본문 바로가기
KERD
논문KCI 등재

용적변조회전 방사선치료 시 Set-up 오차에 따른선량 전달 정확성의 평가

Assessment of Dose Delivery Accuracy with Set-up Error for VMAT

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In order to investigate the effect of set-up error on the dose delivery, this study evaluates the accuracy of dose delivery due to set-up error by applying a virtual brain tumor treatment plan using volumetric modulated rotational radiation therapy (VMAT). After obtaining a crosssectional image with a CT device using Head Phantom, a treatment planning program was used to establish a VMAT treatment plan. Based on the treatment plan established, DQA was performed to assess the difference in dose delivery accuracy with set-up error. When the distance was changed to 7 mm with a distance of 1 mm on the basis of the measured value of 0 in the direction of Lat. and Vrt, a comparison was performed by deriving the average after 5 measurements each using a three dimensional array detector. We applied a virtual brain tumor treatment plan using VMAT. In the absence of set-up error, the measured values were 96.8%, 96.4%, 96.2% in Lat. ±1 mm, 96.2%, 95.9% in Lat. ±2 mm, 95.4%, 95.2% in Lat. ±3 mm, 92.0%, 90.1% in Lat. ±4 mm, 86.9%, 83.7% in Lat. ±5 mm, 85.6%, 80.3% in Lat. ±6 mm, 75.2%, 79.5% in Lat .±7 mm, respectively. 96.1%, 96.1% in Vrt. ±1 mm, 95.9%, 95.7% in Vrt. ±2 mm, 95.1%, 95.1% in Vrt. ±3 mm, 91.4%, 91.5% in Vrt. ±4 mm, 88.6%, 88.3% in Vrt. ±5 mm, 83.1%, 85.6% in Vrt. ±6 mm, 79.1%, 77.2% in Vrt. ±7 mm respectively. As the lateral and vertical error increases, the accuracy of the dose delivery decreases. In order to improve the accuracy of the dose delivery during VMAT treatment, we should try to reduce the error through accurate posture and appropriate image induction.

Abstract

In order to investigate the effect of set-up error on the dose delivery, this study evaluates the accuracy of dose delivery due to set-up error by applying a virtual brain tumor treatment plan using volumetric modulated rotational radiation therapy (VMAT). After obtaining a crosssectional image with a CT device using Head Phantom, a treatment planning program was used to establish a VMAT treatment plan. Based on the treatment plan established, DQA was performed to assess the difference in dose delivery accuracy with set-up error. When the distance was changed to 7 mm with a distance of 1 mm on the basis of the measured value of 0 in the direction of Lat. and Vrt, a comparison was performed by deriving the average after 5 measurements each using a three dimensional array detector. We applied a virtual brain tumor treatment plan using VMAT. In the absence of set-up error, the measured values were 96.8%, 96.4%, 96.2% in Lat. ±1 mm, 96.2%, 95.9% in Lat. ±2 mm, 95.4%, 95.2% in Lat. ±3 mm, 92.0%, 90.1% in Lat. ±4 mm, 86.9%, 83.7% in Lat. ±5 mm, 85.6%, 80.3% in Lat. ±6 mm, 75.2%, 79.5% in Lat .±7 mm, respectively. 96.1%, 96.1% in Vrt. ±1 mm, 95.9%, 95.7% in Vrt. ±2 mm, 95.1%, 95.1% in Vrt. ±3 mm, 91.4%, 91.5% in Vrt. ±4 mm, 88.6%, 88.3% in Vrt. ±5 mm, 83.1%, 85.6% in Vrt. ±6 mm, 79.1%, 77.2% in Vrt. ±7 mm respectively. As the lateral and vertical error increases, the accuracy of the dose delivery decreases. In order to improve the accuracy of the dose delivery during VMAT treatment, we should try to reduce the error through accurate posture and appropriate image induction.

