Agreement between Scheimpflug Camera and the Swept-source Optical Coherence Tomography Measurements in Keratometry and Higher-order Aberrations

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류안과학

초록

Purpose: To evaluate the compatibility of corneal curvature and astigmatism, and higher-order aberrations (HOAs) measuredby the Scheimpflug camera Pentacam HR and the swept-source optical coherence tomography ANTERION. Methods: This prospective study included normal subjects with no ophthalmic history. Steep keratometry (K), flat K, astigmatismand its axis of the anterior and posterior surfaces, total corneal power, and HOAs using the two instruments werecompared. To compare the mean values of the measurements, a paired t-test was used. Bland-Altman analysis was applied toassess the agreement between the two devices. Results: Fifty-three eyes of 53 subjects were evaluated. There were statistically significant differences for steep K, astigmatism,and vector J0, J45 in the anterior surface and total corneal power between the two devices (p < 0.05). There werealso significant differences in the most of the keratometric values of the posterior corneal surface (p < 0.05) except J0 (p =0.410). Both devices showed strong positive correlations in steep K, flat K, astigmatism (r > 0.81, p < 0.001) with wide rangesof a 95% limit of agreement. Vectoral components were significantly correlated (r > 0.78, p < 0.001) with narrow 95% limit ofagreement, except J45 of the posterior surface (r = 0.39, p = 0.004). In the corneal HOAs, there were statistically significantdifferences in the vertical coma, horizontal trefoil, spherical aberration, and root mean square of each fifth- and sixth-orderZernike coefficient (p = 0.043, p = 0.041, p < 0.001, p < 0.001, and p < 0.001, respectively). Other HOAs showed moderateto strong positive correlations (r > 0.37, p < 0.05). Most HOAs, except for the horizontal trefoil, showed clinically acceptableagreements. The total root mean square of HOAs was not significantly different between the two devices (p = 0.122). Conclusions: Most of the keratometric values cannot be used interchangeably. However, the vectoral component of astigmatismshowed clinically good agreement. Several HOAs have statistically significant differences; however, almost all HOAsshowed acceptable agreements, except for the horizontal trefoil.

Abstract

Purpose: To evaluate the compatibility of corneal curvature and astigmatism, and higher-order aberrations (HOAs) measuredby the Scheimpflug camera Pentacam HR and the swept-source optical coherence tomography ANTERION. Methods: This prospective study included normal subjects with no ophthalmic history. Steep keratometry (K), flat K, astigmatismand its axis of the anterior and posterior surfaces, total corneal power, and HOAs using the two instruments werecompared. To compare the mean values of the measurements, a paired t-test was used. Bland-Altman analysis was applied toassess the agreement between the two devices. Results: Fifty-three eyes of 53 subjects were evaluated. There were statistically significant differences for steep K, astigmatism,and vector J0, J45 in the anterior surface and total corneal power between the two devices (p < 0.05). There werealso significant differences in the most of the keratometric values of the posterior corneal surface (p < 0.05) except J0 (p =0.410). Both devices showed strong positive correlations in steep K, flat K, astigmatism (r > 0.81, p < 0.001) with wide rangesof a 95% limit of agreement. Vectoral components were significantly correlated (r > 0.78, p < 0.001) with narrow 95% limit ofagreement, except J45 of the posterior surface (r = 0.39, p = 0.004). In the corneal HOAs, there were statistically significantdifferences in the vertical coma, horizontal trefoil, spherical aberration, and root mean square of each fifth- and sixth-orderZernike coefficient (p = 0.043, p = 0.041, p < 0.001, p < 0.001, and p < 0.001, respectively). Other HOAs showed moderateto strong positive correlations (r > 0.37, p < 0.05). Most HOAs, except for the horizontal trefoil, showed clinically acceptableagreements. The total root mean square of HOAs was not significantly different between the two devices (p = 0.122). Conclusions: Most of the keratometric values cannot be used interchangeably. However, the vectoral component of astigmatismshowed clinically good agreement. Several HOAs have statistically significant differences; however, almost all HOAsshowed acceptable agreements, except for the horizontal trefoil.

참고문헌 (39)

