In order to reduce the radiation exposure dose of the patient and to obtain accurate diagnosis results, the quality control of the diagnostic radiation generator must be conducted periodically In particular, bone density test equipment could be influenced by many factors, and it is far more important because inaccurate measurement would eventually affect the result value. However, the cross-correction phantom of DXA equipment is poorly penetrated due to lack of awareness of the industry and the high cost. Therefore, this study developed a BMD phantom using a 3D printer and Korean BMD phantom with low cost by cross analyzing Korean BMD value from The Korean National Health and Nutrition Examination Survey and evaluated it. The L1, L2, and L3 BMD values of phantoms produced with the 3D printer were measured to be 0.887±0.006 g/cm2, 0.927±0.006 g/cm2, and 0.960±0.005 g/cm2, at 215 mm height and 0.882±0.011 g/cm2, 0.914±0.005 g/cm2, 0.933±0.008 g/cm2 at 155 mm height displaying statistically significant relevance. The result suggests that a proper quality control and cross calibration of DXA device be possible and expected to be an essential data for various medical phantom manufacture development using 3D printer.

Abstract

In order to reduce the radiation exposure dose of the patient and to obtain accurate diagnosis results, the quality control of the diagnostic radiation generator must be conducted periodically In particular, bone density test equipment could be influenced by many factors, and it is far more important because inaccurate measurement would eventually affect the result value. However, the cross-correction phantom of DXA equipment is poorly penetrated due to lack of awareness of the industry and the high cost. Therefore, this study developed a BMD phantom using a 3D printer and Korean BMD phantom with low cost by cross analyzing Korean BMD value from The Korean National Health and Nutrition Examination Survey and evaluated it. The L1, L2, and L3 BMD values of phantoms produced with the 3D printer were measured to be 0.887±0.006 g/cm2, 0.927±0.006 g/cm2, and 0.960±0.005 g/cm2, at 215 mm height and 0.882±0.011 g/cm2, 0.914±0.005 g/cm2, 0.933±0.008 g/cm2 at 155 mm height displaying statistically significant relevance. The result suggests that a proper quality control and cross calibration of DXA device be possible and expected to be an essential data for various medical phantom manufacture development using 3D printer.

참고문헌 (19)

  1. 김영란, "한국인에서 연령에 따른 요추 및 대퇴부에서의 최대 골밀도 및 골소실률에 관한 융합 연구", 한국융합학회논문지, 9(5), pp.77~83 (2018), 한국융합학회[DOI]
  2. Kanis JA, "Assessment of fracture risk and its application to screening for postmenopausal osteoporosis", Osteoporosis International, 4(6), pp.368~381 (1994)
  3. 대한골대사학회 지침 위원회, "골다공증 진단 및 치료 지침 2007", Endocrinology and Metabolism, 23(2), pp.76~108 (2008), 대한내분비학회
  4. 이혜상, "우리나라 성인 남성의 골밀도에 연관성을 보이는 요인 : 국민건강영양조사자료, 2010∼2011", 대한지역사회영양학회지, 22(2), pp.136~144 (2017), 대한지역사회영양학회[DOI]
  5. 김영란, "종합병원 교대근무 여성들의 생활습관 및 영양섭취상태가 골밀도 값에 미치는 영향", 방사선기술과학, 35(1), pp.9~15 (2012), 대한방사선과학회
  6. 김명숙, "서울지역 성인여성의 골밀도 정상군·위험군의 식습관과 식품섭취 관련요인의 비교분석", 대한지역사회영양학회지, 13(1), pp.125~133 (2008), 대한지역사회영양학회
  7. 조동숙, "Bone Mineral Density and Factors Affecting in Female College Students", 여성건강간호학회지, 14(4), pp.297~305 (2008), 한국여성건강간호학회
  8. Gates BJ, "Management of osteoporosis in elderly men", Maturitas, 69(2), pp.113~119 (2011)
  9. 동경래, "골밀도검사의 올바른 질 관리에 따른 임상적용과 해석-이중 에너지 방사선 흡수법을 중심으로-", 방사선기술과학, 31(1), pp.17~24 (2008), 대한방사선과학회
  10. Kang BM, "Manual of bone densitometry" (2009), Cheongwoon
  11. Glüer CC, "Accurate assessment of precision errors : How to measure the reproducibility of bone densitometry techniques", Osteoporosis International, 5(4), pp.262~270 (1995)
  12. 김호성, "골밀도 측정의 정확한 정도관리방법 ― 이중 에너지 방사선 흡수법을 중심으로 ―", 방사선기술과학, 32(4), pp.361~370 (2009), 대한방사선과학회
  13. 김호성, "수도권 지역 병원에서 이중에너지 X-선 흡수측정법 (DXA)을 이용한 골밀도측정기 및 검사자의 정도관리에 관한 연구", 방사선산업학회지, 11(4), pp.283~289 (2017), (사)한국방사선산업학회[DOI]
  14. Yoon MS, "A Study on CT image quality and dose measurement Phantom using 3D printer", master’s, pp.2018
  15. Seo JN, "A phantom production by using 3-dimentional printer andIn-vivo dosimetry for a prostate cancer patient", Journal of Korean Society for Radiotherapeutic Technology, 27(1), pp.53~60 (2015)
  16. 윤명성, "3D프린터를 이용한 CT 선량측정 팬텀 제작 및 비교에 관한 연구", 한국방사선학회논문지, 12(6), pp.737~743 (2018), 한국방사선학회[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (9건)