The objective of this study was to evaluate the accuracy and impact of leaf open time (LOT) and pitch using various machine learning models on EBT film-based delivery quality assurance (DQA) performed on 211 patients of helical tomotherapy (HT). We randomly selected passed (n=191) and failed (n=20) DQA measurements to evaluate the accuracy of the k-nearest neighbor (KNN), support vector machine (SVM), naive Bayes (NB) and logistic regression (LR) models using scale-dependent metrics such as the coefficient of determination (R2), mean squared error (MSE), and root MSE (RMSE). We evaluated the performance of the four prediction models in terms of the accuracy, precision, sensitivity, and F1-score using a confusion matrix, finding the NB and LR models to achieve optimal results. The results of this study are expected to reduce the workload of medical physicists and dosimetrists by predicting DQA results according to LOT and pitch in advance.

Abstract

The objective of this study was to evaluate the accuracy and impact of leaf open time (LOT) and pitch using various machine learning models on EBT film-based delivery quality assurance (DQA) performed on 211 patients of helical tomotherapy (HT). We randomly selected passed (n=191) and failed (n=20) DQA measurements to evaluate the accuracy of the k-nearest neighbor (KNN), support vector machine (SVM), naive Bayes (NB) and logistic regression (LR) models using scale-dependent metrics such as the coefficient of determination (R2), mean squared error (MSE), and root MSE (RMSE). We evaluated the performance of the four prediction models in terms of the accuracy, precision, sensitivity, and F1-score using a confusion matrix, finding the NB and LR models to achieve optimal results. The results of this study are expected to reduce the workload of medical physicists and dosimetrists by predicting DQA results according to LOT and pitch in advance.

참고문헌 (21)

  1. Chang KH, "Clinical implications of High Definition Multileaf Collimator(HDMLC)Dosimetric Leaf Gap(DLG)Variations", Prog Med Phys, 27(3), pp.111~116 (2016)[DOI]
  2. Cho B, "Intensity-modulated radiation therapy : A review with a physics perspective", Radiat Oncol J, 36(1), pp.1~10 (2018)[DOI]
  3. Thiyagarajan R, "Analyzing the performance of ArcCHECK diode array detector for VMAT plan", Reports of Practical Oncology & Radiotherapy, 21(1), pp.50~56 (2016)[DOI]
  4. Chang KH, "Treatment planning guideline of EBT-film based delivery quality assurance using statistical process control in helical tomotherapy", Journal of Radiological Science and Technology, 45(5), pp.439~448 (2022)[DOI]
  5. Chang KH, "A comparison of patient-specific delivery quality assurance (DQA) devices in radiation therapy", Journal of Radiological Science and Technology, 46(3), pp.231~238 (2023)
  6. Guckenberger M, "Precision required for dose-escalated treatment of spinal metastases and implications for image-guided radiation therapy(IGRT)", Radiother Oncol, 84(1), pp.56~63 (2007)[DOI]
  7. Montgomery DC, "Statistical quality control" (2009), Wiley
  8. Chung E, "Clinical implementation of Dosimetry Check™ for TomoTherapy®delivery quality assurance", J Appl Clin Med Phys, 19(6), pp.193~199 (2018)[DOI]
  9. McCowan PM, "A model-based 3D patient-specific pre-treatment QA method for VMAT using the EPID", Phys Med Biol, 62(4), pp.1600~1612 (2017)[DOI]
  10. Chang KH, "Dosimetric comparison of four commercial patient-specific quality assurance devices for helical tomotherapy", J Korean Phys Soc, 76, pp.257~263 (2020)[DOI]
  11. Chang KH, "Statistical analysis of treatment planning parameters for prediction of delivery quality assurance failure for helical tomotherapy", Technol Cancer Res Treat, 19, pp.1533033820979692~ (2020)[DOI]
  12. Siddalingappa R, "K-nearest-neighbor algorithm to predict the survival time and classification of various stages of oral cancer : A machine learning approach", F1000Res, 16(11), pp.70~ (2023)[DOI]
  13. Kubat M, "An introduction to machine learning" (2015), Springer Publishing Company
  14. Cilla S, "Prediction and classification of VMAT dosimetric accuracy using plan complexity and log-files analysis", Phys Med, 103, pp.76~88 (2022)[DOI]
  15. Wall PDH, "Application and comparison of machine learning models for predicting quality assurance outcomes in radiation therapy treatment planning", Informatics in Medicine Unlocked, 18, pp.100292~ (2020)[DOI]
  16. Kononenko I, "Inductive and bayesian learning in medical diagnosis", Appl Artif Intell, 7(4), pp.317~337 (1993)[DOI]
  17. Jierula A, "Study on accuracy metrics for evaluating the predictions of damage locations in deep piles using artificial neural networks with acoustic emission data", Applied Sciences, 11(5), pp.2314~ (2021)[DOI]
  18. Thomas SJ, "Implications of leaf fluence opening factors on transfer of plans between matched helical tomotherapy machines", Biomedical Physics & Engineering Express, 4(1), pp.017001~ (2017)[DOI]
  19. Cavinato S, "Prediction models as decision-support tools for virtual patient-specific quality assurance of helical tomotherapy plans", Phys Imaging Radiat Oncol, 26, pp.100435~ (2023)[DOI]
  20. Zhu H, "Patient-specific quality assurance prediction models based on machine learning for novel dual-layered MLC linac", Med Phys, 50(2), pp.1205~1214 (2023)[DOI]
  21. Kusunoki T, "Evaluation of prediction and classification performances in different machine learning models for patient-specific quality assurance of head-and-neck VMAT plans", Med Phys, 49(1), pp.727~741 (2022)[DOI]

관련 논문AI추천

Feasibility on Statistical Process Control Analysis of Delivery Quality Assurance in Helical Tomotherapy장경환대한방사선과학회 · Journal of Radiological Science and Technology2022 · 제45권 제6호토모테라피에서 통계적공정관리를 이용한 EBT 필름 기반의 선량품질보증의 치료계획 가이드라인장경환대한방사선과학회 · Journal of Radiological Science and Technology2022 · 제45권 제5호Prediction and classification of gamma passing rate for patient-specific quality assurance using machine learning models based on radiomics features and beam parametersDong Zhiwei, Kambondo Alvin, Li Qian, Liu Fangyu, Qie Shuai, Qi Xiaolu, Shi Hongyun, Wang Jie, Zhang Qianru한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제10호소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가강동진, 권경태, 김연래, 민정환, 신영주, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제1호Evaluation of Beam-Matching Accuracy for 8 MV Photon Beam between the Same Model Linear Accelerator강성희, 김연래, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2020 · 제43권 제2호용적변조회전 방사선치료 시 Set-up 오차에 따른선량 전달 정확성의 평가김상현, 나권우, 박성아, 이주훈, 하광은(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제1호

이 간행물의 다른 논문 (5건)