications of leaf fluence opening factors on transfer of plans between matched helical tomotherapy machines", Biomed Phys Eng Express, 4(1), pp.017001~ (2017)
  22. Efficient planning for efficient treatments on tomotherapy, "Vaden R", ASTRO, ~ (2016)
  23. Westerly DC, "Treatment planning to improve delivery accuracy and patient throughput in helical tomotherapy", Int J Radiat Oncol Biol Phys, 74(4), pp.1290~1297 (2009)
  24. Martisikova M, "Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams", Phys Med Biol, 53(24), pp.7013~7027 (2008)

관련 논문AI추천