Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connections
2
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
PURPOSE. This in vitro study aimed to evaluate the effect of implant connection design (external vs. internal) on the fit discrepancy and torque loss of zirconia and titanium abutments. MATERIALS AND METHODS. Two regular platform dental implants, one with external connection (Brånemark, Nobel Biocare AB) and the other with internal connection (Noble Replace, Nobel Biocare AB), were selected. Seven titanium and seven customized zirconia abutments were used for each connection design. Measurements of geometry, marginal discrepancy, and rotational freedom were done using video measuring machine. To measure the torque loss, each abutment was torqued to 35 Ncm and then opened by means of a digital torque wrench. Data were analyzed with two-way ANOVA and t-test at α=0.05 of significance. RESULTS. There were significant differences in the geometrical measurements and rotational freedom between abutments of two connection groups (P<.001). Also, the results showed significant differences between titanium abutments of internal and external connection implants in terms of rotational freedom (P<.001). Not only customized internal abutments but also customized external abutments did not have the exact geometry of prefabricated abutments (P<.001). However, neither connection type (P=.15) nor abutment material (P=.38) affected torque loss. CONCLUSION. Abutments with internal connection showed less rotational freedom. However, better marginal fit was observed in externally connected abutments. Also, customized abutments with either connection could not duplicate the exact geometry of their corresponding prefabricated abutment. However, neither abutment connection nor material affected torque loss values.
Abstract
PURPOSE. This in vitro study aimed to evaluate the effect of implant connection design (external vs. internal) on the fit discrepancy and torque loss of zirconia and titanium abutments. MATERIALS AND METHODS. Two regular platform dental implants, one with external connection (Brånemark, Nobel Biocare AB) and the other with internal connection (Noble Replace, Nobel Biocare AB), were selected. Seven titanium and seven customized zirconia abutments were used for each connection design. Measurements of geometry, marginal discrepancy, and rotational freedom were done using video measuring machine. To measure the torque loss, each abutment was torqued to 35 Ncm and then opened by means of a digital torque wrench. Data were analyzed with two-way ANOVA and t-test at α=0.05 of significance. RESULTS. There were significant differences in the geometrical measurements and rotational freedom between abutments of two connection groups (P<.001). Also, the results showed significant differences between titanium abutments of internal and external connection implants in terms of rotational freedom (P<.001). Not only customized internal abutments but also customized external abutments did not have the exact geometry of prefabricated abutments (P<.001). However, neither connection type (P=.15) nor abutment material (P=.38) affected torque loss. CONCLUSION. Abutments with internal connection showed less rotational freedom. However, better marginal fit was observed in externally connected abutments. Also, customized abutments with either connection could not duplicate the exact geometry of their corresponding prefabricated abutment. However, neither abutment connection nor material affected torque loss values.
참고문헌 (46)
- Vigolo P, "An in vitro evaluation of titanium, zirconia, and alumina procera abutments with hexagonal connection", Int J Oral Maxillofac Implants, 21, pp.575~580 (2006)
- Vigolo P, "An in vitro evaluation of ZiReal abutments with hexagonal connection: in original state and following abutment preparation", Int J Oral Maxillofac Implants, 20, pp.108~114 (2005)
- Park JI, "Comparison of fracture resistance and fit accuracy of customized zirconia abutments with prefabricated zirconia abutments in internal hexagonal implants", Clin Implant Dent Relat Res, 15, pp.769~778 (2013)
- Quinn JB, "A practical and systematic review of Weibull statistics for reporting strengths of dental materials", Dent Mater, 26, pp.135~147 (2010)
- Butz F, "Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation", J Oral Rehabil, 32, pp.838~843 (2005)
- Andersson B, "Ceramic implant abutments for short-span FPDs: a prospective 5-year multicenter study", Int J Prosthodont, 16, pp.640~646 (2003)
- Andersson B, "Alumina ceramic implant abutments used for single-tooth replacement: a prospective 1- to 3-year multicenter study", Int J Prosthodont, 14, pp.432~438 (2001)
- Leutert CR, "Bending moments and types of failure of zirconia and titanium abutments with internal implant-abutment connections:a laboratory study", Int J Oral Maxillofac Implants, 27, pp.505~512 (2012)
- Almeida EO, "Mechanical testing of implant-supported anterior crowns with different implant", abutment connections, 2013, pp.28, 103~108
- Gehrke SA, "Influence of bone insertion level of the implant on the fracture strength of different connection designs: an in vitro study", Clin Oral Investig, 18, pp.715~720 (2014)
