Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

PURPOSE. This study investigated the effects of abutment screw lengths on screw loosening and removal torque in external connection implants after oblique cyclic loading. MATERIALS AND METHODS. External connection implants were secured with abutment screws to straight abutments. The abutment-implant assemblies were classified into seven groups based on the abutment screw length, with each group consisting of five assemblies. A cyclic load of 300 N was applied at a 30o angle to the loading axis until one million cycles were achieved. Removal torque values (RTVs) before and after loading, and RTV differences were evaluated. The measured values were analyzed using repeated measures of analysis of variance with the Student-Newman-Keuls multiple comparisons. RESULTS. All assemblies survived the oblique cyclic loading test without screw loosening. There was a significant decrease in the RTVs throughout the observed abutment screw lengths when the abutmentimplant assemblies were loaded repeatedly (P<.001). However, the abutment screw length did not show significant difference on the RTVs before and after the experiment when the abutment screw length ranged from 1.4 to 3.8 mm (P=.647). CONCLUSION. Within the limit of this experiment, our results indicate that the abutment screw length did not significantly affect RTV differences after oblique cyclic loading when a minimum length of 1.4 mm (3.5 threads) was engaged. These findings suggest that short abutment screws may yield stable clinical outcomes comparable to long screws in terms of load resistance.

Abstract

PURPOSE. This study investigated the effects of abutment screw lengths on screw loosening and removal torque in external connection implants after oblique cyclic loading. MATERIALS AND METHODS. External connection implants were secured with abutment screws to straight abutments. The abutment-implant assemblies were classified into seven groups based on the abutment screw length, with each group consisting of five assemblies. A cyclic load of 300 N was applied at a 30o angle to the loading axis until one million cycles were achieved. Removal torque values (RTVs) before and after loading, and RTV differences were evaluated. The measured values were analyzed using repeated measures of analysis of variance with the Student-Newman-Keuls multiple comparisons. RESULTS. All assemblies survived the oblique cyclic loading test without screw loosening. There was a significant decrease in the RTVs throughout the observed abutment screw lengths when the abutmentimplant assemblies were loaded repeatedly (P<.001). However, the abutment screw length did not show significant difference on the RTVs before and after the experiment when the abutment screw length ranged from 1.4 to 3.8 mm (P=.647). CONCLUSION. Within the limit of this experiment, our results indicate that the abutment screw length did not significantly affect RTV differences after oblique cyclic loading when a minimum length of 1.4 mm (3.5 threads) was engaged. These findings suggest that short abutment screws may yield stable clinical outcomes comparable to long screws in terms of load resistance.

참고문헌 (47)

