Cone beam Computed Tomography(CBCT) is an increasing trend in clinical applications due to its ability to increase the accuracy of radiation therapy. However, this leaded to an increase in exposure dose. In this study, the simulation using Monte Carlo method is performed and the absorbed dose of CBCT is analyzed and standardized data is presented. First, after simulating the CBCT, the photon spectrum was analyzed to secure the reliability and the absorbed dose of the tissue in the human body was evaluated using the MIRD phantom. Compared with SRS-78, the photon spectrum of CBCT showed similar tendency, and the average absorbed dose of MIRD phantom was 8.12 ~ 25.88 mGy depending on the body site. This is about 1% of prescription dose, but dose management will be needed to minimize patient side effects and normal tissue damage.

Abstract

Cone beam Computed Tomography(CBCT) is an increasing trend in clinical applications due to its ability to increase the accuracy of radiation therapy. However, this leaded to an increase in exposure dose. In this study, the simulation using Monte Carlo method is performed and the absorbed dose of CBCT is analyzed and standardized data is presented. First, after simulating the CBCT, the photon spectrum was analyzed to secure the reliability and the absorbed dose of the tissue in the human body was evaluated using the MIRD phantom. Compared with SRS-78, the photon spectrum of CBCT showed similar tendency, and the average absorbed dose of MIRD phantom was 8.12 ~ 25.88 mGy depending on the body site. This is about 1% of prescription dose, but dose management will be needed to minimize patient side effects and normal tissue damage.

참고문헌 (23)

  1. 박천주, "Half-fan 모드를 이용한 방사선치료환자 위치교정을 위한 디지털영상 합성영상기술에 관한 예비연구", 의학물리, 21(4), pp.360~366 (2010), 한국의학물리학회
  2. Do. KM, "Patient Position Verification and Corrective Evaluation Using Cone Beam Computed Tomography (CBCT) in Intensity. modulated Radiation Therapy", The Journal of Korean Society for Radiation Therapy, 21(2), pp.83~88 (2009)
  3. Bang. SJ, "A Study of Image Quality and Exposed Dose by Field Size Changing on CBCT", The Journal of Korean Society for Radiation Therapy, 25(2), pp.175~180 (2013)
  4. 오승종, "kV Cone-beam CT를 사용한 치료준비에서 재구성 영상의품질이 표적 위치 결정에 미치는 영향", 의학물리, 17(4), pp.207~211 (2006), 한국의학물리학회
  5. Yoo. YS, "Implementation of KV Cone Beam CT for Image Guided Radiation Therapy", The Journal of Korean Society for Radiation Therapy, 19(1), pp.43~49 (2007)
  6. 이이성, "콘빔 전산화단층촬영(CBCT) 시스템에서 기계적 오류에 관한 연구", 방사선기술과학, 36(2), pp.123~129 (2013), 대한방사선과학회
  7. Park. BS, "Evaluation of Every Other Day - Cone Beam Computed Tomography in Image Guided Radiation Therapy for Prostate Cancer", Journal of the Korean Society for Radiotherapeutic Technology, 26(2), pp.289~295 (2014)
  8. Semelka RC, "Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI", Journal of Magnetic Resonance Imaging, 25(5), pp.900~909 (2007)[DOI]
  9. Song. YG, "Diagnostic Radiation Imaging Detector Implementation and Image Evaluation using Monte Carlo Simulation", pp.2013
  10. Lee. JK, "The study of radiation dose estimates for computed tomography using Monte Carlo method", pp.2011
  11. Lee. SH, "A comparative study of aluminum, copper, and lead as shielding materials in 6 MeV electron therapy", pp.2013
  12. 이정옥, "MCNPX를 이용한 선형가속기의 6 MeV 전자선에 대한 에너지분포 계산", 의학물리, 17(4), pp.224~231 (2006), 한국의학물리학회
  13. Lim. HJ, "Status of Domestic Linear Accelerator" (2017)
  14. Fouad AA, "On Board Cone Beam CT for Treatment Planning in Image Guided Radiotherapy", pp.2015
  15. Downes P, "Monte Carlo simulation and patient dosimetry for a kilovoltage cone-beam", Medical Physics, 36(9), pp.4156~4167 (2009)[DOI]
  16. Han EY, "Revisions to the ORNL serise of adult and pediatric computational phantoms for use with the MIRD schema", University of Florida Health Phys, 90(4), pp.337~356 (2006)
  17. Valentin J, "ICRP Publication 89:Basic Anatomical and Physiological Data for Use in Radiological Protection:Reference Values" (2003), Elsevier Publishing Company
  18. ICRU REPORT 46: Photon, Electron, Proton and Neutron Interaction Data for Body Tissues, "White. DR", Oxford Univ, ~ (1992)
  19. Park DO, "Development of Computational Procedure for Assessment of Patient Dose in Multi-Detector Computed Tomography", pp.2007
  20. 박일, "몬테카를로 방법론을 이용한 구내촬영 검사 시 환자 방사선량 평가", 방사선산업학회지, 10(3), pp.139~144 (2016), (사)한국방사선산업학회[DOI]
  21. Ding. GX, "Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy", Physics in Medicine and Biology, 52(6), pp.1595~1615 (2007)[DOI]
  22. Palm A, "Absorbed dose and dose rate using the Varian OBI 1.3 and 1.4 CBCT system", Journal of applied clinical medical physics, 11(1), pp.S93~ (2009)
  23. Moon. YM, "A study on the effective dose measurement for cone beam computed tomography using glass dosimeter", pp.2012