초록

The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric M odulated A rc T herapy t reatment p lan. E valuate treatment plan t hrough t he D VH a nalysis, P TV m ean dose (Dmean) and PTV 95% dose(D95) compare the MFX based on original treatment plan, Average error rate was 0.9±0.6%, 1.0±0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9±0.7%, 1.1±0.7% (high dose region), 1.1±0.9%, 1.2±0.7% (low dose region). The average gamma though of MFX and Delta4PT is 98.7±1.2%, 98.4±1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.

Abstract

The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric M odulated A rc T herapy t reatment p lan. E valuate treatment plan t hrough t he D VH a nalysis, P TV m ean dose (Dmean) and PTV 95% dose(D95) compare the MFX based on original treatment plan, Average error rate was 0.9±0.6%, 1.0±0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9±0.7%, 1.1±0.7% (high dose region), 1.1±0.9%, 1.2±0.7% (low dose region). The average gamma though of MFX and Delta4PT is 98.7±1.2%, 98.4±1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.

참고문헌 (18)

  1. Lee L, "RETROSPECTIVE IMRT DOSE RECONSTRUCTION BASED ON CONE-BEAM CT AND MLC LOG-FILE", Int J Radiat Oncol Biol Phys, 70(2), pp.634~644 (2008)
  2. 한영이, "세기변조 방사선치료의 환자 치료 전 선량보증에 대한 고찰", 의학물리, 24(4), pp.213~219 (2013), 한국의학물리학회[DOI]
  3. 박소연, "세기조절방사선치료의 정도관리를 위한 모니터유닛 공간분포 재구성의 효용성 평가", 방사선방어학회지, 36(1), pp.28~34 (2011), 테스트학회
  4. Maria N, "Calibration and quality assurance for rounded leaf-end MLC systems", Med Phys, 28(11), pp.2227~2223 (2001)[DOI]
  5. Nelms B.E, "Per-beam, planar IMRT QA passing rates do not clinically relevant patient dose error", Med Phys, 38(2), pp.1037~1044 (2011)[DOI]
  6. Qian J, "Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files", Phys Med Biol, 55(13), pp.3597~3610 (2010)[DOI]
  7. Markus A, "Guidelines for the verification of IMRT" (2008)
  8. "Mobius 3D USER GUIDE: MOBIUS MEDICAL SYSTEM" (2015)
  9. IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119, "Gary A", ~ (2009)
  10. Nelson C. L, "Commissioning results of an automated treatment planning verification system", JOURNAL OF APPLIED CLINICAL PHYSICS, 15(5), pp.57~65 (2014)
  11. Klein E, "Task Group 142 report: Quality assurance of medical accelerators", Med. Phys, 36(9), pp.4197~4212 (2009)[DOI]
  12. 박달, "단층치료기를 이용한 방사선 치료의 환자별 정도관리 평가를 위한 감마인덱스의 정량화 방법", 의학물리, 21(4), pp.340~347 (2010), 한국의학물리학회
  13. Olasolo-Alonso J, "Evaluation of MLC performance in VMAT and dynamic IMRT by log file analysis", Phys Med, 33, pp.87~94 (2017)[DOI]
  14. Prabaker S, "A study on dosimetric properties of electronic portal imaging device and its use as a quality assurance tool in volumetric modulated arc therapy", Rep Prack Oncol Radiother, 16(6), pp.248~255 (2011)[DOI]
  15. Krzysztof S, "EPID in vivo dosimetry in RapidArc technique", Rep Prack Oncol Radiother, 15(1), pp.8~14 (2010)[DOI]
  16. Dewayne L, "Investigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methods", J Appl Clin Med Phys, 18(4), pp.172~179 (2017)[DOI]
  17. Dharanipahy R, "Catching errors with patient-specific pretreatment machine log file analysis", Practical Radiation Oncology, 3(2), pp.80~90 (2013)[DOI]
  18. "Mobius3D Software based IMRT QA" (2015)