Comparison of Anterior Segment Measurements with a New Multifunctional Unit and Five Other Devices

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류안과학

초록

Purpose: To evaluate the clinical availability of a multifunctional ocular biometric unit, MR-6000, for simultaneous keratometry,tonometry, topography, and pachymetry evaluation, and compare anterior segment measurements with five otherdevices: autokeratometer (KR-1), Scheimpflug camera (Pentacam HR), swept-source optical coherence tomography (IOLMaster700), Placido disk scanning-slit topography (Orbscan II), and noncontact tonometry (FT-1000). Methods: Thirty eyes from thirty patients who visited Severance Hospital for cataract surgery were examined using MR-6000 and the other devices. The mean keratometry, central corneal thickness (CCT), white-to-white (WTW) distance, and intraocularpressure (IOP) values were compared. Repeated measures analysis of variance, Wilcoxon signed-rank test, intraclasscorrelation coefficient (ICC), and Bland-Altman plot were used to assess the correlation and agreement between devices. Results: Thirty eyes of thirty patients were evaluated. Statistically significant differences in mean keratometry between MR-6000, KR-1, Pentacam HR, and IOLMaster 700 were not observed (p > 0.05). All five devices, including Orbscan II, had almostperfect agreement in measuring keratometry (ICC > 0.80, p < 0.05). CCT measured by MR-6000 was significantly differentfrom that of Pentacam HR and Orbscan II measurements (p < 0.05) but correlated with that of Pentacam HR and OrbscanII measurements (ICC > 0.60, p < 0.05). The WTW distance measured by MR-6000 was not significantly different from thatmeasured by IOLMaster 700 but was different from that measured by Orbscan II. IOP measured by MR-6000 was not correlatedwith FT-1000. Conclusions: Keratometric values obtained through MR-6000 can be used interchangeably with other devices based on goodcorrelation and agreement. However, the CCT, WTW, and IOP values were not interchangeable with a single multifunctionalunit for cataract surgery preoperative examination.

Abstract

Purpose: To evaluate the clinical availability of a multifunctional ocular biometric unit, MR-6000, for simultaneous keratometry,tonometry, topography, and pachymetry evaluation, and compare anterior segment measurements with five otherdevices: autokeratometer (KR-1), Scheimpflug camera (Pentacam HR), swept-source optical coherence tomography (IOLMaster700), Placido disk scanning-slit topography (Orbscan II), and noncontact tonometry (FT-1000). Methods: Thirty eyes from thirty patients who visited Severance Hospital for cataract surgery were examined using MR-6000 and the other devices. The mean keratometry, central corneal thickness (CCT), white-to-white (WTW) distance, and intraocularpressure (IOP) values were compared. Repeated measures analysis of variance, Wilcoxon signed-rank test, intraclasscorrelation coefficient (ICC), and Bland-Altman plot were used to assess the correlation and agreement between devices. Results: Thirty eyes of thirty patients were evaluated. Statistically significant differences in mean keratometry between MR-6000, KR-1, Pentacam HR, and IOLMaster 700 were not observed (p > 0.05). All five devices, including Orbscan II, had almostperfect agreement in measuring keratometry (ICC > 0.80, p < 0.05). CCT measured by MR-6000 was significantly differentfrom that of Pentacam HR and Orbscan II measurements (p < 0.05) but correlated with that of Pentacam HR and OrbscanII measurements (ICC > 0.60, p < 0.05). The WTW distance measured by MR-6000 was not significantly different from thatmeasured by IOLMaster 700 but was different from that measured by Orbscan II. IOP measured by MR-6000 was not correlatedwith FT-1000. Conclusions: Keratometric values obtained through MR-6000 can be used interchangeably with other devices based on goodcorrelation and agreement. However, the CCT, WTW, and IOP values were not interchangeable with a single multifunctionalunit for cataract surgery preoperative examination.

참고문헌 (46)

