Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류안과학

초록

Purpose: To evaluate repeatability and agreement of chord mu between Scheimpflug tomography (Pentacam HR) and sweptsource optical coherence tomography–based optical biometer (IOLMaster 700). Methods: In this retrospective study, 63 eyes from 33 patients were included. Chord mu, X and Y Cartesian distances between the corneal vertex and the pupil center (Px and Py), and the pupil diameter were compared using two instruments. Repeatability was evaluated using intraclass correlation coefficient (ICC), coefficient of variation (CoV), and within-subject standard deviation (Sw). Interdevice agreement was evaluated using paired t-tests and Bland-Altman plots. Results: Although Sw values for all parameters were similar between the two devices, CoV values of chord mu and pupil diameter were lower, and ICC values of those parameters were higher, in the IOLMaster 700 than in the Pentacam HR. Chord mu and pupil diameter values were higher in IOLMaster 700 than Pentacam HR (p < 0.01). The width of the 95% limit of agreement was wide for all parameters. Conclusions: IOLMaster 700 showed better repeatability than Pentacam HR in chord mu, Px, Py, and pupil diameter values. Because there were statistically significant differences and a low level of agreement in chord mu and pupil diameter values between the two devices, they cannot be used interchangeably.

Abstract

Purpose: To evaluate repeatability and agreement of chord mu between Scheimpflug tomography (Pentacam HR) and sweptsource optical coherence tomography–based optical biometer (IOLMaster 700). Methods: In this retrospective study, 63 eyes from 33 patients were included. Chord mu, X and Y Cartesian distances between the corneal vertex and the pupil center (Px and Py), and the pupil diameter were compared using two instruments. Repeatability was evaluated using intraclass correlation coefficient (ICC), coefficient of variation (CoV), and within-subject standard deviation (Sw). Interdevice agreement was evaluated using paired t-tests and Bland-Altman plots. Results: Although Sw values for all parameters were similar between the two devices, CoV values of chord mu and pupil diameter were lower, and ICC values of those parameters were higher, in the IOLMaster 700 than in the Pentacam HR. Chord mu and pupil diameter values were higher in IOLMaster 700 than Pentacam HR (p < 0.01). The width of the 95% limit of agreement was wide for all parameters. Conclusions: IOLMaster 700 showed better repeatability than Pentacam HR in chord mu, Px, Py, and pupil diameter values. Because there were statistically significant differences and a low level of agreement in chord mu and pupil diameter values between the two devices, they cannot be used interchangeably.

참고문헌 (29)

  1. Hashemi H, "Distribution of angle kappa measurements with Orbscan II in a population-based survey", J Refract Surg, 26, pp.966~971 (2010)
  2. Moshirfar M, "Angle kappa and its importance in refractive surgery", Oman J Ophthalmol, 6, pp.151~158 (2013)
  3. Fu Y, "Influence of angle kappa and angle alpha on visual quality after implantation of multifocal intraocular lenses", J Cataract Refract Surg, 45, pp.1258~1264 (2019)
  4. Khakshoor H, "Use of angle kappa in myopic photorefractive keratectomy", Clin Ophthalmol, 9, pp.193~195 (2015)
  5. Hayashi K, "Correlation between pupillary size and intraocular lens decentration and visual acuity of a zonal-progressive multifocal lens and a monofocal lens", Ophthalmology, 108, pp.2011~2017 (2001)
  6. de Vries NE, "Dissatisfaction after implantation of multifocal intraocular lenses", J Cataract Refract Surg, 37, pp.859~865 (2011)
  7. Harrer A, "Variability in angle κ and its influence on higher-order aberrations in pseudophakic eyes", J Cataract Refract Surg, 43, pp.1015~1019 (2017)
  8. Chang DH, "The subject-fixated coaxially sighted corneal light reflex : a clinical marker for centration of refractive treatments and devices", Am J Ophthalmol, 158, pp.863~874 (2014)
  9. Garza-Leon M, "Repeatability of ocular biometry with IOLMaster 700in subjects with clear lens", Int Ophthalmol, 37, pp.1133~1138 (2017)
  10. Michael R, "Feasibility and repeatability of ocular biometry measured with IOLMaster 700 in a large population-based study", Ophthalmic Physiol Opt, 43, pp.860~873 (2023)
  11. Park CY, "Measurement of angle kappa and centration in refractive surgery", Curr Opin Ophthalmol, 23, pp.269~275 (2012)
  12. Qi Y, "Role of angle κ in visual quality in patients with a trifocal diffractive intraocular lens", J Cataract Refract Surg, 44, pp.949~954 (2018)
  13. Narang P, "Application of Purkinje images for pinhole pupilloplasty and relevance to chord length mu", J Cataract Refract Surg, 45, pp.745~751 (2019)
  14. Bland M, "An introduction to medical statistics" (2000), Oxford University Press
  15. Lackner B, "Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan", Optom Vis Sci, 82, pp.858~861 (2005)
  16. Bullimore MA, "An evaluation of the IOLMaster 700", Eye Contact Lens, 45, pp.117~123 (2019)
  17. de Vries NE, "Intraocular straylight after implantation of the multifocal AcrySof ReSTOR SA60D3 diffractive intraocular lens", J Cataract Refract Surg, 34, pp.957~962 (2008)
  18. Dominguez-Vicent A, "Measurement of angle kappa with Orbscan II and Galilei G4 : effect of accommodation", Graefes Arch Clin Exp Ophthalmol, 252, pp.249~255 (2014)
  19. Basmak H, "Measurement of angle kappa with synoptophore and Orbscan II in a normal population", J Refract Surg, 23, pp.456~460 (2007)
  20. Martinez-Plaza E, "Reliability and agreement of apparent chord mu measurements between static and dynamic evaluations", J Cataract Refract Surg, 49, pp.21~28 (2023)
  21. Perez-Bartolome F, "Anterior-segment swept-source ocular coherence tomography and Scheimpflug imaging agreement for keratometry and pupil measurements in healthy eyes", J Clin Med, 10, pp.5789~ (2021)
  22. 여준형, "Measurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography", Korean Journal of Ophthalmology, 31(3), pp.257~262 (2017), 대한안과학회
  23. Qin M, "Comparison of preoperative angle kappa measurements in the eyes of cataract patients obtained from Pentacam Scheimpf lug system, optical low-coherence reflectometry, and ray-tracing aberrometry", BMC Ophthalmol, 22, pp.153~ (2022)
  24. Holladay JT, "Apparent chord mu and actual chord mu and their clinical value", J Cataract Refract Surg, 45, pp.1198~1199 (2019)
  25. Rodrigues PF, "Measurement of the visual axis through two different methods : quantification and differences for measuring chord μ", Arq Bras Oftalmol (2023)
  26. Rodriguez-Vallejo M, "Avoiding misinterpretations of kappa angle for clinical research studies with Pentacam", J Optom, 12, pp.71~73 (2019)
  27. Kanellopoulos AJ, "Scheimpflug vs scanning-slit corneal tomography : comparison of corneal and anterior chamber tomography indices for repeatability and agreement in healthy eyes", Clin Ophthalmol, 14, pp.2583~2592 (2020)
  28. Guclu H, "Agreement of anterior segment parameters between Schiempflug topography and swept-source optic coherence based optic biometry in keratoconus and healthy subjects", Eye Contact Lens, 47, pp.539~545 (2021)
  29. Guillon M, "The effects of age, refractive status, and luminance on pupil size", Optom Vis Sci, 93, pp.1093~1100 (2016)

관련 논문AI추천

이 간행물의 다른 논문 (12건)