본문 바로가기
KERD
논문KCI 등재

Ferucarbotran-Enhanced Hepatic MRI at 3T Unit: Quantitative and Qualitative Comparison of Fast Breathhold Imaging Sequences

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Purpose : To compare the relative values of various fast breath-hold imaging sequences for superparamagnetic iron-oxide (SPIO)-enhanced hepatic MRI for the assessment of solid focal lesions with a 3T MRI unit. Materials and Methods : 102 consecutive patients with one or more solid malignant hepatic lesions were evaluated by spoiled gradient echo (GRE) sequences with three different echo times (2.4 msec [GRE_2.4], 5.8 msec [GRE_5.8], and 10 msec [GRE_10]) for T2*-weighted imaging in addition to T2-weighted turbo spin echo (TSE) sequence following intravenous SPIO injection. Image qualities of the hepatic contour, vascular landmarks and artifacts were rated by two independent readers using a four-point scale. For quantitative analysis, contrast-to-noise ratio (CNR) was measured in 170 solid focal lesions larger than 1 cm (107 hepatocellular carcinomas, nine cholangiocarcinomas and 54 metastases). Results : GRE_5.8 showed the highest mean points for hepatic contour, vascular anatomy and imaging artifact presence among all of the subjected sequences (p<0.001) and was comparable (p=0.414) with GRE_10 with regard to lesion conspicuity. The mean CNRs were significantly higher (p<0.001) in the following order: GRE_10 (24.4±14.5), GRE_5.8 (14.8±9.4), TSE (9.7±6.3), and GRE_2.4 (7.9±6.4). The mean CNRs of CCCs and metastases were higher than those of HCCs for all imaging sequences (p<0.05). Conclusion : Regarding overall performances, GRE using a moderate echo time of 5.8 msec can provide the most reliable data among the various fast breath-hold SPIOenhanced hepatic MRI sequences at 3T unit despite the lower CNR of GRE_5.8 compared to that of GRE_10.

Abstract

Purpose : To compare the relative values of various fast breath-hold imaging sequences for superparamagnetic iron-oxide (SPIO)-enhanced hepatic MRI for the assessment of solid focal lesions with a 3T MRI unit. Materials and Methods : 102 consecutive patients with one or more solid malignant hepatic lesions were evaluated by spoiled gradient echo (GRE) sequences with three different echo times (2.4 msec [GRE_2.4], 5.8 msec [GRE_5.8], and 10 msec [GRE_10]) for T2*-weighted imaging in addition to T2-weighted turbo spin echo (TSE) sequence following intravenous SPIO injection. Image qualities of the hepatic contour, vascular landmarks and artifacts were rated by two independent readers using a four-point scale. For quantitative analysis, contrast-to-noise ratio (CNR) was measured in 170 solid focal lesions larger than 1 cm (107 hepatocellular carcinomas, nine cholangiocarcinomas and 54 metastases). Results : GRE_5.8 showed the highest mean points for hepatic contour, vascular anatomy and imaging artifact presence among all of the subjected sequences (p<0.001) and was comparable (p=0.414) with GRE_10 with regard to lesion conspicuity. The mean CNRs were significantly higher (p<0.001) in the following order: GRE_10 (24.4±14.5), GRE_5.8 (14.8±9.4), TSE (9.7±6.3), and GRE_2.4 (7.9±6.4). The mean CNRs of CCCs and metastases were higher than those of HCCs for all imaging sequences (p<0.05). Conclusion : Regarding overall performances, GRE using a moderate echo time of 5.8 msec can provide the most reliable data among the various fast breath-hold SPIOenhanced hepatic MRI sequences at 3T unit despite the lower CNR of GRE_5.8 compared to that of GRE_10.

참고문헌 (29)