  1. Anayol MA, "Comparison of central corneal thickness, thinnest corneal thickness, anterior chamber depth, and simulated keratometry using galilei, Pentacam, and Sirius devices", Cornea, 33, pp.582~586 (2014)
  2. Huang J, "Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry", J Cataract Refract Surg, 37, pp.341~348 (2011)
  3. Hernandez-Camarena JC, "Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry", J Refract Surg, 30, pp.616~621 (2014)
  4. Shammas HJ, "Repeatability and reproducibility of biometry and keratometry measurements using a noncontact optical low-coherence reflectometer and keratometer", Am J Ophthalmol, 153, pp.55~61 (2012)
  5. Menassa N, "Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems", J Cataract Refract Surg, 34, pp.1742~1747 (2008)
  6. Rabsilber TM, "Anterior chamber measurements using Pentacam rotating Scheimpflug camera", J Cataract Refract Surg, 32, pp.456~459 (2006)
  7. 이동민, "두 가지 장비를 이용한 각막 지형도 검사의 비교", 대한안과학회지, 53(11), pp.1584~1590 (2012), 대한안과학회
  8. Speicher L, "Intra-ocular lens calculation status after corneal refractive surgery", Curr Opin Ophthalmol, 12, pp.17~29 (2001)
  9. Rose LT, "Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound : biometry for intraocular lens calculation", Clin Exp Ophthalmol, 31, pp.121~124 (2003)
  10. Connors R 3rd, "Accuracy and reproducibility of biometry using partial coherence interferometry", J Cataract Refract Surg, 28, pp.235~238 (2002)
  11. Kim KY, "Comparison study of the axial length measured using the new swept-source optical coherence tomography ANTERION and the partial coherence interferometry IOL Master", PLoS One, 15, pp.e0244590~ (2020)
  12. Sheng H, "Ocular component measurement using the Zeiss IOLMaster", Optom Vis Sci, 81, pp.27~34 (2004)
  13. Hoffer KJ, "Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry", J Cataract Refract Surg, 42, pp.1165~1172 (2016)
  14. Wang Q, "A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices", PLoS One, 7, pp.e45607~ (2012)
  15. Landis JR, "The measurement of observer agreement for categorical data", Biometrics, 33, pp.159~174 (1977)
  16. Mamalis N, "Complications of foldable intraocular lenses requiring explanation or secondary intervention : 1998 survey", J Cataract Refract Surg, 26, pp.766~772 (2000)
  17. Holladay JT, "Improving the predictability of intraocular lens power calculations", Arch Ophthalmol, 104, pp.539~541 (1986)
  18. Norrby S, "Sources of error in intraocular lens power calculation", J Cataract Refract Surg, 34, pp.368~376 (2008)
  19. 조동현, "각막굴절교정수술 후 인공수정체 도수 결정시 Orbscan II 비디오각막경의 유용성", 대한안과학회지, 50(11), pp.1730~1734 (2009), 대한안과학회
  20. 김승희, "종합병원의 외래진료대기시간과 지연요인 분석", 보건과 복지, 10, pp.107~120 (2008), 한국보건복지학회[DOI]
  21. 김일권, "의료서비스의 대기시간 요인이 고객만족과 서비스 가치에 미치는 영향", 병원경영학회지, 15(3), pp.47~68 (2010), 한국병원경영학회
  22. 한정모, "자동각막굴절계, IOL Master, 각막지형도의 각막굴절력 및 각막난시 측정의 비교 분석", 대한안과학회지, 52(12), pp.1427~1433 (2011), 대한안과학회
  23. Whang WJ, "Comparison of refractive outcomes using five devices for the assessment of preoperative corneal power", Clin Exp Ophthalmol, 40, pp.425~432 (2012)
  24. 김소윤, "여러 가지 기기에서의 각막곡률 비교", 대한안과학회지, 53(5), pp.618~625 (2012), 대한안과학회[DOI]
  25. Lee YW, "Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems", J Cataract Refract Surg, 41, pp.1018~1029 (2015)
  26. Tajbakhsh Z, "Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer", Ophthalmic Physiol Opt, 32, pp.539~546 (2012)
  27. Kumar M, "Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes", Indian J Ophthalmol, 65, pp.288~294 (2017)
  28. Meyer JJ, "Repeatability and agreement of Orbscan II, Pentacam HR, and Galilei tomography systems in corneas with keratoconus", Am J Ophthalmol, 175, pp.122~128 (2017)
  29. Buehl W, "Comparison of three methods of measuring corneal thickness and anterior chamber depth", Am J Ophthalmol, 141, pp.7~12 (2006)
  30. 박주연, "세 가지 검사방법을 이용한 각막두께 및 전방깊이의 비교", 대한안과학회지, 50(5), pp.664~669 (2009), 대한안과학회
  31. 신영주, "두 가지 각막지형도 검사에서의 각막 측정값 비교", 대한안과학회지, 48(5), pp.637~641 (2007), 대한안과학회
  32. Hashemi H, "Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia", J Cataract Refract Surg, 33, pp.1701~1707 (2007)
  33. Lackner B, "Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound", Optom Vis Sci, 82, pp.892~899 (2005)
  34. 윤상문, "에이엘 스캔과 펜타캠을 이용한 전안부 계측값의 비교", 대한안과학회지, 55(6), pp.801~808 (2014), 대한안과학회
  35. Savini G, "Comparison of anterior segment measurements by 3 Scheimpflug tomographers and 1 Placido corneal topographer", J Cataract Refract Surg, 37, pp.1679~1685 (2011)
  36. Kojima T, "Optimization of an implantable collamer lens sizing method using high-frequency ultrasound biomicroscopy", Am J Ophthalmol, 153, pp.632~637 (2012)
  37. Reinstein DZ, "Comparison of postoperative vault height predictability using whiteto-white or sulcus diameter-based sizing for the visian implantable collamer lens", J Refract Surg, 29, pp.30~35 (2013)
  38. 정윤, "IOLMaster, Lenstar, Orbscan 및 수동 방법을 이용한 각막윤부직경 비교", 대한안과학회지, 54(8), pp.1187~1192 (2013), 대한안과학회
  39. Dinc UA, "Assessment and comparison of anterior chamber dimensions using various imaging techniques", Ophthalmic Surg Lasers Imaging, 41, pp.115~122 (2010)
  40. Oh J, "Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy", Ophthalmology, 114, pp.1685~1688 (2007)
  41. Domínguez-Vicent A, "Device interchangeability on anterior chamber depth and white-to-white measurements : a thorough literature review", Int J Ophthalmol, 9, pp.1057~1065 (2016)
  42. Reinstein DZ, "Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using artemis VHF digital ultrasound", J Refract Surg, 25, pp.185~194 (2009)
  43. Nam SW, "Buffering zone of implantable Collamer lens sizing in V4c", BMC Ophthalmol, 17, pp.260~ (2017)
  44. Graf M, "Significance of the corneal thickness in non-contact tonometry", Klin Monbl Augenheilkd, 199, pp.183~186 (1991)
  45. Recep OF, "Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry", J Cataract Refract Surg, 27, pp.1787~1791 (2001)
  46. Altman DG, "Statistics notes : the normal distribution", BMJ, 310, pp.298~ (1995)

관련 논문AI추천

이 간행물의 다른 논문 (11건)