Digital evaluation of axial displacement by implant-abutment connection type: an in vitro study

2

조회

0

다운로드

4

인용

0

참고문헌

키워드

초록

PURPOSE. To measure axial displacement of different implant–abutment connection types and materials during screw tightening at the recommended torque by using a contact scanner for two-dimensional (2D) and threedimensional (3D) analyses. MATERIALS AND METHODS. Twenty models of missing mandibular left second premolars were 3D-printed and implant fixtures were placed at the same position by using a surgical guide. External and internal fixtures were used. Three implant–abutment internal connection (INT) types and one implant–abutment external connection (EXT) type were prepared. Two of the INT types used titanium abutment and zirconia abutment; the other INT type was a customized abutment, fabricated by using a computercontrolled milling machine. The EXT type used titanium abutment. Screws were tightened at 10 N·cm, simulating hand tightening, and then at the manufacturers’ recommended torque (30 N·cm) 10 min later. Abutments and adjacent teeth were subsequently scanned with a contact scanner for 2D and 3D analyses using a 3D inspection software. RESULTS. Significant differences were observed in axial displacement according to the type of implant–abutment connection (P<.001). Vertical displacement of abutments was greater than overall displacement, and significant differences in vertical and overall displacement were observed among the four connection types (P<.05). CONCLUSION. Displacement according to connection type and material should be considered in choosing an implant abutment. When adjusting a prosthesis, tightening the screw at the manufacturers’ recommended torque is advisable, rather than the level of hand tightening.

Abstract

PURPOSE. To measure axial displacement of different implant–abutment connection types and materials during screw tightening at the recommended torque by using a contact scanner for two-dimensional (2D) and threedimensional (3D) analyses. MATERIALS AND METHODS. Twenty models of missing mandibular left second premolars were 3D-printed and implant fixtures were placed at the same position by using a surgical guide. External and internal fixtures were used. Three implant–abutment internal connection (INT) types and one implant–abutment external connection (EXT) type were prepared. Two of the INT types used titanium abutment and zirconia abutment; the other INT type was a customized abutment, fabricated by using a computercontrolled milling machine. The EXT type used titanium abutment. Screws were tightened at 10 N·cm, simulating hand tightening, and then at the manufacturers’ recommended torque (30 N·cm) 10 min later. Abutments and adjacent teeth were subsequently scanned with a contact scanner for 2D and 3D analyses using a 3D inspection software. RESULTS. Significant differences were observed in axial displacement according to the type of implant–abutment connection (P<.001). Vertical displacement of abutments was greater than overall displacement, and significant differences in vertical and overall displacement were observed among the four connection types (P<.05). CONCLUSION. Displacement according to connection type and material should be considered in choosing an implant abutment. When adjusting a prosthesis, tightening the screw at the manufacturers’ recommended torque is advisable, rather than the level of hand tightening.

참고문헌 (36)

