초록
Objectives: This study assessed the reliability of digital color measurements using images of resin composite specimens captured with a cellphone. Materials and Methods: The reference color of cylindrical specimens built-up with the use of resin composite (shades A1, A2, A3, and A4) was measured with a portable spectrophotometer (CIELab). Images of the specimens were obtained individually or pairwise (compared shades in the same photograph) under standardized parameters. The color of the specimens was measured in the images using RGB system and converted to CIELab system using image processing software. Whiteness index (WID) and color differences (ΔE00) were calculated for each color measurement method. For the cellphone, the ΔE00 was calculated between the pairs of shades in separate images and in the same image. Data were analyzed using 2-way repeated-measures analysis of variance (α = 0.05). Linear regression models were used to predict the reference ΔE00 values of those calculated using color measured in the images. Results: Images captured with the cellphone resulted in different WID values from the spectrophotometer only for shades A3 and A4. No difference to the reference ΔE00 was observed when individual images were used. In general, a similar ranking of ΔE00 among resin composite shades was observed for all methods. Stronger correlation coefficients with the reference ΔE00 were observed using individual than pairwise images. Conclusions: This study showed that the use of cellphone images to measure the color difference seems to be a feasible alternative providing outcomes similar to those obtained with the spectrophotometer.
Abstract
Objectives: This study assessed the reliability of digital color measurements using images of resin composite specimens captured with a cellphone. Materials and Methods: The reference color of cylindrical specimens built-up with the use of resin composite (shades A1, A2, A3, and A4) was measured with a portable spectrophotometer (CIELab). Images of the specimens were obtained individually or pairwise (compared shades in the same photograph) under standardized parameters. The color of the specimens was measured in the images using RGB system and converted to CIELab system using image processing software. Whiteness index (WID) and color differences (ΔE00) were calculated for each color measurement method. For the cellphone, the ΔE00 was calculated between the pairs of shades in separate images and in the same image. Data were analyzed using 2-way repeated-measures analysis of variance (α = 0.05). Linear regression models were used to predict the reference ΔE00 values of those calculated using color measured in the images. Results: Images captured with the cellphone resulted in different WID values from the spectrophotometer only for shades A3 and A4. No difference to the reference ΔE00 was observed when individual images were used. In general, a similar ranking of ΔE00 among resin composite shades was observed for all methods. Stronger correlation coefficients with the reference ΔE00 were observed using individual than pairwise images. Conclusions: This study showed that the use of cellphone images to measure the color difference seems to be a feasible alternative providing outcomes similar to those obtained with the spectrophotometer.
참고문헌 (34)
- Trifkovic B, "Color adjustment potential of resin composites", Clin Oral Investig, 22, pp.1601~1607 (2018)
- Westland S, "Investigation of the perceptual thresholds of tooth whiteness", J Dent, 67S, pp.S11~S14 (2017)
- Correa AC, "The impact of a customized tray on in-office bleaching tooth sensitivity: a randomized clinical trial", Oper Dent, 41, pp.15~22 (2016)
- Fernandes MT, "Preemptive use of naproxen on tooth sensitivity caused by in-office bleaching: a triple-blind, crossover, randomized clinical trial", Oper Dent, 42, pp.486~496 (2017)
- Peixoto AC, "Preemptive use of piroxicam on tooth sensitivity caused by in-office bleaching: a randomized clinical trial", Braz Dent J, 30, pp.498~504 (2019)
