Correlation between Treatment Planning Parameters and PSQA Results for Two Quality Assurance Devices

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Many institutions still perform patient-specific quality assurance (PSQA) using 2D array devices. However, no re- ports in the literature thus far have compared the correlations between PSQA results and treatment planning parameters to evaluate the impact of QA on IMRT. The purpose of this study is to compare the results of PSQA between MapCHECK and Octavius for radiation therapy. A total of 50 patients with passing PSQA results were included. The analysis threshold was set at 10% of the global maximum. Point dose differences were within the acceptable range of ±5% for all measurements. Gamma passing rates (GPRs) based on 3%/3 mm, 2%/2 mm, and 1%/1 mm were analyzed for all measurements. Correlations between treatment plan parameters and PSQA results were analyzed using Pearson’s correlation, to investigate the impacts of the PSQA results among all of the clinical cases. Plan complexity was evaluated to analyze the dosimetric accuracy by quantifying the degree of modulation according to each plan. A paired two-tailed Student’s t-test was performed to calculate the p-value using R software. The average difference in GPRs between the two detectors was statistically significant—except in the H&N, rectum, and breast cases with 3%/3 mm criteria. There were no significant correlations found between the modulation index and GPR results in any of the cases. These PSQA systems can be used interchangeably for routine PSQA during the delivery of radiation therapy

Abstract

Many institutions still perform patient-specific quality assurance (PSQA) using 2D array devices. However, no re- ports in the literature thus far have compared the correlations between PSQA results and treatment planning parameters to evaluate the impact of QA on IMRT. The purpose of this study is to compare the results of PSQA between MapCHECK and Octavius for radiation therapy. A total of 50 patients with passing PSQA results were included. The analysis threshold was set at 10% of the global maximum. Point dose differences were within the acceptable range of ±5% for all measurements. Gamma passing rates (GPRs) based on 3%/3 mm, 2%/2 mm, and 1%/1 mm were analyzed for all measurements. Correlations between treatment plan parameters and PSQA results were analyzed using Pearson’s correlation, to investigate the impacts of the PSQA results among all of the clinical cases. Plan complexity was evaluated to analyze the dosimetric accuracy by quantifying the degree of modulation according to each plan. A paired two-tailed Student’s t-test was performed to calculate the p-value using R software. The average difference in GPRs between the two detectors was statistically significant—except in the H&N, rectum, and breast cases with 3%/3 mm criteria. There were no significant correlations found between the modulation index and GPR results in any of the cases. These PSQA systems can be used interchangeably for routine PSQA during the delivery of radiation therapy

관련 논문AI추천

A Comparison of Patient-specific Delivery Quality Assurance (DQA) Devices in Radiation Therapy장경환대한방사선과학회 · Journal of Radiological Science and Technology2023 · 제46권 제3호Feasibility on Statistical Process Control Analysis of Delivery Quality Assurance in Helical Tomotherapy장경환대한방사선과학회 · Journal of Radiological Science and Technology2022 · 제45권 제6호Evaluation of DQA for Tomotherapy using 3D Volumetric Phantom김정구, 이상욱대한방사선과학회 · Journal of Radiological Science and Technology2016 · 제39권 제4호Clinical performance of FractionLab in patient-specific quality assurance for intensity-modulated radiotherapy: a retrospective study김승엽, 박재원, 예지원, 오세안, Jaehyeon Park영남대학교 의과대학 · Journal of Yeungnam Medical Science2022 · 제39권 제2호소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가강동진, 권경태, 김연래, 민정환, 신영주, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제1호Prediction and classification of gamma passing rate for patient-specific quality assurance using machine learning models based on radiomics features and beam parametersDong Zhiwei, Kambondo Alvin, Li Qian, Liu Fangyu, Qie Shuai, Qi Xiaolu, Shi Hongyun, Wang Jie, Zhang Qianru한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제10호

이 간행물의 다른 논문 (10건)