참고문헌 (13)

  1. 박진호, "콘빔CT (CBCT)를 이용한 온라인 영상유도방사선치료(On-line Image Guided Radiation Therapy)", Radiation oncology journal, 24(4), pp.294~299 (2006), 대한방사선종양학회
  2. Bruno S, "Clinical application of image-guided radiotherapy, IGRT (on the Varian OBI platform)", Cancer Radioth, 10(5), pp.252~257 (2006)[DOI]
  3. Francisco CH, "A Probability Approach to the Study on Uncertainty Effects on Gamma Index Evaluations in Radiation Therapy", Int. J. Comput. Math., 2011, pp.1~10 (2011)
  4. Guckenberger M, "Dosimetric consequences of translation and rotational errors in frame-less image-guided radiosurgery", Radiat. Oncol., 7(1), pp.1~8 (2012)[DOI]
  5. Holtzer N, "Roll and Pitch Correction in Headand-Neck Patients by Rapid Adaptation of Volumetric Modulated Arc Treatment Plans", Int. J. Radiat. Oncol. Biol. Phys., 84(3), pp.S28~S29 (2012)
  6. Hussein M, "A comparison of the gamma index analysis in various commercial IMRT", VMAT QA systems, 2013, pp.109(3), 370~376
  7. 김동현, "뇌ㆍ두경부 방사선치료 시 전자조사문영상장치를 이용한 세트업 오차 확인에서 제2경추 치상돌기 위치의 임상적 의의", Radiation oncology journal, 29(2), pp.107~114 (2011), 대한방사선종양학회
  8. Kim JD, "Analyses of the Setup Errors using on Board Imager (OBI)", J. Korean Soc. Radiat Ther., 19(1), pp.1~5 (2007)
  9. Kim SH, "Evaluate the Implementation of Volumetric Modulated Arc Therapy QA in the Radiation Therapy Treatment according to Various Factors by Using the Portal Dosimetry", J. Korean Soc. Radia Ther., 27(2), pp.167~174 (2016)
  10. Lee DH, "The Accuracy Evaluation according to Dose Delivery Interruption and Restart for Volumetric Modulated Arc Therapy", J. Korean Soc. Radiat Ther., 25(1), pp.77~85 (2013)
  11. Li G, "Evaluation of the ArcCHECK QA system for IMRT and VMAT verification", Phys. Medica, 29(3), pp.295~303 (2013)[DOI]
  12. Stasi M, "Pretreatment patient-specific IMRT quality assurance : a correlation study between gamma index and patient clinical dose volume histogram", Med. Phys., 39(12), pp.7626~7634 (2012)[DOI]
  13. Stock M, "Interpretation and evaluation of the <TEX>${\gamma}$<", TEX> index and the <TEX>${\gamma}$<, Phys. Med. Biol., pp.2005, 50(3)[DOI]

관련 논문AI추천

소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가강동진, 권경태, 김연래, 민정환, 신영주, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제1호AAPM TG-119 보고서를 통한 세기조절방사선치료(IMRT)와부피적세기조절회전치료(VMAT)의 치료 전 환자별 정도관리박승, 강동진, 김연래, 김종하, 손승창, 신영주, 이근섭, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2012 · 제35권 제3호동적조형회전조사 시 표적종양의 위치변위와 조사반경의 변화에 따른 선량전달 오류분석강민영, 김유현, 이보람, 이정우대한방사선과학회 · Journal of Radiological Science and Technology2011 · 제34권 제1호두경부 IMRT 및 VMAT 시 체적 감소가 전산화치료계획에 미치는 영향김가중, 백금문, 엄기천대한방사선과학회 · Journal of Radiological Science and Technology2023 · 제46권 제3호VMAT 기법을 활용한 HA-WBRT에서 빔 배열 변화에 따른 선량학적 인자 비교: 팬텀 연구김동현, 박성국대한방사선과학회 · Journal of Radiological Science and Technology2024 · 제47권 제4호Deep Inspiration Breath Holding을 적용한 유방암 세기변조방사선치료 시 위치잡이오차 분석을 통한 선량 평가정강교, 조평곤, 함일식대한방사선과학회 · Journal of Radiological Science and Technology2019 · 제42권 제2호

이 간행물의 다른 논문 (13건)