본문 바로가기
KERD
논문KCI 등재

Uncertainty evaluation of absorbed dose measurement based on temperature rise using a water calorimeter for 60Co gamma rays

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

A water calorimeter has been used to establish the measurement standard of absorbed dose to water for 60Co gamma-ray beams. While heat defect and heat conduction have been thoroughly investigated in previous studies, the uncertainty in temperature rise measurement has not received the same level of attention. This work reevaluated that component, identifying the dominant contributions from type A uncertainty due to repeated irradiation experiments (0.10 %), sensitivity calibration of the bridge circuit (0.075 %), and the extrapolation procedure used to determine signal change (0.07 %). Other sources, such as thermistor calibration and resistance measurements, contributed 0.047 %. As a result, the standard uncertainty of the temperature rise measurement was estimated to be 0.15 %, leading to a combined standard uncertainty of 0.26 % for the absorbed dose to water.

Abstract

A water calorimeter has been used to establish the measurement standard of absorbed dose to water for 60Co gamma-ray beams. While heat defect and heat conduction have been thoroughly investigated in previous studies, the uncertainty in temperature rise measurement has not received the same level of attention. This work reevaluated that component, identifying the dominant contributions from type A uncertainty due to repeated irradiation experiments (0.10 %), sensitivity calibration of the bridge circuit (0.075 %), and the extrapolation procedure used to determine signal change (0.07 %). Other sources, such as thermistor calibration and resistance measurements, contributed 0.047 %. As a result, the standard uncertainty of the temperature rise measurement was estimated to be 0.15 %, leading to a combined standard uncertainty of 0.26 % for the absorbed dose to water.

관련 논문AI추천

Comparison of the standards for absorbed dose to water of the IAEA and the KRISS, Korea in accelerator photon beams김윤호, 김인중, 박종인, 이철영, Czap L., Msimang Z.한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제7호국제표준규격의 137Cs 감마선장에서 전리함선량계 종류에 따른 공기커마율 측정 및 평가결과의 유효성 검증안현, 강영록, 계용욱, 김효진, 목성윤, 백정화, 이지은, 이창근, 장희진, 조월순(사)한국방사선산업학회 · 방사선산업학회지2025 · 제19권 제3호Development of a device to improve the precision of water surface identification for MeV electron beam dosimetry김인중, 박종인, Okky Agassy F.한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제4호Evaluating the Kerma coefficients relevant to various water-equivalent materials at different photon energiesAbdi Zahra, Baghani Hamid Reza, Shamsabadi Reza한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제2호Monte Carlo simulation of neutron interaction in Mn bath used for primary standard neutron dosimetryAbdelsalam A., Bakr Reham Hamdy, Elsaied Mohamed A., El-Sersy A.R., Mostafa Khaled, Osman W.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제7호Monte Carlo-based identification of electron and proton edges for calibration of miniaturized tissue equivalent proportional counter엄민기, 예성준, 윤석원한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제11호

이 간행물의 다른 논문 (59건)