The effect of a titanium socket with a zirconia abutment on screw loosening after thermocycling in an internally connected implant: a preliminary study

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류치의학

초록

Purpose: The aim of this study was to investigate the effects of a titanium component for the zirconia abutment in the internal connection implant system on screw loosening under thermocycling conditions. Materials and Methods: Internal connection titanium abutments and external connection zirconia abutments with titanium sockets were connected respectively to screwshaped internal connection type titanium implants with 30 Ncm tightening. These implant-screw-abutment assemblies were divided into two groups of five specimens each; titanium abutments as control and zirconia abutments with titanium sockets as experimental group. The specimens were subjected to 2,000 thermocycles in water baths at 5°C and 55°C, with 60 seconds of immersion at each temperature. The removal torque values (RTVs) of the abutment screws of the specimen were measured before and after thermocycling. RTVs pre- and post-thermocycling were investigated in statistics. Results: There was not screw loosening identified by tactile and visual inspection in any of the specimens during or after thermocycling. The mean RTV difference for the control group and the experimental group were -1.34 ± 2.53 Ncm and -1.26 ± 2.06 Ncm, respectively. Statistical analysis using an independent t-test revealed that no significant differences were found in the mean RTV difference of the groups (P > 0.05). Conclusion: Within the limitations of this in vitro study, the titanium socket for the zirconia abutment did not show a significant effect on screw loosening under thermal stress compared to the titanium abutment in the internal connection implant.

Abstract

Purpose: The aim of this study was to investigate the effects of a titanium component for the zirconia abutment in the internal connection implant system on screw loosening under thermocycling conditions. Materials and Methods: Internal connection titanium abutments and external connection zirconia abutments with titanium sockets were connected respectively to screwshaped internal connection type titanium implants with 30 Ncm tightening. These implant-screw-abutment assemblies were divided into two groups of five specimens each; titanium abutments as control and zirconia abutments with titanium sockets as experimental group. The specimens were subjected to 2,000 thermocycles in water baths at 5°C and 55°C, with 60 seconds of immersion at each temperature. The removal torque values (RTVs) of the abutment screws of the specimen were measured before and after thermocycling. RTVs pre- and post-thermocycling were investigated in statistics. Results: There was not screw loosening identified by tactile and visual inspection in any of the specimens during or after thermocycling. The mean RTV difference for the control group and the experimental group were -1.34 ± 2.53 Ncm and -1.26 ± 2.06 Ncm, respectively. Statistical analysis using an independent t-test revealed that no significant differences were found in the mean RTV difference of the groups (P > 0.05). Conclusion: Within the limitations of this in vitro study, the titanium socket for the zirconia abutment did not show a significant effect on screw loosening under thermal stress compared to the titanium abutment in the internal connection implant.

참고문헌 (15)

  1. Yildirim M, "In vivo fracture resistance of implant-supported allceramic restorations", J Prosthet Dent, 90, pp.325~331 (2003)
  2. Apholt W, "Flexural strength of Cerec 2 machined and jointed In-Ceram-Alumina and InCeram-Zirconia bars", Dent Mater, 17, pp.260~267 (2001)
  3. Garvie RC, "Ceramic steel?", Nature, 258, pp.703~704 (1975)
  4. Adatia ND, "Fracture resistance of yttria-stabilized zirconia dental implant abutments", J Prosthodont, 18, pp.17~22 (2009)
  5. Sailer I, "In vitro study of the influence of the type of connection on the fracture load of zirconia abutments with internal and external implant-abutment connections", Int J Oral Maxillofac Implants, 24, pp.850~858 (2009)
  6. Chun HJ, "Fracture strength study of internally connected zirconia abutments reinforced with titanium inserts", Int J Oral Maxillofac Implants, 30, pp.346~350 (2015)
  7. Ekelund JA, "Implant treatment in the edentulous mandible: a prospective study on Brånemark system implants over more than 20 years", Int J Prosthodont, 16, pp.602~608 (2003)
  8. Eckert SE, "Analysis of incidence and associated factors with fractured implants:a retrospective study", Int J Oral Maxillofac Implants, 15, pp.662~667 (2000)
  9. Taylor TD, "Implant prosthodontics:current perspective and future directions", Int J Oral Maxillofac Implants, 15, pp.66~75 (2000)
  10. Winkler S, "Implant screw mechanics and the settling effect: overview", J Oral Implantol, 29, pp.242~245 (2003)
  11. Halliday D, "Fundamentals of physics extended", pp.586~591 (2008), Wiley
  12. McGlumphy EA, "Implant screw mechanics", Dent Clin North Am, 42, pp.71~89 (1998)
  13. Burguete RL, "Tightening characteristics for screwed joints in osseointegrated dental implants", J Prosthet Dent, 71, pp.592~599 (1994)
  14. Traini T, "In vitro evaluation of thermomechanic coupling in conical implant-to-abutment joint", Implant Dent, 16, pp.379~386 (2007)
  15. Low D, "Thermal expansion coefficient of titanium casting", J Oral Rehabil, 28, pp.239~242 (2001)

관련 논문AI추천

Removal torque of a two-piece zirconia abutment with a novel titanium component in an internal connection implant system after dynamic cyclic loading이주희대한치과보철학회 · 대한치과보철학회지2017 · 제55권 제2호Influence of internal connection length on screw loosening in internal connection implants김선재, 김성태, 김지선, 김현철, 문홍석, 박영범, 이재훈, 최현민대한치과보철학회 · 대한치과보철학회지2017 · 제55권 제3호Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connectionsElaheh Beyabanaki, Hakimeh Siadat, Marzieh Alikhasi, Niloufar Mousavi대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제4호Digital evaluation of axial displacement by implant-abutment connection type: an in vitro study김성준, 이규복, 손큰바다대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제5호Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading이주희, 차현석대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제6호The assessment of abutment screw stability between the external and internal hexagonal joint under cyclic loading김성훈, 양재호, 이재봉, 이태식, 한중석대한치과보철학회 · 대한치과보철학회지2008 · 제46권 제6호

이 간행물의 다른 논문 (12건)