  1. Konstantopoulos A, "Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?", Br J Ophthalmol, 91, pp.551~557 (2007)
  2. Chen W, "Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: repeatability and agreement", J Cataract Refract Surg, 38, pp.1626~1632 (2012)
  3. Ventura BV, "Comparison of corneal power, astigmatism, and wavefront aberration measurements obtained by a point-source color light-emitting diode-based topographer, a Placido-disk topographer, and a combined Placido and dual Scheimpflug device", J Cataract Refract Surg, 41, pp.1658~1671 (2015)
  4. Fabian E, "Prediction accuracy of total keratometry compared to standard keratometry using different intraocular lens power formulas", J Refract Surg, 35, pp.362~368 (2019)
  5. Lawless M, "Total keratometry in intraocular lens power calculations in eyes with previous laser refractive surgery", Clin Exp Ophthalmol, 48, pp.749~756 (2020)
  6. Reitblat O, "Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: comparison of methodologies", J Cataract Refract Surg, 42, pp.217~225 (2016)
  7. Savini G, "An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with toric intraocular lenses", Invest Ophthalmol Vis Sci, 56, pp.827~835 (2015)
  8. Savini G, "Optimized keratometry and total corneal astigmatism for toric intraocular lens calculation", J Cataract Refract Surg, 43, pp.1140~1148 (2017)
  9. Kern C, "Comparing refractive outcomes of a standard industry toric IOL calculator using anterior corneal astigmatism and total corneal refractive power", Graefes Arch Clin Exp Ophthalmol, 258, pp.345~350 (2020)
  10. Gordon-Shaag A, "Aberrations and topography in normal, keratoconus-suspect, and keratoconic eyes", Optom Vis Sci, 89, pp.411~418 (2012)
  11. Pinero DP, "Characterization of corneal structure in keratoconus", J Cataract Refract Surg, 38, pp.2167~2183 (2012)
  12. Yamaguchi T, "Effect of spherical aberration on visual function under photopic and mesopic conditions after cataract surgery", J Cataract Refract Surg, 35, pp.57~63 (2009)
  13. Iseli HP, "Corneal and total wavefront aberrations in phakic and pseudophakic eyes after implantation of monofocal foldable intraocular lenses", J Cataract Refract Surg, 32, pp.762~771 (2006)
  14. Song IS, "Higher-order aberrations associated with better near visual acuity in eyes with aspheric monofocal IOLs", J Refract Surg, 30, pp.442~446 (2014)
  15. Moshirfar M, "Use of presbyopia-correcting intraocular lenses in patients with prior corneal refractive surgery", Curr Opin Ophthalmol, 32, pp.45~53 (2021)
  16. Tana-Rivero P, "Repeatability of whole-cornea measurements using a new swept-source optical coherence tomographer", Eur J Ophthalmol, 31, pp.1709~1719 (2021)
  17. Montes-Mico R, "Assessment of anterior segment measurements using a high-resolution imaging device", Expert Rev Med Devices, 17, pp.969~979 (2020)
  18. Ruiz-Mesa R, "Ocular biometric repeatability using a new high-resolution swept-source optical coherence tomographer", Expert Rev Med Devices, 17, pp.591~597 (2020)
  19. Fisus AD, "Comparison of 2 sweptsource optical coherence tomography-based biometry devices", J Cataract Refract Surg, 47, pp.87~92 (2021)
  20. Pardeshi AA, "Intradevice repeatability and interdevice agreement of ocular biometric measurements: a comparison of two swept-source anterior segment OCT devices", Transl Vis Sci Technol, 9, pp.14~ (2020)
  21. Tana-Rivero P, "Agreement of white-to-white measurements with sweptsource OCT, Scheimpflug and color LED devices", Int Ophthalmol, 41, pp.57~65 (2021)
  22. Tana-Rivero P, "Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer", J Cataract Refract Surg, 47, pp.488~495 (2021)
  23. Fan R, "Applications of corneal topography and tomography: a review", Clin Exp Ophthalmol, 46, pp.133~146 (2018)
  24. McAlinden C, "Corneal aberrations measured with a high-resolution Scheimpflug tomographer: repeatability and reproducibility", J Cataract Refract Surg, 46, pp.581~590 (2020)
  25. Asam JS, "High resolution imaging in microscopy and ophthalmology: new frontiers in biomedical optics", pp.285~299 (2019), Springer
  26. Thibos LN, "Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error", Optom Vis Sci, 74, pp.367~375 (1997)
  27. Thibos LN, "Power vector analysis of the optical outcome of refractive surgery", J Cataract Refract Surg, 27, pp.80~85 (2001)
  28. Klijn S, "Accuracy of total corneal astigmatism measurements with a Scheimpflug imager and a color light-emitting diode corneal topographer", Am J Ophthalmol, 167, pp.72~78 (2016)
  29. Park DY, "Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation", BMC Ophthalmol, 17, pp.156~ (2017)
  30. Ozyol P, "Agreement between swept-source optical biometry and Scheimpf lug-based topography measurements of anterior segment parameters", Am J Ophthalmol, 169, pp.73~78 (2016)
  31. Cui XH, "Comparison of keratometric measurements between color light-emitting diode topography and Scheimpflug camera", BMC Ophthalmol, 19, pp.98~ (2019)
  32. Pan C, "A comparative study of total corneal power using a ray tracing method obtained from 3different Scheimpflug camera devices", Am J Ophthalmol, 216, pp.90~98 (2020)
  33. Savini G, "Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras", J Cataract Refract Surg, 44, pp.471~478 (2018)
  34. Sel S, "Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements", Cont Lens Anterior Eye, 40, pp.318~322 (2017)
  35. Piccinini AL, "Higher-order aberration measurements: comparison between Scheimpflug and dual Scheimpflug-Placido technology in normal eyes", J Cataract Refract Surg, 45, pp.490~494 (2019)
  36. 김다영, "발광다이오드 각막형태분석기와 샤임플러그 전안부사진기의 각막난시 및 각막고위수차 비교", 대한안과학회지, 60(10), pp.922~928 (2019), 대한안과학회
  37. 신지영, "두 가지 장비를 이용한 각막곡률 및 각막고위수차의 비교", 대한안과학회지, 55(12), pp.1758~1764 (2014), 대한안과학회
  38. Karimian F, "Higher-order aberrations in myopic eyes", J Ophthalmic Vis Res, 5, pp.3~9 (2010)
  39. McAlinden C, "A comprehensive evaluation of the precision (repeatability and reproducibility)of the Oculus Pentacam HR", Invest Ophthalmol Vis Sci, 52, pp.7731~7737 (2011)

관련 논문AI추천

이 간행물의 다른 논문 (13건)