- Lin MI, "A retrospective study of implant-abutment connections on crestal bone level", J Dent Res, 92, pp.202S~207S (2013)
- Peñarrocha-Diago MA, "Influence of implant neck design and implant-abutment connection type on peri-implant health. Radiological study", Clin Oral Implants Res, 24, pp.1192~1200 (2013)
- Asvanund P, "Photoelastic stress analysis of external versus internal implant-abutment connections", J Prosthet Dent, 106, pp.266~271 (2011)
- Gracis S, "Internal vs. external connections for abutments", reconstructions: a systematic review, 2012, pp.23, 202~216
- Asvanund P, "A strain gauge analysis comparing external and internal implant-abutment connections", Implant Dent, 23, pp.206~211 (2014)
- Davi LR, "In vitro integrity of implant external hexagon after application of surgical placement torque simulating implant locking", Braz Oral Res, 22, pp.125~131 (2008)
- Khraisat A, "Fatigue resistance of two implant", abutment joint designs, 2002, pp.88, 604~610
- Norton MR, "An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design", Clin Oral Implants Res, 8, pp.290~298 (1997)
- Möllersten L, "Comparison of strength and failure mode of seven implant systems: an in vitro test", J Prosthet Dent, 78, pp.582~591 (1997)
- Adatia ND, "Fracture resistance of yttria-stabilized zirconia dental implant abutments", J Prosthodont, 18, pp.17~22 (2009)
- Kelly JR, "Developing meaningful systematic review of CAD", CAM reconstructions and fiber-reinforced composites, 2007, pp.18, 205~217
- Strub JR, "Computer-aided design and fabrication of dental restorations: current systems and future possibilities", J Am Dent Assoc, 137, pp.1289~1296 (2006)
- Fuster-Torres MA, "CAD", CAM dental systems in implant dentistry: update, 2009, pp.14, 141~
- Beuer F, "Precision of fit: zirconia three-unit fixed dental prostheses", Clin Oral Investig, 13, pp.343~349 (2009)
- Kapos T, "Computer-aided design and computer-assisted manufacturing in prosthetic implant dentistry", Int J Oral Maxillofac Implants, 24, pp.110~117 (2009)
- Karataşli O, "Comparison of the marginal fit of different coping materials and designs produced by computer aided manufacturing systems", Dent Mater J, 30, pp.97~102 (2011)
- Alikhasi M, "A comparison of precision of fit, rotational freedom, and torque loss with copy-milled zirconia and prefabricated titanium abutments", Int J Oral Maxillofac Implants, 28, pp.996~1002 (2013)
- Lang LA, "The influence of abutment screw tightening on screw joint configuration", J Prosthet Dent, 87, pp.74~79 (2002)
- Att W, "Marginal adaptation of all-ceramic crowns on implant abutments", Clin Implant Dent Relat Res, 10, pp.218~225 (2008)
- Abrahamsson I, "The mucosal attachment at different abutments. An experimental study in dogs", J Clin Periodontol, 25, pp.721~727 (1998)
- Baixe S, "Microgap between zirconia abutments and titanium implants", Int J Oral Maxillofac Implants, 25, pp.455~460 (2010)
- Weinberg LA, "A comparison of implant", prosthesis loading with four clinical variables, 1995, pp.8, 421~433
- Hecker DM, "Cyclic loading of implantsupported prostheses: comparison of gaps at the prostheticabutment interface when cycled abutments are replaced with as-manufactured abutments", J Prosthet Dent, 95, pp.26~32 (2006)
- Binon PP, "The effect of implant", abutment hexagonal misfit on screw joint stability, 1996, pp.9, 149~160
- Binon PP, "The effect of eliminating implant", abutment rotational misfit on screw joint stability, 1996, pp.9, 511~519
- Jörnéus L, "Loads and designs of screw joints for single crowns supported by osseointegrated implants", Int J Oral Maxillofac Implants, 7, pp.353~359 (1992)
- Kano SC, "The effect of casting procedures on rotational misfit in castable abutments", Int J Oral Maxillofac Implants, 22, pp.575~579 (2007)
- Vigolo P, "Measurement of the dimensions and abutment rotational freedom of gold-machined 3i UCLA-type abutments in the as-received condition, after casting with a noble metal alloy and porcelain firing", J Prosthet Dent, 84, pp.548~553 (2000)
- Binon PP, "Evaluation of machining accuracy and consistency of selected implants, standard abutments, and laboratory analogs", Int J Prosthodont, 8, pp.162~178 (1995)
- Binon PP, "Evaluation of three slip fit hexagonal implants", Implant Dent, 5, pp.235~248 (1996)
- de Barros Carrilho GP, "Comparison of external and internal hex implants’ rotational freedom: a pilot study", Int J Prosthodont, 18, pp.165~166 (2005)
- Kano SC, "Effect of casting procedures on screw loosening in UCLA-type abutments", J Prosthodont, 15, pp.77~81 (2006)
- Barbosa GA, "Relation between implant", abutment vertical misfit and torque loss of abutment screws, 2008, pp.19, 358~363
- Yüzügüllü B, "The implant-abutment interface of alumina and zirconia abutments", Clin Implant Dent Relat Res, 10, pp.113~121 (2008)
- Kerstein RB, "A comparison of fabrication precision and mechanical reliability of 2 zirconia implant abutments", Int J Oral Maxillofac Implants, 23, pp.1029~1036 (2008)
- Garine WN, "Measurement of the rotational misfit and implant-abutment gap of all-ceramic abutments", Int J Oral Maxillofac Implants, 22, pp.928~938 (2007)