  1. Branemark PI, "Intra-osseous anchorage of dental prostheses. I. Experimental studies", Scand J Plast Reconstr Surg, 3, pp.81~100 (1969)[DOI]
  2. Branemark PI, "Osseointegration and its experimental background", J Prosthet Dent, 50, pp.399~410 (1983)[DOI]
  3. Branemark PI, "Ten-year survival rates of fixed prostheses on four or six implants ad modum Branemark in full edentulism", Clin Oral Implants Res, 6, pp.227~231 (1995)[DOI]
  4. Sutter F, "The new restorative concept of the ITI dental implant system: design and engineering", Int J Periodontic Restor Dent, 13, pp.408~431 (1993)
  5. Ugurel CS, "Mechanical resistance of screwless morse taper and screw-retained implant-abutment connections", Clin Oral Implants Res, 26, pp.137~142 (2015)[DOI]
  6. Branemark PI, "Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period", Scand J Plast Reconstr Surg Suppl, 16, pp.1~132 (1977)
  7. Norton MR, "An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design", Clin Oral Implants Res, 8, pp.290~298 (1997)[DOI]
  8. Maeda Y, "In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication", J Oral Rehabil, 33, pp.75~78 (2006)[DOI]
  9. Schwarz MS, "Mechanical complications of dental implants", Clin Oral Implants Res, 11, pp.156~158 (2000)[DOI]
  10. Dailey B, "Axial displacement of abutments into implants and implant replicas, with the tapered cone-screw internal connection, as a function of tightening torque", Int J Oral Maxillofac Implants, 24, pp.251~256 (2009)
  11. Yilmaz B, "Displacement of screw-retained single crowns into implants with conical internal connections", Int J Oral Maxillofac Implants, 28, pp.803~806 (2013)[DOI]
  12. Winkler S, "Implant screw mechanics and the settling effect: overview", J Oral Implantol, 29, pp.242~245 (2003)[DOI]
  13. Lee JH, "Axial displacements in external and internal implant-abutment connection", Clin Oral Implants Res, 25, pp.e83~e89 (2014)[DOI]
  14. Spector MR, "An evaluation of impression techniques for osseointegrated implants", J Prosthet Dent, 63, pp.444~447 (1990)[DOI]
  15. Liu DY, "Accuracy of different implant impression techniques: Evaluation of new tray design concept", J Prosthodont (2017)
  16. Tan KB, "The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint", Int J Oral Maxillofac Implants, 17, pp.175~183 (2002)
  17. Dellinges MA, "A measurement of torque values obtained with hand-held drivers in a simulated clinical setting", J Prosthodont, 2, pp.212~214 (1993)[DOI]
  18. Kanawati A, "Measurement of clinicians' ability to hand torque dental implant components", J Oral Implantol, 35, pp.185~188 (2009)[DOI]
  19. 문승진, "Sinking and fit of abutment of locking taper implant system", The Journal of Advanced of Prosthodontics, 1(2), pp.97~101 (2009), 대한치과보철학회[DOI]
  20. Gilbert AB, "Three-Dimensional Displacement of Nine Different Abutments for an Implant with an Internal Hexagon Platform", Int J Oral Maxillofac Implants, 30, pp.781~788 (2015)[DOI]
  21. Hakimeh Siadat, "Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connections", The Journal of Advanced of Prosthodontics, 9(4), pp.271~277 (2017), 대한치과보철학회[DOI]
  22. Goodacre CJ, "Clinical complications of osseointegrated implants", J Prosthet Dent, 81, pp.537~552 (1999)[DOI]
  23. Tagger Green N, "Fracture of dental implants: literature review and report of a case", Implant Dent, 11, pp.137~143 (2002)[DOI]
  24. Kim KS, "Variation in the total lengths of abutment", implant assemblies generated with a function of applied tightening torque in external and internal implant-abutment connection, 2011, pp.22, 834~839
  25. 심혜원, "Long-term cumulative survival and mechanical complications of singletooth Ankylos Implants: focus on the abutment neck fractures", The Journal of Advanced of Prosthodontics, 7(6), pp.423~430 (2015), 대한치과보철학회[DOI]
  26. Messias A, "Effect of intentional abutment disconnection on the micromovements of the implant-abutment assembly: a 3D digital image correlation analysis", Clin Oral Implants Res, 28, pp.9~16 (2017)
  27. Rebeeah HA, "Comparison of 3D displacements of screw-retained zirconia implant crowns into implants with different internal connections with respect to screw tightening", J Prosthet Dent, 119, pp.132~137 (2018)[DOI]
  28. Schaefer O, "Marginal and internal fit of pressed lithium disilicate partial crowns in vitro: a three-dimensional analysis of accuracy and reproducibility", Dent Mater, 28, pp.320~326 (2012)[DOI]
  29. Rudolph H, "Computer-aided analysis of the influence of digitizing and surfacing on the accuracy in dental CAD", CAM technology, 2007, pp.37, 579~587
  30. 이예규, "Comparison of intraoral scanning and conventional impression techniques using 3-dimensional superimposition", The Journal of Advanced of Prosthodontics, 7(6), pp.460~467 (2015), 대한치과보철학회[DOI]
  31. 박지만, "Comparative analysis on reproducibility among 5 intraoral scanners: sectional analysis according to restoration type and preparation outline form", The Journal of Advanced of Prosthodontics, 8(5), pp.354~362 (2016), 대한치과보철학회[DOI]
  32. Hye-Nan Park, "A comparison of the accuracy of intraoral scanners using an intraoral environment simulator", The Journal of Advanced of Prosthodontics, 10(1), pp.58~64 (2018), 대한치과보철학회[DOI]
  33. Hayama H, "Trueness and precision of digital impressions obtained using an intra-oral scanner with different head size in the partially edentulous mandible", J Prosthodont Res, 62, pp.347~352 (2018)[DOI]
  34. Patzelt SB, "Accuracy of full-arch scans using intraoral scanners", Clin Oral Investig, 18, pp.1687~1694 (2014)[DOI]
  35. Persson A, "A three-dimensional evaluation of a laser scanner and a touchprobe scanner", J Prosthet Dent, 95, pp.194~200 (2006)[DOI]
  36. Dimitrova M, "A 3D evaluation of the repeatability of accuracy in optical and contact scanners", pp.2017

관련 논문AI추천

Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connectionsElaheh Beyabanaki, Hakimeh Siadat, Marzieh Alikhasi, Niloufar Mousavi대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제4호Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force곽재영, 김성균, 허성주, Yuseung Yi대한치과보철학회 · The Journal of Advanced of Prosthodontics2019 · 제11권 제6호Effect of cyclic loading on axial displacement of abutment into implant with internal tapered connection: a pilot study곽재영, 김성균, 설현우, 한종현, 허성주대한치과보철학회 · 대한치과보철학회지2013 · 제51권 제4호Influence of the implant-abutment connection design and diameter on the screw joint stability신현모, 윤미정, 전영찬, 정창모, 허중보, Brian Myung Chang대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제2호Influence of the implant abutment types and the dynamic loading on initial screw loosening김은숙, 신수연대한치과보철학회 · The Journal of Advanced of Prosthodontics2013 · 제5권 제1호Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connectionGiuseppe Troiano, Khrystyna Zhurakivska, Luigi Laino, Rosario Prisco대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제6호

이 간행물의 다른 논문 (8건)