Anterior Ocular Biometrics Using Placido-scanning-slit System, Rotating Scheimpflug Tomography, and Swept-source Optical Coherence Tomography

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류안과학

초록

Purpose: To compare anterior biometry measurements using placido-scanning-slit topography, rotating Scheimpflug tomography,and swept-source optical coherence tomography. Methods: A retrospective review consisted of 80 eyes of 49 participants who underwent anterior chamber depth (ACD),central corneal thickness (CCT), and keratometry examination on the same day. We used placido-scanning-slit topography(ORBscan II), rotating Scheimpflug tomography (Pentacam HR), and swept-source optical coherence tomography (CASIA SS-1000). The intraclass correlation coefficients and Bland-Altman plots were used to evaluate the agreement and differencesbetween measurements. Results: The mean ACD values were 2.88 ± 0.43, 2.82 ± 0.50, and 2.68 ± 0.44 mm; and the mean CCT values were 536.96 ±31.19, 543.79 ± 31.04, and 561.41 ± 32.60 μm; and the mean keratometry (Km) were 43.81 ± 1.69, 43.81 ± 1.77, and 44.65± 1.95 diopters; as measured by CASIA SS-1000, Pentacam HR, and ORBscan II, respectively. Among the three devices, ACDwas deepest to shallowest in the order of CASIA SS-1000, Pentacam HR, and ORBscan II (p < 0.05). The CCT was thickest tothinnest in the order of ORBscan II, Pentacam HR, and CASIA SS-1000 (p < 0.05). No significant differences in Km values wereexamined between CASIA SS-1000 and Pentacam HR, whereas ORBscan II overestimated Km with a statistically significantdifference compared to the other two devices. Conclusions: High level of agreement was found between CASIA SS-1000 and Pentacam HR for anterior parameters, includingACD, CCT, and Km, suggesting interchangeability. However, ORBscan II measurements differed considerably with the measurementsobtained from the other two devices; therefore, it should not be used interchangeably. However, further studieswith repeatability test should be considered in order to elucidate the reliability of each device.

Abstract

Purpose: To compare anterior biometry measurements using placido-scanning-slit topography, rotating Scheimpflug tomography,and swept-source optical coherence tomography. Methods: A retrospective review consisted of 80 eyes of 49 participants who underwent anterior chamber depth (ACD),central corneal thickness (CCT), and keratometry examination on the same day. We used placido-scanning-slit topography(ORBscan II), rotating Scheimpflug tomography (Pentacam HR), and swept-source optical coherence tomography (CASIA SS-1000). The intraclass correlation coefficients and Bland-Altman plots were used to evaluate the agreement and differencesbetween measurements. Results: The mean ACD values were 2.88 ± 0.43, 2.82 ± 0.50, and 2.68 ± 0.44 mm; and the mean CCT values were 536.96 ±31.19, 543.79 ± 31.04, and 561.41 ± 32.60 μm; and the mean keratometry (Km) were 43.81 ± 1.69, 43.81 ± 1.77, and 44.65± 1.95 diopters; as measured by CASIA SS-1000, Pentacam HR, and ORBscan II, respectively. Among the three devices, ACDwas deepest to shallowest in the order of CASIA SS-1000, Pentacam HR, and ORBscan II (p < 0.05). The CCT was thickest tothinnest in the order of ORBscan II, Pentacam HR, and CASIA SS-1000 (p < 0.05). No significant differences in Km values wereexamined between CASIA SS-1000 and Pentacam HR, whereas ORBscan II overestimated Km with a statistically significantdifference compared to the other two devices. Conclusions: High level of agreement was found between CASIA SS-1000 and Pentacam HR for anterior parameters, includingACD, CCT, and Km, suggesting interchangeability. However, ORBscan II measurements differed considerably with the measurementsobtained from the other two devices; therefore, it should not be used interchangeably. However, further studieswith repeatability test should be considered in order to elucidate the reliability of each device.

참고문헌 (41)

