초록

Mechanisms of ceramic liner fractures have not yet been fully elucidated. Impingement between a stem and ceramic liner is a proposed cause of ceramic liner fractures. We experienced a case of ceramic liner fracture caused by direct impingement between the stem shoulder and the ceramic liner. This type of impingement, unlike impingements with a stem neck, has not been previously reported. While we assume that certain characteristics of the stem contributed to the impingement, we report this case to note that caution may be needed when using certain stem designs.

Abstract

Mechanisms of ceramic liner fractures have not yet been fully elucidated. Impingement between a stem and ceramic liner is a proposed cause of ceramic liner fractures. We experienced a case of ceramic liner fracture caused by direct impingement between the stem shoulder and the ceramic liner. This type of impingement, unlike impingements with a stem neck, has not been previously reported. While we assume that certain characteristics of the stem contributed to the impingement, we report this case to note that caution may be needed when using certain stem designs.

참고문헌 (10)

  1. Lee YK, "Midterm results of the BIOLOX delta ceramic-on-ceramic total hip arthroplasty", Bone Joint J, 99, pp.741~748 (2017)
  2. Lim SJ, "Clinical outcomes and bearing-specific complications following fourth-generation alumina ceramic-on-ceramic total hip arthroplasty : a single-surgeon series of 749 hips at a minimum of 5-year follow-up", J Arthroplasty, 33, pp.2182~2186.e1 (2018)
  3. Ha YC, "Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty", Clin Orthop Relat Res, 458, pp.106~110 (2007)
  4. Park YS, "Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing", J Bone Joint Surg Am, 88, pp.780~787 (2006)
  5. Massin P, "Does Biolox Delta ceramic reduce the rate of component fractures in total hip replacement?", Orthop Traumatol Surg Res, 100(6 Suppl), pp.S317~321 (2014)
  6. Howard DP, "Ceramic-on-ceramic bearing fractures in total hip arthroplasty : an analysis of data from the National Joint Registry", Bone Joint J, 99, pp.1012~1019 (2017)
  7. Lee YK, "Metal neck and liner impingement in ceramic bearing total hip arthroplasty", J Orthop Res, 29, pp.218~222 (2011)
  8. Traina F, "Risk factors for ceramic liner fracture after total hip arthroplasty", Hip Int, 22, pp.607~614 (2012)
  9. Lee YK, "Effect of inner taper angle of acetabular metal shell on the malseating and dissociation force of ceramic liner", J Arthroplasty, 32, pp.1360~1362 (2017)
  10. Bachhal V, "A new method of measuring acetabular cup anteversion on simulated radiographs", Int Orthop, 36, pp.1813~1818 (2012)