- Gokce HS, "Shade matching performance of normal and color vision-deficient dental professionals with standard daylight and tungsten illuminants", J Prosthet Dent, 103, pp.139~147 (2010)
- Imbery TA, "Dental shade matching and value discernment abilities of first-year dental students", J Prosthodont, 27, pp.821~827 (2018)
- Corcodel N, "Evaluation of small-group education on the shade determination ability of preclinical dental students-a controlled clinical trial", Eur J Dent Educ, 22, pp.e582~e587 (2018)
- Rezende M, "Effect of topical application of dipyrone on dental sensitivity reduction after in-office dental bleaching: a randomized, triple-blind multicenter clinical trial", J Am Dent Assoc, 149, pp.363~371 (2018)
- Klaric Sever E, "Clinical and patient reported outcomes of bleaching effectiveness", Acta Odontol Scand, 76, pp.30~38 (2018)
- Vaez SC, "Is a single preliminary session of in-office bleaching beneficial for the effectiveness of at-home tooth bleaching? a randomized controlled clinical trial", Oper Dent, 44, pp.E180~E189 (2019)
- Pereira R, "Dental prophylaxis influence in tooth color assessment-clinical study", J Esthet Restor Dent, 32, pp.586~592 (2020)
- Wee AG, "Lighting conditions used during visual shade matching in private dental offices", J Prosthet Dent, 115, pp.469~474 (2016)
- Igiel C, "Reliability of visual and instrumental color matching", J Esthet Restor Dent, 29, pp.303~308 (2017)
- Sullivan C, "A yellowness index for use in dentistry", J Dent, 91, pp.103244~ (2019)
- Chu SJ, "Dental color matching instruments and systems. Review of clinical and research aspects", J Dent, 38(Supplement 2), pp.e2~e16 (2010)
- Joiner A, "Tooth colour and whiteness: a review", J Dent, 67S, pp.S3~S10 (2017)
- Gómez-Polo C, "Comparison of the CIELab and CIEDE2000 color difference formulas", J Prosthet Dent, 115, pp.65~70 (2016)
- Pérez MM, "Development of a customized whiteness index for dentistry based on CIELAB color space", Dent Mater, 32, pp.461~467 (2016)
- Luo MR, "The development of the CIE 2000 colour‐difference formula: CIEDE2000", Color Res Appl, 26, pp.340~350 (2001)
- Brook AH, "The clinical measurement of tooth colour and stain", Int Dent J, 57, pp.324~330 (2007)
- Bhandari R, "In vivo comparative evaluation of esthetics after microabrasion and microabrasion followed by casein phosphopeptide-amorphous calcium fluoride phosphate on molar incisor hypomineralization-affected incisors", Contemp Clin Dent, 10, pp.9~15 (2019)
- Jarad FD, "The use of digital imaging for colour matching and communication in restorative dentistry", Br Dent J, 199, pp.43~49 (2005)
- Yamanel K, "Assessment of color parameters of composite resin shade guides using digital imaging versus colorimeter", J Esthet Restor Dent, 22, pp.379~388 (2010)
- Tam WK, "Dental shade matching using a digital camera", J Dent, 40(Supplement 2), pp.e3~e10 (2012)
- Guan YH, "The measurement of tooth whiteness by image analysis and spectrophotometry: a comparison.", J Oral Rehabil, 32, pp.7~15 (2005)
- Pérez MM, "Whiteness difference thresholds in dentistry", Dent Mater, 35, pp.292~297 (2019)
- Pecho OE, "Relevant optical properties for direct restorative materials", Dent Mater, 32, pp.e105~e112 (2016)
- Santos SMM, "Color changes caused by reduction on the dentin shade composite thickness", Braz Dent J, 29, pp.469~474 (2018)
- Dain SJ, "Illuminant and observer metamerism and the Hardy-Rand-Rittler color vision test editions", Vis Neurosci, 23, pp.685~694 (2006)
- Spink LS, "Comparison of an absolute and surrogate measure of relative translucency in dental ceramics", Dent Mater, 29, pp.702~707 (2013)
- Tung OH, "Development of digital shade guides for color assessment using a digital camera with ring flashes", Clin Oral Investig, 15, pp.49~56 (2011)
- Lazar R, "The accuracy of dental shade matching using cross-polarization photography", Int J Comput Dent, 22, pp.343~351 (2019)
- Mahn E, "Comparison of visual analog shade matching, a digital visual method with a cross-polarized light filter, and a spectrophotometer for dental color matching", J Prosthet Dent, 125, pp.511~516 (2021)