골밀도검사에서 피검자의 장기선량 측정 및 차폐기구의 효용성 평가

Evaluation of the Effectiveness of the Shielding Device and the Organ Dose of Subject During Bone Mineral Density

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Bone mineral density is a examination to measure the amount of bone in patients with metabolic bone disease. It is a low dose, but may cause unnecessary exposure to the gonads and other organs located in the periphery when examining the lumbar and proximal femurs. Therefore, the purpose of this study was to evaluated the exposure dose for each organ exposed during the bone mineral density through simulation, and analyzed the applicability of the subject to radiation shielding devices using 3D printing materials. As a result, the highest dose was shown at 11.47 uSv in the breast during lumbar examination and 8.98 uSv in the testis during proximal femur examination. Also, the farther away from the examination site, the lower the effect of the scattering-ray. The shielding effect of using 3D printing shielding device showed high results in proportion to the effective atomic number and specific gravity of the printing material. Among t he p rinting materials, A BS + W showed an e ffect o f at l east 78.72 to 9 6.3 9% c ompared to t he e xisting lead material.

Abstract

Bone mineral density is a examination to measure the amount of bone in patients with metabolic bone disease. It is a low dose, but may cause unnecessary exposure to the gonads and other organs located in the periphery when examining the lumbar and proximal femurs. Therefore, the purpose of this study was to evaluated the exposure dose for each organ exposed during the bone mineral density through simulation, and analyzed the applicability of the subject to radiation shielding devices using 3D printing materials. As a result, the highest dose was shown at 11.47 uSv in the breast during lumbar examination and 8.98 uSv in the testis during proximal femur examination. Also, the farther away from the examination site, the lower the effect of the scattering-ray. The shielding effect of using 3D printing shielding device showed high results in proportion to the effective atomic number and specific gravity of the printing material. Among t he p rinting materials, A BS + W showed an e ffect o f at l east 78.72 to 9 6.3 9% c ompared to t he e xisting lead material.

참고문헌 (18)

  1. ICRP, "Managing Patient Dose in Digital Radiology. ICRP Publication 93", Annals of the ICRP, 34(1) (2004)
  2. ICRP, "2007 Recommndations of the International Commission on Radiological Protection. Publication 103", Annals of the ICRP, 137(2-4) (2007), Pergamon Press
  3. 한만석, "이중에너지 X-ray 흡수법을 적용한 근위 대퇴골 및 요추부 골밀도 검사의 상관관계 분석", 한국컴퓨터정보학회논문지, 17(9), pp.165~169 (2012), 한국컴퓨터정보학회
  4. Shin YR, "Assessment of bone mineral density", Journal of Korean Society of Pediatric Endocrinology, 11(2), pp.123~130 (2006)
  5. CHO Alliance, "Medical 3D printing, bio-organ, and bio-material market conditions and technology development strategy" (2017)
  6. OsteoSys Co. Ltd, "DEXXUM T User Manual, DOC. No. : OT08-2F7430-MUL-01", ~
  7. Guidelines for Diagnostic Reference Levels in Plain Radiography, "Korea Centers for Disease Control and Prevention", ~, pp.Medical Radiation Series, No.16 (2019)
  8. Guidelines for Diagnostic Reference Levels in Mammography, "Korea Centers for Disease Control and Prevention", ~, pp.Medical Radiation Series, No.17 (2019)
  9. Guidelines for Diagnostic Reference Levels in Computed Tomography(CT), "Korea Centers for Disease Control and Prevention", ~, pp.Medical Radiation Series, No.10 (2017)
  10. "Dual-energy X-ray absorptiometry (DE XA)" (2017)[URL]
  11. Guglielmi G, "Radiation exposure in X-raybased imaging techniques used in osteoporosis", European Radiology, 20(11), pp.2707~2714 (2010)
  12. Jung DR, "Genital gland dose and reduction due to scattered radiation measured bone mineral density with dual energy X-ray absorptiometry", Master, Department of Radiological Science, pp.2016
  13. 한채연, "골밀도 검사 시 산란선에 의한 유방의 피폭선량을 경감시키기 위하여 제작한 차폐체(Breast shield)의 유용성 평가", 석사, 의학물리전공, pp.2013
  14. Karimian A, "Assessment of the absorbed dose to organs from bone mineral density scan by using TLDs and the monte carlo method", Nuclear Technology & Radiation Protection, 29(4), pp.289~295 (2014)
  15. Yoon JS, "The study of radiation dose on BMD using dual energy x-ray absorptiometry", Master, Department of Radiological Science, pp.2017
  16. 이준호, "3D 프린팅을 이용한 골밀도 팬텀 개발", 한국방사선학회논문지, 13(2), pp.185~192 (2019), 한국방사선학회[DOI]
  17. 박훈희, "3D 프린팅 기법 기반의 Hoffman Brain 팬텀 개발 및 유용성 평가", 방사선기술과학, 42(6), pp.441~446 (2019), 대한방사선과학회
  18. 안기송, "전자선치료 시 3D 프린터로 제작한 환자 맞춤형 차폐체의 유용성 평가", 방사선기술과학, 42(6), pp.447~454 (2019), 대한방사선과학회

관련 논문AI추천

이 간행물의 다른 논문 (5건)