Repeated Ceramic Head Fracture after Ceramic-on-ceramic Total Hip Arthroplasty
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Alumina ceramic-on-ceramic total hip arthroplasty is a widely-adopted option, especially for younger patients, due low wear, scratch resistance, the wettable characteristics of alumina ceramics, and because of the relatively low biological reactivites of ceramic wear particles. However, fracture remains a significant problem because ceramics are inherently brittle, and few reports have been issued on short head fractures. In the described case, a ceramic head fracture occurred only 8 months after replacing the original head and with a new 28 mm head and liner. Both of these incidents may have been caused by head-to-liner impingement due to excessive anteversion of the metal shell. During the second revision surgery, this situation was managed successfully using a conventional elevated-polyethylene liner and a CoCr metal head, while retaining all metal implants.
Abstract
Alumina ceramic-on-ceramic total hip arthroplasty is a widely-adopted option, especially for younger patients, due low wear, scratch resistance, the wettable characteristics of alumina ceramics, and because of the relatively low biological reactivites of ceramic wear particles. However, fracture remains a significant problem because ceramics are inherently brittle, and few reports have been issued on short head fractures. In the described case, a ceramic head fracture occurred only 8 months after replacing the original head and with a new 28 mm head and liner. Both of these incidents may have been caused by head-to-liner impingement due to excessive anteversion of the metal shell. During the second revision surgery, this situation was managed successfully using a conventional elevated-polyethylene liner and a CoCr metal head, while retaining all metal implants.
참고문헌 (24)
- Barrack, RL, "Concerns about ceramics in THA", Clin Orthop Relat Res, 429, pp.73~79 (2004)
- Bierbaum BE, "Ceramic-on-ceramic bearings in total hip arthroplasty", Clin Orthop Relat Res, 405, pp.158~163 (2002)
- Sedel L, "Evolution of alumina-on-alumina implants: a review", Clin Orthop Relat Res, 379, pp.48~54 (2000)
- Hannouche D, "Fractures of ceramic bearings: history and present status", Clin Orthop Relat Res, 417, pp.19~26 (2003)
- Arenas A, "Ceramic femoral head fractures in total hip replacement", Int Orthop, 23, pp.351~352 (1999)
- Murphy SB, "Two- to 9-year clinical results of alumina ceramic-on-ceramic THA", Clin Orthop Relat Res, 453, pp.97~102 (2006)
- Pulliam IT, "Fracture of a ceramic femoral head after a revision operation. A case report", J Bone Joint Surg Am, 79, pp.118~121 (1997)
- Toran MM, "Fracture of a ceramic femoral head after ceramic-onceramic total hip arthroplasty", J Arthroplasty, 21, pp.1072~1073 (2006)
- Diwanji SR, "Fracture of the ABC ceramic liner: a report of three cases", Clin Orthop Relat Res, 464, pp.242~246 (2007)
- Ha YC, "Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty", Clin Orthop Relat Res, 458, pp.106~110 (2007)
- Hasegawa M, "Alumina ceramic-onceramic total hip replacement with a layered acetabular component", J Bone Joint Surg Br, 88, pp.877~882 (2006)
- Skinner HB, "Ceramic bearing surfaces", Clin Orthop Relat Res, 369, pp.83~91 (1999)
- D’Antonio J, "New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty", J Arthroplasty, 17, pp.390~397 (2002)
- Garino JP, "Modern ceramic-on-ceramic total hip systems in the United States: early results", Clin Orthop Relat Res, 379, pp.41~47 (2000)
- Park YS, "Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing", J Bone Joint Surg Am, 88, pp.780~787 (2006)
- Hasegawa M, "Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup", J Arthroplasty, 18, pp.658~661 (2003)
- Callaway GH, "Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplasty", J Arthroplasty, 10, pp.855~859 (1995)
- Maccauro G, "Fracture of a YTZP ceramic femoral head. Analysis of a fault", J Bone Joint Surg Br, 86, pp.1192~1196 (2004)
- Masonis JL, "Zirconia femoral head fractures: a clinical and retrieval analysis", J Arthroplasty, 19, pp.898~905 (2004)
- Willmann G, "Ceramic femoral head retrieval data", Clin Orthop Relat Res, 379, pp.22~28 (2000)
- Koo KH, "Isolated fracture of the ceramic head after thirdgeneration alumina-on-alumina total hip arthroplasty", J Bone Joint Surg Am, 90, pp.329~336 (2008)
- Kim YH, "Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear", J Bone Joint Surg Am, 87, pp.577~582 (2005)
- Luchetti WT, "Drain entrapment and titanium to ceramic head deposition: two unique complications following closed reduction of a dislocated total hip arthroplasty", J Arthroplasty, 13, pp.713~717 (1998)
- Yoo JJ, "Damage of an alumina-onalumina bearing surface from a difficult reduction of a total hip arthroplasty. A report of three cases", J Bone Joint Surg Am, 86-A, pp.376~378 (2004)