  1. Guber I, "Correcting interdevice bias of horizontal white-to-white and sulcus-to-sulcus measures used for implantable collamer lens sizing", Am J Ophthalmol, 161, pp.116~125 (2016)
  2. Randleman JB, "Risk assessment for ectasia after corneal refractive surgery", Ophthalmology, 115, pp.37~50 (2008)
  3. Li X, "Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness", Ann Transl Med, 8, pp.857~ (2020)
  4. Wylegala E, "Anterior segment imaging : Fourier-domain optical coherence tomography versus time-domain optical coherence tomography", J Cataract Refract Surg, 35, pp.1410~1414 (2009)
  5. Amano S, "Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography", Ophthalmology, 113, pp.937~941 (2006)
  6. Pflugfelder SC, "Corneal thickness indices discriminate between keratoconus and contact lens-induced corneal thinning", Ophthalmology, 109, pp.2336~2341 (2002)
  7. Randleman JB, "Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography", Br J Ophthalmol, 99, pp.1176~1181 (2015)
  8. Binder PS, "Analysis of ectasia after laser in situ keratomileusis : risk factors", J Cataract Refract Surg, 33, pp.1530~1538 (2007)
  9. Reddy AR, "Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and IOLMaster", J Cataract Refract Surg, 30, pp.1268~1271 (2004)
  10. Pinero DP, "Technologies for anatomical and geometric characterization of the corneal structure and anterior segment : a review", Semin Ophthalmol, 30, pp.161~170 (2015)
  11. Ghoreishi SM, "Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes", Int Ophthalmol, 37, pp.965~971 (2017)
  12. Fukuda S, "Comparison of three-dimensional optical coherence tomography and combining a rotating Scheimpflug camera with a Placido topography system for forme fruste keratoconus diagnosis", Br J Ophthalmol, 97, pp.1554~1559 (2013)
  13. Lee YW, "Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems", J Cataract Refract Surg, 41, pp.1018~1029 (2015)
  14. Szalai E, "Reliability and repeatability of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus", J Cataract Refract Surg, 38, pp.485~494 (2012)
  15. Paul T, "Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system", J Cataract Refract Surg, 34, pp.1906~1912 (2008)
  16. Lackner B, "Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound", Optom Vis Sci, 82, pp.892~899 (2005)
  17. 심형섭, "중심각막두께측정에 대한 전안부 광간섭단층촬영계의 유용성", 대한안과학회지, 48(12), pp.1643~1648 (2007), 대한안과학회
  18. Ho T, "Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia", J Cataract Refract Surg, 33, pp.1177~1182 (2007)
  19. Hashemi H, "Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia", J Cataract Refract Surg, 33, pp.1701~1707 (2007)
  20. Kanellopoulos AJ, "Scheimpflug vs scanning-slit corneal tomography : comparison of corneal and anterior chamber tomography indices for repeatability and agreement in healthy eyes", Clin Ophthalmol, 14, pp.2583~2592 (2020)
  21. Angmo D, "Evaluation of anterior segment parameters with two anterior segment optical coherence tomography systems: Visante and Casia, in primary angle closure disease", Indian J Ophthalmol, 67, pp.500~504 (2019)
  22. Hoshikawa R, "Comparison of conventional keratometry and total keratometry in normal eyes", Biomed Res Int, 2020, pp.8075924~ (2020)
  23. Cairns G, "Orbscan computerized topography:attributes, applications, and limitations", J Cataract Refract Surg, 31, pp.205~220 (2005)
  24. Zhao Y, "The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging", Eye Vis(Lond), 7, pp.32~ (2020)
  25. Pardeshi AA, "Intradevice repeatability and interdevice agreement of ocular biometric measurements : a comparison of two swept-source anterior segment OCT devices", Transl Vis Sci Technol, 9, pp.14~ (2020)
  26. Kumar M, "Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes", Indian J Ophthalmol, 65, pp.288~294 (2017)
  27. 박주연, "세 가지 검사방법을 이용한 각막두께 및 전방깊이의 비교", 대한안과학회지, 50(5), pp.664~669 (2009), 대한안과학회
  28. Utine CA, "Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes", Acta Ophthalmol, 87, pp.386~391 (2009)
  29. Yazici AT, "Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam", J Refract Surg, 26, pp.127~133 (2010)
  30. Zhang XF, "Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation", Int J Ophthalmol, 13, pp.625~631 (2020)
  31. Obstfeld H, "Crystalline lens accommodation and anterior chamber depth", Ophthalmic Physiol Opt, 9, pp.36~40 (1989)
  32. Rozema JJ, "Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices", Am J Ophthalmol, 158, pp.1111~1120 (2014)
  33. Yekta AA, "Comparison of central corneal thickness measurements with Pentacam, Orbscan II, and ultrasound pachymeter", Iran J Ophthalmol, 21, pp.51~57 (2009)
  34. Buehl W, "Comparison of three methods of measuring corneal thickness and anterior chamber depth", Am J Ophthalmol, 141, pp.7~12 (2006)
  35. Rosa N, "Comparison of central corneal thickness measured with Orbscan and Pentacam", J Refract Surg, 23, pp.895~899 (2007)
  36. Jhanji V, "Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes", Clin Exp Ophthalmol, 41, pp.735~745 (2013)
  37. Krysik K, "Scheimpflug camera and swept-source optical coherence tomography in pachymetry evaluation of diabetic patients", J Ophthalmol, 2019, pp.4532657~ (2019)
  38. Schroder S, "Comparison of corneal tomography: repeatability, precision, misalignment, mean elevation, and mean pachymetry", Curr Eye Res, 43, pp.709~716 (2018)
  39. Doughty MJ, "The Orbscan acoustic(correction)factor for central corneal thickness measures of normal human corneas", Eye Contact Lens, 36, pp.106~115 (2010)
  40. Tajbakhsh Z, "Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer", Ophthalmic Physiol Opt, 32, pp.539~546 (2012)
  41. Meyer JJ, "Repeatability and agreement of Orbscan II, Pentacam HR, and Galilei tomography systems in corneas with keratoconus", Am J Ophthalmol, 175, pp.122~128 (2017)

관련 논문AI추천

이 간행물의 다른 논문 (13건)