전산화단층 모의치료장치의 정도관리 항목 제안

Proposal of CT Simulator Quality Assurance Items

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

A quality assurance of computed tomography(CT) have done seven items that were water attenuation coefficient, noise, homogeneity, spatial resolution, contrast resolution, slice thickness, artifact using by standard phantom. But there is no quality assurance items and methods for CT simulator at domestic institutions yet. Therefore the study aimed to access the CT dose index(CTDI), table tilting, image distortion, laser accuracy, table movement accuracy and CT seven items for CT simulator quality assurance. The CTDI at the center of the head phantom was 0.81 for 80 kVp, 1.55 for 100 kVp, 2.50 for 120 ㎜, 0.22 for 80 kVp at the center of the body phantom, 0.469 for 100 kVp, and 0.81 for 120 kVp. The table tilting was within the tolerance range of ±1.0° or less. Image distortion had 1 ㎜ distortion in the left and right images based on the center, and the laser accuracy was measured within ±2 ㎜ tolerance. The purpose of this study is to improve the quality assurance items suitable for the current situation in Korea in order to protect the normal tissues during the radiation treatment process and manage the CT simulator that is implemented to find the location of the tumor more clearly. In order to improve the accuracy of the CT simulator when looking at the results, the error range of each item should be small. It is hoped that the quality assurance items of the CT simulator will be improved by suggesting the quality assurance direction of the CT simulator in this study, and the results of radiation therapy will also improve.

Abstract

A quality assurance of computed tomography(CT) have done seven items that were water attenuation coefficient, noise, homogeneity, spatial resolution, contrast resolution, slice thickness, artifact using by standard phantom. But there is no quality assurance items and methods for CT simulator at domestic institutions yet. Therefore the study aimed to access the CT dose index(CTDI), table tilting, image distortion, laser accuracy, table movement accuracy and CT seven items for CT simulator quality assurance. The CTDI at the center of the head phantom was 0.81 for 80 kVp, 1.55 for 100 kVp, 2.50 for 120 ㎜, 0.22 for 80 kVp at the center of the body phantom, 0.469 for 100 kVp, and 0.81 for 120 kVp. The table tilting was within the tolerance range of ±1.0° or less. Image distortion had 1 ㎜ distortion in the left and right images based on the center, and the laser accuracy was measured within ±2 ㎜ tolerance. The purpose of this study is to improve the quality assurance items suitable for the current situation in Korea in order to protect the normal tissues during the radiation treatment process and manage the CT simulator that is implemented to find the location of the tumor more clearly. In order to improve the accuracy of the CT simulator when looking at the results, the error range of each item should be small. It is hoped that the quality assurance items of the CT simulator will be improved by suggesting the quality assurance direction of the CT simulator in this study, and the results of radiation therapy will also improve.

참고문헌 (14)

  1. 신동호, "의료용 모의치료기의 품질관리에 관한 연구", 의학물리, 13(2), pp.90~97 (2002), 한국의학물리학회
  2. Ministry of Health and Welfare, "Rules for the installation and operation of special medical equipment", ~ (2019)
  3. Quality assurance for computed-tomography simulators and the computed tomography-simulation process, "Sasa M", American Association of Physicists in Medicine, ~, pp.Report of the AAPM radiation therapy committee task group No. 66 (2005)
  4. Philippe D, "COMP report : CPQR technical quality control guidelines for CT simulators", J Appl Clin Med Phys, 19(2), pp.12~17 (2018)
  5. Zakaria GA, "Quality assurance in radiotherapy:Simulators, cone beam CT, EPID and immobilization devices" (2012), Academic Teaching Hospital of the University of Cologne
  6. 김영균, "Dual-energy 컴퓨터단층촬영에서 장비 제공선량과 측정선량 비교", 방사선기술과학, 41(5), pp.405~411 (2018), 대한방사선과학회[DOI]
  7. Hwang SH, "Efficiency evaluation of CT simulator QA phantom", The Journal of Korean Society for Radiation Therapy, 21(2), pp.89~95 (2009)
  8. Min CK, "A CT simulator phantom for geometrical test", The Journal of the Korean Society for Therapeutic Radiology and Oncology, 18(4), pp.329~336 (2000)
  9. Huaiqun G, "A positioning QA procedure of 2D", 2D(kV, 3D(CT, pp.CBCT)image matching for radiotherapy patient setup, Journal of Applied Clinical Medical Physics[DOI]
  10. Tomic N, "Image quality of radiotherapy CT simulator with different scanner bore size", Physica Medica, 45, pp.65~71 (2018)
  11. Li XN, "Quality assurance on accuracy of the external laser centre in CT simulator", Chinese Journal of Radiation Oncology, (6), pp.481~483 (2006)
  12. 구은회, "진단용 CT-모의치료기 테이블의 효율적인 교정 방법", 한국콘텐츠학회 논문지, 9(11), pp.254~261 (2009), 한국콘텐츠학회
  13. 안현준, "Acceptance Test and Clinical Commissioning of CT Simulator", 의학물리, 30(4), pp.160~166 (2019), 한국의학물리학회[DOI]
  14. 윤명성, "삼차원 프린팅 기술을 이용한 전산화단층영상 품질 측정용 팬텀 제작 및 비교 연구", 방사선기술과학, 41(6), pp.595~602 (2018), 대한방사선과학회[DOI]