Performance Evaluation of Stealth Chamber as a Novel Reference Chamber for Measuring Percentage Depth Dose and Profile of VitalBeam Linear Accelerator

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The purpose of this study is to evaluate the performance of a “stealth chamber” as a novel reference chamber for measuring percentage depth dose (PDD) and profile of 6, 8 and 10 MV photon energies. The PDD curves and dose profiles with fields ranging from 3 × 3 to 25 × 25 cm2 were acquired from measurements by using the stealth chamber and CC 13 chamber as reference chamber. All measurements were performed with Varian VitalBeam linear accelerator. In order to assess the performance of stealth chamber, PDD curves and profiles measured with stealth chamber were compared with measurement data using CC13 chamber. For PPDs measured with both chambers, the dosimetric parameters such as dmax (depth of maximum dose), D50 (PDD at 50 ㎜ depth), and D100 (PDD at 100 ㎜ depth) were analyzed. Moreover, root mean square error (RMSE) values for profiles at dmax and 100 ㎜ depth were evaluated. The measured PDDs and profiles between the stealth chamber and CC13 chamber as reference detector had almost comparable. For PDDs, the evaluated dosimetric parameters were observed small difference (<1%) for all energies and field sizes, except for dmax less than 2 ㎜. In addition, the difference of RMSEs for profiles at dmax and 100 ㎜ depth was similar for both chambers. This study confirmed that the use of stealth chamber for measuring commission beam data is a feasible as reference chamber for fields ranging from 3 × 3 to 20 × 20 cm2. Furthermore, it has an advantage with respect to measurement of the small fields (less than 3 × 3 cm2 field) although not performed in this study.

Abstract

The purpose of this study is to evaluate the performance of a “stealth chamber” as a novel reference chamber for measuring percentage depth dose (PDD) and profile of 6, 8 and 10 MV photon energies. The PDD curves and dose profiles with fields ranging from 3 × 3 to 25 × 25 cm2 were acquired from measurements by using the stealth chamber and CC 13 chamber as reference chamber. All measurements were performed with Varian VitalBeam linear accelerator. In order to assess the performance of stealth chamber, PDD curves and profiles measured with stealth chamber were compared with measurement data using CC13 chamber. For PPDs measured with both chambers, the dosimetric parameters such as dmax (depth of maximum dose), D50 (PDD at 50 ㎜ depth), and D100 (PDD at 100 ㎜ depth) were analyzed. Moreover, root mean square error (RMSE) values for profiles at dmax and 100 ㎜ depth were evaluated. The measured PDDs and profiles between the stealth chamber and CC13 chamber as reference detector had almost comparable. For PDDs, the evaluated dosimetric parameters were observed small difference (<1%) for all energies and field sizes, except for dmax less than 2 ㎜. In addition, the difference of RMSEs for profiles at dmax and 100 ㎜ depth was similar for both chambers. This study confirmed that the use of stealth chamber for measuring commission beam data is a feasible as reference chamber for fields ranging from 3 × 3 to 20 × 20 cm2. Furthermore, it has an advantage with respect to measurement of the small fields (less than 3 × 3 cm2 field) although not performed in this study.

참고문헌 (11)

  1. IJ Das, "Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM", Med PhyS., 35(9), pp.4186~4215 (2008)[DOI]
  2. J Van Dyk, "Commissioning and quality assurance of treatment planning computers", Int J Radiat Oncol Biol Phys, 26(2), pp.261~273 (1993)[DOI]
  3. R Palta, "Linear Accelerator Acceptance Testing and Commissioning", Med Phys, 30(6), pp.1356~1367 (2003)
  4. B Fraass, "American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning", Med Phys, 25(10), pp.1773~1829 (1998)[DOI]
  5. Stereotactic beam characterization using the IBA stealth chamber reference detector. IBA Dosimetry Whitepaper, "J Gersh", ~ (2014)
  6. J Wuerfel, "Dose measurements in small fields", Med Phys Int J, 1(1), pp.81~90 (2013)
  7. GX Ding, "Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods", Phys Med Biol, 51(10), pp.2549~ (2006)[DOI]
  8. LAV Quino, "Evaluation of a novel reference chabmer "stealth chamber" through Monte Carlo simulations and experimental data", Int J Cancer Ther Oncol, 3(2), pp.3222~ (2015)[DOI]
  9. 구기래, "선형가속기의 소조사면에 대한 빔 자료 측정에서 검출기의 선량 특성 분석", 방사선기술과학, 35(3), pp.265~274 (2012), 대한방사선과학회
  10. Quino LAV, "Clinical experience with a novel reference chamber "stealth chamber" by IBA", AIP Conf. Proc, 1747, pp.040006~ (2016)
  11. Azcona JD, "Commissioning of small field size radiosurgery cones in a 6-MV flattening filter-free beam", Med Dosim, 42, pp.282~288 (2017)[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (9건)