본문 바로가기
KERD
논문KCI 등재

K3에 축적된 내부 응력이 피로 파절에 미치는 영향

Effect of internal stress on cyclic fatigue failure in K3

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Objectives: This study aimed to evaluate the relationship between the cyclic fatigue of a K3 file and internal stress intentionally induced until the activation of the autostop function of the torque-controlled motor. Materials and Methods: K3 (Sybron Endo) .04 and .06 taper, size 25, 30, 35, 40 and 45 were used in this study. To give the internal stress, the K3 files were put into the .02 taper Endo-Training-Bloc (Dentsply Maillefer) until the activation of the auto-stop function of the torque-controlled motor. The rotation speed was 300 rpm and torque value was 1.0 N·cm. K3 were grouped by the number of induced internal stress and randomly distributed to 4 experimental groups (n = 10, Stress 0 [control], Stress 1, Stress 2 and Stress 3). For measuring the cyclic fatigue failure, the K3 files were worked against a sloped glass block and time for file separation was recorded. Data was statistically analyzed Statistical analyses were performed using two-way ANOVA and Duncan post-hoc test at p < 0.05 level. Results:Except .04 taper size 30 in Stress 1 group, there were statistically significant differences in time for file separation between control and all experimental groups. K3 with .04 taper showed higher cyclic fatigue resistance than those of .06 taper. Conclusion:In the limitation of this study, the cyclic fatigue of the K3 file was influenced by the accumulated internal stress from use until the auto-stop function was activated by the torque-controlled motor. Therefore, clinicians should avoid the reuse of the K3 file that has undergone auto-stops.

Abstract

Objectives: This study aimed to evaluate the relationship between the cyclic fatigue of a K3 file and internal stress intentionally induced until the activation of the autostop function of the torque-controlled motor. Materials and Methods: K3 (Sybron Endo) .04 and .06 taper, size 25, 30, 35, 40 and 45 were used in this study. To give the internal stress, the K3 files were put into the .02 taper Endo-Training-Bloc (Dentsply Maillefer) until the activation of the auto-stop function of the torque-controlled motor. The rotation speed was 300 rpm and torque value was 1.0 N·cm. K3 were grouped by the number of induced internal stress and randomly distributed to 4 experimental groups (n = 10, Stress 0 [control], Stress 1, Stress 2 and Stress 3). For measuring the cyclic fatigue failure, the K3 files were worked against a sloped glass block and time for file separation was recorded. Data was statistically analyzed Statistical analyses were performed using two-way ANOVA and Duncan post-hoc test at p < 0.05 level. Results:Except .04 taper size 30 in Stress 1 group, there were statistically significant differences in time for file separation between control and all experimental groups. K3 with .04 taper showed higher cyclic fatigue resistance than those of .06 taper. Conclusion:In the limitation of this study, the cyclic fatigue of the K3 file was influenced by the accumulated internal stress from use until the auto-stop function was activated by the torque-controlled motor. Therefore, clinicians should avoid the reuse of the K3 file that has undergone auto-stops.

참고문헌 (24)

  1. Walia HM, "An initial investigation of the bending and torsional properties of Nitinol root canal files", J Endod, 14, pp.346~351 (1988)
  2. Gambill JM, "Comparison of nickeltitanium and stainless steel hand-file instrumentation using computed tomography", J Endod, 22, pp.369~375 (1996)
  3. Glossen CR, "A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-Flex endodontic instruments", J Endod, 21, pp.146~151 (1995)
  4. Elliott LM, "Cutting pattern of nickel-titanium files using two preparation techniques", Endod Dent Traumatol, 14, pp.10~15 (1998)
  5. Cheung GSP, "Instrument fracture: mechanisms, removal of fragments, and clinical outcomes", Endodontic Topics, 16, pp.1~26 (2009)
  6. Kramkowski TR, "An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT Series X rotary nickel-titanium files", J Endod, 35, pp.404~407 (2009)
  7. Pruett JP, "Cyclic fatigue testing of nickel-titanium endodontic instruments", J Endod, 23, pp.77~85 (1997)
  8. Sattapan B, "Defects in rotary nickel-titanium files after clinical use", J Endod, 26, pp.161~165 (2000)
  9. Kuhn G, "Influence of structure on nickel-titanium endodontic instruments failure", J Endod, 27, pp.516~520 (2001)
  10. Gambarini G, "Advantages and disadvantages of new torque-controlled endodontic motors and low-torque NiTi rotary instrumentation", Aust Endod J, 27, pp.99~104 (2001)
  11. Berutti E, "Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments", J Endod, 30, pp.228~230 (2004)
  12. Barbosa FO, "Fractographic analysis of K3 nickel-titanium rotary instruments submitted to different modes of mechanical loading", J Endod, 34, pp.994~998 (2008)
  13. Li UM, "Cyclic fatigue of endodontic nickel titanium rotary instruments: static and dynamic tests", J Endod, 28, pp.448~451 (2002)
  14. 박세희, "근관치료용 전기모터의 회전 안정성 평가", Restorative Dentistry and Endodontics, 35(4), pp.246~256 (2010), 대한치과보존학회
  15. Melo MC, "Dimensional characterization and mechanical behaviour of K3 rotary instruments", Int Endod J, 41, pp.329~338 (2008)
  16. Plotino G, "A review of cyclic fatigue testing of nickel-titanium rotary instruments", J Endod, 35, pp.1469~1476 (2009)
  17. Bahia MG, "Decrease in the fatigue resistance of nickel-titanium rotary instruments after clinical use in curved root canals", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 100, pp.249~255 (2005)
  18. Haikel Y, "Dynamic fracture of hybrid endodontic hand instruments compared with traditional files", J Endod, 17, pp.217~220 (1991)
  19. Haïkel Y, "Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments", J Endod, 25, pp.434~440 (1999)
  20. Cheung GS, "Fatigue testing of a NiTi rotary instrument. Part 1: strain-life relationship", Int Endod J, 40, pp.612~618 (2007)
  21. Schäfer E, "Bending properties of rotary nickel-titanium instruments", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 96, pp.757~763 (2003)
  22. Ullmann CJ, "Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments", J Endod, 31, pp.183~186 (2005)
  23. 신유미, "전동식 니켈 티타늄 파일의 표면 결함 및 단면 형태가 반복응력 하에서 피로 파절에 미치는 영향", Restorative Dentistry and Endodontics, 29(3), pp.267~272 (2004), 대한치과보존학회
  24. 이정규, "Fractographic 분석을 통한 HEROShaper 니켈티타늄 전동 파일의 피로파절에 미치는 표면결함의 역할", Restorative Dentistry and Endodontics, 32(2), pp.130~137 (2007), 대한치과보존학회

관련 논문AI추천

이 간행물의 다른 논문 (10건)