임플란트 안정성 평가를 위한 타진 기기의 측정 각도에 따른 반복성 분석: 생체외 연구

Measuring Device According to the Measurement Angle: an In Vitro Study

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Damping capacity analysis (DCA) devices used for evaluating dental implant stability are limited by their applicable measurement angles. Recently developed DCA devices have been designed to function over a wide range of angles from 0° to 90°. The purpose of this study was to compare implant stability values (ISV) obtained at various measurement angles using a newly developed DCA device and to assess the reproducibility of ISV measurements at each angle. Bone blocks (8.0 × 8.0 × 30 mm) were prepared from the central third of porcine ribs, and implants (4.0 mm diameter, 10.0 mm length) were placed. Initial stability was evaluated using peak insertion torque and the implant stability quotient (ISQ) values. Using a rotatable table, ISV were measured at 0°, 45°, and 90°, with three repeated measurements at each angle to calculate the intraclass correlation coefficient (ICC). Repeated measures analysis of variance (RMANOVA) was performed on the mean ISV values obtained at each of the three angles to determine whether any significant differences existed among the measurement angles. ICC analysis showed high reproducibility at all angles, with ICCs of 0.998 at 0°, 0.995 at 45°, and 0.987 at 90°. RMANOVA revealed a statistically significant difference in ISV values among the three angles (p<.001). The mean ISV was 72.097±3.425 at 0°, 70.965±3.702 at 45°, and 69.930±3.649 at 90°. Paired t-test with Bonferroni correction revealed significant differences between all angle pairs: between 0° and 45° (p<.001), between 45° and 90° (p<001), and between 0° and 90° (p<.001). The newly developed DCA device showed excellent reproducibility at all tested angles (0°, 45°, and 90°). The decrease in ISV with increasing measurement angles was statistically significant.

Abstract

Damping capacity analysis (DCA) devices used for evaluating dental implant stability are limited by their applicable measurement angles. Recently developed DCA devices have been designed to function over a wide range of angles from 0° to 90°. The purpose of this study was to compare implant stability values (ISV) obtained at various measurement angles using a newly developed DCA device and to assess the reproducibility of ISV measurements at each angle. Bone blocks (8.0 × 8.0 × 30 mm) were prepared from the central third of porcine ribs, and implants (4.0 mm diameter, 10.0 mm length) were placed. Initial stability was evaluated using peak insertion torque and the implant stability quotient (ISQ) values. Using a rotatable table, ISV were measured at 0°, 45°, and 90°, with three repeated measurements at each angle to calculate the intraclass correlation coefficient (ICC). Repeated measures analysis of variance (RMANOVA) was performed on the mean ISV values obtained at each of the three angles to determine whether any significant differences existed among the measurement angles. ICC analysis showed high reproducibility at all angles, with ICCs of 0.998 at 0°, 0.995 at 45°, and 0.987 at 90°. RMANOVA revealed a statistically significant difference in ISV values among the three angles (p<.001). The mean ISV was 72.097±3.425 at 0°, 70.965±3.702 at 45°, and 69.930±3.649 at 90°. Paired t-test with Bonferroni correction revealed significant differences between all angle pairs: between 0° and 45° (p<.001), between 45° and 90° (p<001), and between 0° and 90° (p<.001). The newly developed DCA device showed excellent reproducibility at all tested angles (0°, 45°, and 90°). The decrease in ISV with increasing measurement angles was statistically significant.

관련 논문AI추천

Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study구기태, 안정섭, 이용무, 이재현, 이정원, 조현재, 표세욱대한치주과학회 · Journal of Periodontal & Implant Science2020 · 제50권 제1호Effects of Striking and Bone Conditions on Damping Capacity Assessment: An in vitro Study강대영, 신현승, 안현성, 조인우대한구강악안면임플란트학회 · 대한구강악안면임플란트학회지2023 · 제27권 제2호Reliability of implant stability measuring devices depending on various clinical conditions: an in vitro study김명섭, 류재준, 심지석, 이정열, 이한나, Xu Zihan대한치과보철학회 · The Journal of Advanced of Prosthodontics2023 · 제15권 제3호임플란트 healing abutment 높이와 타진각도에 따른 타진방식 임플란트 안정성 측정기기의 수치 차이박양훈, 백장현, 이석원, 이성복, 이정열대한치과보철학회 · 대한치과보철학회지2018 · 제56권 제4호Comparative analysis of novel damping capacity analysis devices for assessing implant stability in clinical practiceBo-Ah Lee, Jung-Seok Lee, Su-Bin Hong, Woo-Ri Lee, Young-Taek Kim대한치주과학회 · Journal of Periodontal & Implant Science2025 · 제55권 제6호브릿지 형태의 상부보철물 및 상부보철물의 나사풀림에 따른 타진식 임플란트 안정성 측정기기의 수치비교강선우, 권연우, 김복주, 김정한, 김철훈, 성현미대한치과의사협회 · 대한치과의사협회지2022 · 제61권 제1호

이 간행물의 다른 논문 (4건)