  1. Laura Gaviria, "Current trends in dental implants", 대한구강악안면외과학회지, 40(2), pp.50~60 (2014), 대한구강악안면외과학회
  2. Chee W, "Failures in implant dentistry", Br Dent J, 202, pp.123~129 (2007)
  3. Di Iorio D, "Biomechanical aspects in late implant failures: scanning electron microscopy analysis of four clinical cases", J Contemp Dent Pract, 12, pp.356~360 (2011)
  4. Weber HP, "Does the type of implant prosthesis affect outcomes in the partially edentulous patient", Int J Oral Maxillofac Implants, 22, pp.140~172 (2007)
  5. Gracis S, "Internal vs. external connections for abutments", reconstructions: a systematic review, 2012, pp.23, 202~216
  6. Sailer I, "Cemented and screw-retained implant reconstructions: a systematic review of the survival and complication rates", Clin Oral Implants Res, 23, pp.163~201 (2012)
  7. Theoharidou A, "Abutment screw loosening in single-implant restorations: a systematic review", Int J Oral Maxillofac Implants, 23, pp.681~690 (2008)
  8. Vigolo P, "Internal- vs external-connection single implants: A retrospective study in an italian population treated by certified prosthodontists", Int J Oral Maxillofac Implants, 31, pp.1385~1396 (2016)
  9. Katsuta Y, "Abutment screw loosening of endosseous dental implant body", abutment joint by cyclic torsional loading test at the initial stage, 2015, pp.34, 896~902
  10. Sakamoto K, "Influence of eccentric cyclic loading on implant components: Comparison between external joint system and internal joint system", Dent Mater J, 35, pp.929~937 (2016)
  11. Khraisat A, "Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system", J Prosthet Dent, 91, pp.326~334 (2004)
  12. Tsuge T, "Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants", Dent Mater J, 28, pp.373~381 (2009)
  13. Gupta S, "Technical complications of implant-causes and management: A comprehensive review", Natl J Maxillofac Surg, 6, pp.3~8 (2015)
  14. do Nascimento C, "Leakage of saliva through the implant-abutment interface: in vitro evaluation of three different implant connections under unloaded and loaded conditions", Int J Oral Maxillofac Implants, 27, pp.551~560 (2012)
  15. Taylor TD, "Prosthodontic problems and limitations associated with osseointegration", J Prosthet Dent, 79, pp.74~78 (1998)
  16. Stüker RA, "Preload and torque removal evaluation of three different abutment screws for single standing implant restorations", J Appl Oral Sci, 16, pp.55~58 (2008)
  17. Winkler S, "Implant screw mechanics and the settling effect: overview", J Oral Implantol, 29, pp.242~245 (2003)
  18. Dhingra A, "Analysis of dimensional changes in the screw and the surface topography at the interface of a titanium screw and a zirconia abutment under cyclic loading: an in vitro study", Int J Oral Maxillofac Implants, 38, pp.661~669 (2013)
  19. Siamos G, "Relationship between implant preload and screw loosening on implant-supported prostheses", J Oral Implantol, 28, pp.67~73 (2002)
  20. Yeo IS, "The effect of abutment screw length on screw loosening in dental implants with external abutment connections after thermocycling", Int J Oral Maxillofac Implants, 29, pp.59~62 (2014)
  21. Kim BJ, "The effect of screw length on fracture load and abutment strain in dental implants with external abutment connections", Int J Oral Maxillofac Implants, 27, pp.820~823 (2012)
  22. "ISO 14801. Dentistry - Implants - Dynamic fatigue test for endosseous dental implants", International Organization for Standardization, ~ (2007)
  23. Khraisat A, "Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading", Clin Implant Dent Relat Res, 6, pp.157~164 (2004)
  24. Pjetursson BE, "A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years", Clin Oral Implants Res, 23, pp.22~38 (2012)
  25. Bianco G, "Osseointegrated implant for singletooth replacement: a retrospective multicenter study on routine use in private practice", Clin Implant Dent Relat Res, 2, pp.152~158 (2000)
  26. Jemt T, "A 3-year follow-up study on single implant treatment", J Dent, 21, pp.203~208 (1993)
  27. Att W, "Fracture resistance of single-tooth implant-supported all-ceramic restorations: an in vitro study", J Prosthet Dent, 95, pp.111~116 (2006)
  28. Hnd Hadi Mohammed, "Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants", The Journal of Advanced of Prosthodontics, 8(1), pp.62~69 (2016), 대한치과보철학회[DOI]
  29. Craig RG, "Restorative dental materials", pp.75~77 (1993), Mosby, St
  30. Dittmer MP, "Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading", J Prosthodont Res, 56, pp.19~24 (2012)
  31. Mericske-Stern R, "In vivo measurements of some functional aspects with mandibular fixed prostheses supported by implants", Clin Oral Implants Res, 7, pp.153~161 (1996)
  32. Alkan I, "Influence of occlusal forces on stress distribution in preloaded dental implant screws", J Prosthet Dent, 91, pp.319~325 (2004)
  33. 최진호, "치과용 임플랜트에서 지대주 나사의 길이 및 반복 조임 횟수가 지내주 나사의 풀림에 미치는 영향", 대한치과보철학회지, 44(4), pp.432~442 (2006), 대한치과보철학회
  34. Sopwith DG, "The distribution of load in screw threads", Proc Inst Mech Eng, 159, pp.373~383 (1948)
  35. Jemt T, "Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup", Int J Oral Maxillofac Implants, 6, pp.270~276 (1991)
  36. Jemt T, "Failures and complications in 127 consecutively placed fixed partial prostheses supported by Brånemark implants: from prosthetic treatment to first annual checkup", Int J Oral Maxillofac Implants, 7, pp.40~44 (1992)
  37. Luterbacher S, "Fractured prosthetic abutments in osseointegrated implants: a technical complication to cope with", Clin Oral Implants Res, 11, pp.163~170 (2000)
  38. 이유승, "Comparison of implant component fractures in external and internal type: A 12-year retrospective study", The Journal of Advanced of Prosthodontics, 10(2), pp.155~162 (2018), 대한치과보철학회[DOI]
  39. Binon PP, "The effect of implant", abutment hexagonal misfit on screw joint stability, 1996, pp.9, 149~160
  40. McGlumphy EA, "Implant screw mechanics", Dent Clin North Am, 42, pp.71~89 (1998)
  41. Dixon DL, "Comparison of screw loosening, rotation, and deflection among three implant designs", J Prosthet Dent, 74, pp.270~278 (1995)
  42. Coppedê AR, "Mechanical comparison of experimental conical-head abutment screws with conventional flat-head abutment screws for external-hex and internal tri-channel implant connections: an in vitro evaluation of loosening torque", Int J Oral Maxillofac Implants, 28, pp.e321~e329 (2013)
  43. 신현모, "Influence of the implant-abutment connection design and diameter on the screw joint stability", The Journal of Advanced of Prosthodontics, 6(2), pp.126~132 (2014), 대한치과보철학회[DOI]
  44. Feitosa PC, "Stability of external and internal implant connections after a fatigue test", Eur J Dent, 7, pp.267~271 (2013)
  45. Assunção WG, "Preload evaluation of different screws in external hexagon joint", Implant Dent, 21, pp.46~50 (2012)
  46. Maeda Y, "In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication", J Oral Rehabil, 33, pp.75~78 (2006)
  47. Stockton LW, "Factors affecting retention of post systems: a literature review", J Prosthet Dent, 81, pp.380~385 (1999)

관련 논문AI추천

이 간행물의 다른 논문 (5건)