브릿지 형태의 상부보철물 및 상부보철물의 나사풀림에 따른 타진식 임플란트 안정성 측정기기의 수치비교

Study of the Differences in a Percussion-type Implant Stability Measuring Instrument according to Bridge prosthesis and screw loosening

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

This study aimed to investigate the effect of different conditions on the implant stability test (IST) value of three-unit implant bridges. Fifteen pairs of specimens were prepared. Two internal conical-type implants were placed in artificial bone using a torque of 35 to 45 Ncm. Three-unit monolithic zirconia bridges (Perfit ZR) were fabricated using CAD-CAM. Implant Stability Quotient (ISQ) values were measured for each implant. IST value was measured under different conditions: healing abutment, without cementation, with cementation, one prosthesis unscrewed, and two prostheses unscrewed. The results showed statistically significant differences according to the conditions (p <0.05). The lowest IST level was observed in the groups tested on healing abutment and the prosthesis without cementation. The highest IST level was observed in the groups with cemented prostheses. There was no significant difference according to screw loosening. IST values for all groups measured above 79, indicating high stability with consistent tendency. The present study implied that IST values could be used as a reference restrictedly for assessing implant stability and conditions on bridge-type prostheses.

Abstract

This study aimed to investigate the effect of different conditions on the implant stability test (IST) value of three-unit implant bridges. Fifteen pairs of specimens were prepared. Two internal conical-type implants were placed in artificial bone using a torque of 35 to 45 Ncm. Three-unit monolithic zirconia bridges (Perfit ZR) were fabricated using CAD-CAM. Implant Stability Quotient (ISQ) values were measured for each implant. IST value was measured under different conditions: healing abutment, without cementation, with cementation, one prosthesis unscrewed, and two prostheses unscrewed. The results showed statistically significant differences according to the conditions (p <0.05). The lowest IST level was observed in the groups tested on healing abutment and the prosthesis without cementation. The highest IST level was observed in the groups with cemented prostheses. There was no significant difference according to screw loosening. IST values for all groups measured above 79, indicating high stability with consistent tendency. The present study implied that IST values could be used as a reference restrictedly for assessing implant stability and conditions on bridge-type prostheses.

참고문헌 (20)

  1. Brånemark PI, "Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period", Scand J Plast Reconstr Surg Suppl, 16, pp.1~132 (1977)
  2. Mitsias M, "Immediate, early(6 weeks)and delayed loading(3 months)of single, partial and full fixed implant supported prostheses : 1-year post-loading data from a multicentre randomised con-trolled trial", Eur J Oral Implantol, 11(1), pp.63~75 (2018)
  3. Al-Sawai AA, "Success of immediate loading implants compared to conventionally-loaded implants : a literature review", J Investig Clin Dent, 7(3), pp.217~224 (2016)
  4. Mistry G, "Measuring im-plant stability : A review of different methods", J Dent Implant, 4, pp.165~169 (2014)
  5. Roberts WE, "Osseous adaptation to continuous loading of rigid endosseous implants", Am J Orthod, 86(2), pp.95~111 (1984)
  6. Atsumi M, "Methods used to assess im-plant stability : current status", Int J Oral Maxillofac Implants, 22, pp.743~548 (2007)
  7. Herrero-Climent M, "Assessment of Osstell ISQ's reliability for implant stability measurement : a cross-sectional clinical study", Med Oral Patol Oral Cir Bucal, 18(6), pp.e877~82 (2013)
  8. Schulte W, "Periotest for measuring periodontal characteristics- correlation with periodontal bone loss", J Periodont Res (1992)
  9. Hugo Chavez, "Assessment of oral implant mobility", The Journal of Prosthetic Dentistry, 70(5), pp.421~426 (1993), Elsevier BV[DOI]
  10. Hakan Bilhan, "The Evaluation of the Reliability of Periotest for Implant Stability Measurements: An In Vitro Study", Journal of Oral Implantology, 41(4), pp.e90~e95 (2015), American Academy of Implant Dentistry[DOI]
  11. Teerlinck J, "Perio test : An object clinical diagnosis of bone apposition toward im-plants", Int J Oral Maxillofac Implants, 6, pp.55~61 (1991)
  12. Olive J, "The periotest method as a measure of os-seointegrated oral implant stability", Int J Oral Maxillofac Im-plants, 5, pp.390~400 (1990)
  13. 이정원, "Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study", Journal of Periodontal & Implant Science, 50(1), pp.56~66 (2020), 대한치주과학회[DOI]
  14. 옥성미, "상부 보철물에 따른 타진식 임플란트 안정성 측정기기의 수치 변화", 대한통합치과학회지, 10(1), pp.1~7 (2021), 대한통합치과학회[DOI]
  15. 박양훈, "임플란트 healing abutment 높이와 타진각도에 따른 타진방식 임플란트 안정성 측정기기의 수치 차이", 대한치과보철학회지, 56(4), pp.278~286 (2018), 대한치과보철학회[DOI]
  16. Elias CN, "Mechanical and clinical properties of titanium and titanium-based alloys (Ti G2, Ti G4 cold worked nanostructured and Ti G5) for biomedical applications", J Mater Res Technol, 8, pp.1060~1069 (2019)
  17. Candido LM, "Mechanical and surface properties of monolithic zirconia", Oper Dent, 43, pp.E119~28 (2018)
  18. Bozkaya, D., "Mechanics of the tapered interference fit in dental implants", Journal of Biomechanics, 36, pp.1649~1658 (2003)
  19. Lee JH, "Axial displacements in external and internal implant-abutment connection", Clin Oral Implants Res, 25(2), pp.e83~9 (2014)
  20. Meredith N, "Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis", Clin Oral Implants Res, 7, pp.261~267 (1996)

관련 논문AI추천

이 간행물의 다른 논문 (4건)