Development of a device to improve the precision of water surface identification for MeV electron beam dosimetry
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
The study aimed to develop a laser-based distance meter (LDM) to improve water surface identification for clinical MeV electron beam dosimetry, as inaccurate water surface determination can lead to imprecise positioning of ionization chambers (ICs). The LDM consisted of a laser ranging sensor, a signal processing microcontroller, and a tablet PC for data acquisition. I50 (the water depth at which ionization current drops to 50 % of its maximum) measurements of electron beams were performed using six different types of ICs and compared to other water surface identification methods. The LDM demonstrated reproducible I50 measurements with a level of 0.01 cm for all six ICs. The uncertainty of water depth was evaluated at 0.008 cm with the LDM. The LDM also exposed discrepancies between I50 measurements using different ICs, which was partially reduced by applying an optimum shift of IC’s point of measurement (POM) or effective point of measurement (EPOM). However, residual discrepancies due to the energy dependency of the cylindrical chamber’s EPOM caused remained. The LDM offers straightforward and efficient means for precision water surface identification, minimizing reliance on individual operator skills.
Abstract
The study aimed to develop a laser-based distance meter (LDM) to improve water surface identification for clinical MeV electron beam dosimetry, as inaccurate water surface determination can lead to imprecise positioning of ionization chambers (ICs). The LDM consisted of a laser ranging sensor, a signal processing microcontroller, and a tablet PC for data acquisition. I50 (the water depth at which ionization current drops to 50 % of its maximum) measurements of electron beams were performed using six different types of ICs and compared to other water surface identification methods. The LDM demonstrated reproducible I50 measurements with a level of 0.01 cm for all six ICs. The uncertainty of water depth was evaluated at 0.008 cm with the LDM. The LDM also exposed discrepancies between I50 measurements using different ICs, which was partially reduced by applying an optimum shift of IC’s point of measurement (POM) or effective point of measurement (EPOM). However, residual discrepancies due to the energy dependency of the cylindrical chamber’s EPOM caused remained. The LDM offers straightforward and efficient means for precision water surface identification, minimizing reliance on individual operator skills.
참고문헌 (19)
- B. R. Muir, "Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types", Med. Phys, 41(11) (2014)
- D. W. O. Rogers, "A new approach to electron-beam reference dosimetry", Med. Phys, 25(3), pp.310~320 (1998)
- P. R. Almond, "AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams", Med. Phys, 26(9), pp.1847~1870 (1999)
- International Atomic Energy Agency, "AEA TRS 398:Absorbed Dose Determination in External Beam Radiotherapy: an International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, vol. 12", ~ (2000)
- B. R. Muir, "Determination of relative ion chamber calibration coefficients from depth-ionization measurements in clinical electron beams", Phys. Med. Biol, 59(19), pp.5953~5969 (2014)
- M. McEwen, "Addendum to the AAPM’s TG-51 protocol for clinical reference dosimetry of highenergy photon beams", Med. Phys, 41(4), pp.1~20 (2014)
- J. M. Osinga-Bl¨attermann, "Direct determination of kQ for Farmer-type ionization chambers in a clinical scanned carbon ion beam using water calorimetry", Phys. Med. Biol, 62(6), pp.2033~2054 (2017)
- "2015", 42(11), pp.10.1118, 1.4931970
- D. W. O. Rogers, "Clinical Dosimetry Measurements in Radiotherapy, Medical Physics Monograph No 34", pp.239~298 (2009)
- R. C. Tailor, "Depth of ionization chamber in water", Phys. Med. Biol, 40(8), pp.1389~1392 (1995)
- P. von Voigts-Rhetz, "SU-F-T-72 : experimental determination of the position uncertainties for ROOS ionization chambers in clinical electron beams", Med. Phys, 43, pp.3478~3478 (2016)
- J. V. Siebers, "Quantitative ionization chamber alignment to a water surface : theory and simulation", Med. Phys, 44(7), pp.3794~3804 (2017)
- J. D. Ververs, "Quantitative ionization chamber alignment to a water surface : performance of multiple chambers", Med. Phys, 44(7), pp.3839~3847 (2017)
- DIN Deutsches Institut für Normung, "Procedures of Dosimetry with Probe-type Detectors for Photon and Electron Radiation – Part 2: Ionization Chamber Dosimetry of High Energy Photon and Electron Radiation", ~ (2020)
- "Baumer, Datasheet of Laser Sensor OM-30-P0350" (2021)[URL]
- "Effective point of measurement for parallel plate chambers using PTW TRUFIX" (2021)[URL]
- "2014", 24(3), pp.10.1016, j.zemedi.2013.12.001
- I. J. Das, "Accelerator beam data commissioning equipment and procedures : report of the TG-106 of the Therapy Physics Committe of the AAPM", Med. Phys, 35(9), pp.4186~4215 (2008)
- M. R. McEwen, "The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams", Med. Phys, 35(3), pp.950~958 (2008)