  1. Van Beers BE, "Contrast-enhanced MR iaging of the liver", Radiology, 203, pp.297~306 (1997)
  2. Reimer P, "Hepatic MRI with SPIO: detection and characterization of focal liver lesions", Eur Radiol, 8, pp.1198~1204 (1998)
  3. von Falkenhausen M, "Intraindividual comparison of image contrast in SPIO-enhanced liver MRI at 1.5T and 3.0T", Eur Radiol, 17, pp.1256~1261 (2007)
  4. Chang JM, "Superparamagnetic iron oxide-enhanced liver magnetic resonance imaging-comparison of 1.5T and 3.0T imaging for detection of focal malignant liver lesions", Invest Radiol, 41, pp.168~174 (2006)
  5. Ramalho M, "Zapparoli M, Semelka R. Liver MR imaging: 1.5T versus 3T", Magn Reson Imaging Clin N Am, 15, pp.321~347 (2007)
  6. von Falkenhausen MM, "High-field-strength MR imaging of the liver at 3.0T: Intraindividual comparative study with MR imaging at 1.5T", Radiology, 241, pp.156~166 (2006)
  7. Kim T, "Effect of superparamagnetic iron oxide on tumor-to-liver contrast at T2*-weighted gradient-echo MRI:comparison between 3.0T and 1.5T MR systems", J Magn Reson Imaging, 29, pp.595~600 (2009)
  8. Nakada T, "Clinical Experience on 3.0T systems in Niigata, 1996 to 2002", Invest Radiol, 38, pp.377~384 (2003)
  9. Ward J, "Colorectal hepatic metastases: detection with SPIO-enhanced breath-hold MR imaging-comparison of optimized sequences", Radiology, 228, pp.709~718 (2003)
  10. Kim MJ, "Optimal TE for SPIOenhanced gradient-recalled echo MRI for the detection of focal hepatic lesions", AJR Am J Roentgenol, 187, pp.W255~W266 (2006)
  11. de Bazelaire CM, "MR Imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0T : preliminary results", Radiology, 230, pp.652~659 (2004)
  12. Bottomley PA, "A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age", Med Phys, 11, pp.425~448 (1984)
  13. Zapparoli M, "3.0-T MRI evaluation of patients with chronic liver diseases : initial observations", Magn Reson Imaging, 26, pp.650~660 (2008)
  14. Hahn PF, "Clinical application of superparamagnetic iron oxide to MR imaging of tissue perfusion in vascular liver tumors", Radiology, 174, pp.361~366 (1990)
  15. Schneider G, "Contrast agents in abdominal imaging-current and future directions", Top Magn Reson Imaging, 16, pp.107~124 (2005)
  16. Bellin MF, "Liver metastases: safety and Efficacy of detection with superparamagnetic iron oxide in MR imaging", Radiology, 193, pp.657~663 (1994)
  17. Matsuo M, "Detection of malignant hepatic tumors with ferumoxides-enhanced MRI: Comparison of five gradient-recalled echo sequences with different TEs", AJR Am J Roentgenol, 182, pp.235~242 (2004)
  18. Yoshikawa T, "Gradient- and spin-echo T2-weighted imaging for SPIO-enhanced detection and characterization of focal liver lesions", J Magn Reson Imaging, 23, pp.712~719 (2006)
  19. Ward J, "Hepatic lesion detection after superparamagnetic iron oxide enhancement: comparison of five T2-weighted sequences at 1.0 T by using alternativefree response receiver operating characteristic analysis", Radiology, 214, pp.159~166 (2000)
  20. Kurokawa H, "Experimental study of fast and ultrafast T2-weighted imaging sequences using AMI-25 superparamagnetic iron oxide (SPIO)", Acta Med Okayama, 55, pp.41~50 (2001)
  21. Kumano S, "Using superparamagnetic iron oxide-enhanced MRI to differentiate metastatic hepatic tumors and nonsolid benign lesions", AJR Am J Roentgenol, 181, pp.1335~1339 (2003)
  22. Tanimoto A, "Superparamagnetic iron oxide-mediated hepatic signal intensity change in patients with and without cirrhosis: pulse sequence effects and Kupffer cell function", Radiology, 222, pp.661~666 (2002)
  23. Kanematsu M, "Malignant hepatic tumor detection with ferumoxides-enhanced MR imaging with a 1.5-T system: comparison of four imaging pulse sequences", J Magn Reson Imaging, 13, pp.249~257 (2001)
  24. Elizondo G, "Hepatic cirrhosis and hepatitis: MR imaging enhanced with superparamagnetic iron oxide", Radiology, 174, pp.797~801 (1990)
  25. Zech CJ, "hoenberg SO. Black-blood diffusion-weighted EPI acquisition of the liver with paraller imaging. Comparison with a standard T2-weighted sequence for detection of focal liver lesions", Invest Radiol, 43, pp.261~266 (2008)
  26. Kuwatsuru R, "Definition of liver tumors in the presence of diffuse liver disease: comparison of findings at MR imaging with positive and negative contrast agents", Radiology, 202, pp.131~138 (1997)
  27. Tanimoto A, , Radiology, 222, pp.661~666 (2002)
  28. Tanimoto A, "Relaxation effects of clustered particles", J Magn Reson Imaging, 14, pp.72~77 (2001)
  29. Tang Y, "Detection of hepatocellular carcinoma arising in cirrhotic livers: comparison of gadolinium and ferumoxides-enhanced MR imaging", AJR Am J Roentgenol, 172, pp.1547~1554 (1999)

관련 논문AI추천

Comparison of Three, Motion-Resistant MR Sequences on Hepatobiliary Phase for Gadoxetic Acid (Gd-EOB-DTPA)- Enhanced MR Imaging of the Liver고명주, 고수연, 김두리, 김봉수, 김승형, 송병철, 이경렬, 이무숙, 이정섭, 최국명대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2017 · 제21권 제2호Focal Liver Lesion Detection in Gadoxeic Acid-enhanced Liver MRI : Effects of Scan Delay, Hepatic Fuction, and Magnetic Field Strength김기황, 김명진, 김화선, 박미숙, 임준석, 정용은, 정재준대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2011 · 제15권 제3호Usefulness of Gadobenate Dimeglumine - Enhanced Hepatobiliary Phase MR Imaging on Predicting Histological Grade of Hepatocellular Carcinoma김기황, 김명진, 박성호, 임준석, 최진영대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2011 · 제15권 제3호Optimization of the Flip Angle and Scan Timing in Hepatobiliary Phase Imaging Using T1-Weighted, CAIPIRINHA GRE ImagingBong Soo Kim, Guk Myung Choi, Ho Kyu Lee, Jeongjae Kim, Jeong Sub Lee, Joon Hyuk Park, Kyung Ryeol Lee, Mu Sook Lee, Seung Hyoung Kim, Seung Tae Woo대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2018 · 제22권 제1호Advanced Methods in Dynamic Contrast Enhanced Arterial Phase Imaging of the Liver김윤철대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2019 · 제23권 제1호Image Quality and Focal Lesion Detectability Analysis of Multiband Variable-Rate Selective Excitation Diffusion-Weighted Imaging of the Liver Using 3.0-T MRI신재승, 정용은, Eunju Kim, Ja Kyung Yoon대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2024 · 제28권 제1호

이 간행물의 다른 논문 (9건)