Removal torque of a two-piece zirconia abutment with a novel titanium component in an internal connection implant system after dynamic cyclic loading

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Purpose: The aim of this study was to evaluate screw removal torque of the two-piece zirconia abutment with the novel titanium component compared to the conventional onepiece titanium abutment in the internal connection implant before and after dynamic cyclic loading. Materials and methods: Two types of the abutment assemblies with internal connection were prepared and divided into the groups; titanium abutment-titanium abutment screw assemblies as control, and zirconia abutment-titanium socket-titanium abutment screw assemblies as experimental group. A total of 12 abutments and implants were used of six assemblies each group. Each assembly was tightened to 30 Ncm. A cyclic load of 300 N at an angle of 30 degrees in reference to the loading axis was applied until one million cycles or failure. The removal torque values (RTVs) of the abutment screws were measured with a digital torque gauge before and after cyclic loading. The RTVs of the pre load and post load were analyzed with t-test, and P-values < .05 were considered statistically significant. Results: The assemblies of both groups survived all after the dynamic cyclic loading test without screw loosening. The statistically significant differences were found between the mean RTVs before and after the cyclic loading in both groups (P < .05). The RTV differences for the control and the experimental group were -7.25 ± 1.50 Ncm and -7.33 ± 0.93 Ncm, respectively. Statistical analysis revealed that the RTV differences in both groups were not significantly different from each other (P > .05). Conclusion: Within the limitation of this study, the two-piece zirconia abutment with the titanium component did not show a significant RTV difference of the abutment screw compared to the titanium abutment after dynamic cyclic loading.

Abstract

Purpose: The aim of this study was to evaluate screw removal torque of the two-piece zirconia abutment with the novel titanium component compared to the conventional onepiece titanium abutment in the internal connection implant before and after dynamic cyclic loading. Materials and methods: Two types of the abutment assemblies with internal connection were prepared and divided into the groups; titanium abutment-titanium abutment screw assemblies as control, and zirconia abutment-titanium socket-titanium abutment screw assemblies as experimental group. A total of 12 abutments and implants were used of six assemblies each group. Each assembly was tightened to 30 Ncm. A cyclic load of 300 N at an angle of 30 degrees in reference to the loading axis was applied until one million cycles or failure. The removal torque values (RTVs) of the abutment screws were measured with a digital torque gauge before and after cyclic loading. The RTVs of the pre load and post load were analyzed with t-test, and P-values < .05 were considered statistically significant. Results: The assemblies of both groups survived all after the dynamic cyclic loading test without screw loosening. The statistically significant differences were found between the mean RTVs before and after the cyclic loading in both groups (P < .05). The RTV differences for the control and the experimental group were -7.25 ± 1.50 Ncm and -7.33 ± 0.93 Ncm, respectively. Statistical analysis revealed that the RTV differences in both groups were not significantly different from each other (P > .05). Conclusion: Within the limitation of this study, the two-piece zirconia abutment with the titanium component did not show a significant RTV difference of the abutment screw compared to the titanium abutment after dynamic cyclic loading.

참고문헌 (23)

  1. Prestipino V, "Esthetic high-strength implant abutments. Part I", J Esthet Dent, 5, pp.29~36 (1993)
  2. Manicone PF, "An overview of zirconia ceramics: basic properties and clinical applications", J Dent, 35, pp.819~826 (2007)
  3. Yildirim M, "In vivo fracture resistance of implant-supported all-ceramic restorations", J Prosthet Dent, 90, pp.325~331 (2003)
  4. Apholt W, "Flexural strength of Cerec 2 machined and jointed InCeram-Alumina and InCeram-Zirconia bars", Dent Mater, 17, pp.260~267 (2001)
  5. Adatia ND, "Fracture resistance of yttria-stabilized zirconia dental implant abutments", J Prosthodont, 18, pp.17~22 (2009)
  6. Sailer I, "In vitro study of the influence of the type of connection on the fracture load of zirconia abutments with internal and external implant-abutment connections", Int J Oral Maxillofac Implants, 24, pp.850~858 (2009)
  7. Gehrke P, "Retentive strength of two-piece CAD", CAM zirconia implant abutments, 2014, pp.16, 920~925
  8. Chun HJ, "Fracture strength study of internally connected zirconia abutments reinforced with titanium inserts", Int J Oral Maxillofac Implants, 30, pp.346~350 (2015)
  9. Theoharidou A, "Abutment screw loosening in single-implant restorations: a systematic review", Int J Oral Maxillofac Implants, 23, pp.681~690 (2008)
  10. Taylor TD, "Prosthodontic problems and limitations associated with osseointegration", J Prosthet Dent, 79, pp.74~78 (1998)
  11. Stüker RA, "Preload and torque removal evaluation of three different abutment screws for single standing implant restorations", J Appl Oral Sci, 16, pp.55~58 (2008)
  12. Dhingra A, "Analysis of dimensional changes in the screw and the surface topography at the interface of a titanium screw and a zirconia abutment under cyclic loading:an in vitro study", Int J Oral Maxillofac Implants, 28, pp.661~669 (2013)
  13. Siamos G, "Relationship between implant preload and screw loosening on implant-supported prostheses", J Oral Implantol, 28, pp.67~73 (2002)
  14. Tsuge T, "Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants", Dent Mater J, 28, pp.373~381 (2009)
  15. Craig RG, "Restorative dental materials", pp.75~77 (1993), Mosby
  16. Att W, "Fracture resistance of singletooth implant-supported all-ceramic restorations: an in vitro study", J Prosthet Dent, 95, pp.111~116 (2006)
  17. Dittmer MP, "Influence of the interface design on the yield force of the implantabutment complex before and after cyclic mechanical loading", J Prosthodont Res, 56, pp.19~24 (2012)
  18. Khraisat A, "Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading", Clin Implant Dent Relat Res, 6, pp.157~164 (2004)
  19. Khraisat A, "Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system", J Prosthet Dent, 91, pp.326~334 (2004)
  20. Gracis S, "Internal vs. external connections for abutments", reconstructions:a systematic review, 2012, pp.23, 202~216
  21. Delben JA, "Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems", Dent Mater, 30, pp.408~416 (2014)
  22. Kim JS, "In vitro assessment of three types of zirconia implant abutments under static load", J Prosthet Dent, 109, pp.255~263 (2013)
  23. Dixon DL, "Comparison of screw loosening, rotation, and deflection among three implant designs", J Prosthet Dent, 74, pp.270~278 (1995)

관련 논문AI추천

The effect of a titanium socket with a zirconia abutment on screw loosening after thermocycling in an internally connected implant: a preliminary study경규영, 이주희, 차현석대한턱관절교합학회 · 구강회복응용과학지2017 · 제33권 제2호The assessment of abutment screw stability between the external and internal hexagonal joint under cyclic loading김성훈, 양재호, 이재봉, 이태식, 한중석대한치과보철학회 · 대한치과보철학회지2008 · 제46권 제6호Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading이주희, 차현석대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제6호Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants배지명, 이진한, 조혜원, Hnd Hadi Mohammed대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제1호Influence of internal connection length on screw loosening in internal connection implants김선재, 김성태, 김지선, 김현철, 문홍석, 박영범, 이재훈, 최현민대한치과보철학회 · 대한치과보철학회지2017 · 제55권 제3호Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connectionsElaheh Beyabanaki, Hakimeh Siadat, Marzieh Alikhasi, Niloufar Mousavi대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제4호

이 간행물의 다른 논문 (12건)