The purpose of this study is to evaluate the dose distribution by gantry rotation and MLC moving speed on treatment planning system(TPS) and linear accelerator. The dose analyzer phantom(Delta 4) was scanned by CT simulator for treatment planning. The planning target volumes(PTVs) of prostate and pancreas was prescribed 6,500 cGy, 5,000 cGy on VMAT(Volumetric Modulated Arc Therapy) by TPS while MLC speed changed. The analyzer phantom was irradiated linear accelerator using by planned parameters. Dose distribution of PTVs were evaluated by the homogeneity index, conformity index, dose volume histogram of organ at risk(rectum, bladder, spinal cord, kidney). And irradiated dose analysis were evaluated dose distribution and conformity by gamma index. The PTV dose of pancreas was 4,993 cGy during 0.1 ㎝/deg leaf and gantry that was the most closest prescribed dose(5,000 cGy). The dose of spinal cord, left kidney, and right kidney were accessed the lowest during 0.1 ㎝/deg, 1.5 ㎝/deg, 0.3 ㎝/deg. The PTV dose of prostate was 6,466 cGy during 0.1 ㎝/deg leaf and gantry that was the most closest prescribed dose(6,500 cGy). The dose of bladder and rectum were accessed the lowest during 0.3 ㎝/deg, 2.0 ㎝/deg. For gamma index, pancreas and prostate were analyzed the lowest error 100% at 0.8, 1.0 ㎝/deg and 99.6% at 0.3, 0.5 ㎝/deg. We should used the optimal leaf speed according to the gantry rotation if the treatment cases are performed VMAT.

Abstract

The purpose of this study is to evaluate the dose distribution by gantry rotation and MLC moving speed on treatment planning system(TPS) and linear accelerator. The dose analyzer phantom(Delta 4) was scanned by CT simulator for treatment planning. The planning target volumes(PTVs) of prostate and pancreas was prescribed 6,500 cGy, 5,000 cGy on VMAT(Volumetric Modulated Arc Therapy) by TPS while MLC speed changed. The analyzer phantom was irradiated linear accelerator using by planned parameters. Dose distribution of PTVs were evaluated by the homogeneity index, conformity index, dose volume histogram of organ at risk(rectum, bladder, spinal cord, kidney). And irradiated dose analysis were evaluated dose distribution and conformity by gamma index. The PTV dose of pancreas was 4,993 cGy during 0.1 ㎝/deg leaf and gantry that was the most closest prescribed dose(5,000 cGy). The dose of spinal cord, left kidney, and right kidney were accessed the lowest during 0.1 ㎝/deg, 1.5 ㎝/deg, 0.3 ㎝/deg. The PTV dose of prostate was 6,466 cGy during 0.1 ㎝/deg leaf and gantry that was the most closest prescribed dose(6,500 cGy). The dose of bladder and rectum were accessed the lowest during 0.3 ㎝/deg, 2.0 ㎝/deg. For gamma index, pancreas and prostate were analyzed the lowest error 100% at 0.8, 1.0 ㎝/deg and 99.6% at 0.3, 0.5 ㎝/deg. We should used the optimal leaf speed according to the gantry rotation if the treatment cases are performed VMAT.

참고문헌 (16)

  1. 고혜진, "전유방절제술 환자에서 IMRT와 VMAT을 이용한 방사선치료시 선량 분포의 비교", 한국콘텐츠학회 논문지, 14(8), pp.225~232 (2014), 한국콘텐츠학회[DOI]
  2. 강동진, "AAPM TG-119 보고서를 통한 세기조절방사선치료(IMRT)와부피적세기조절회전치료(VMAT)의 치료 전 환자별 정도관리", 방사선기술과학, 35(3), pp.255~264 (2012), 대한방사선과학회
  3. Rao M, "Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: Plan quality delivery efficiency and accuracy", Med Phy, 37, pp.1350~1359 (2010)
  4. 정재용, "선택적 최적화 변수(Selectable Optimization Parameters)에 따른 부피적조절회전방사선치료(VMAT)의 선량학적 영향", 의학물리, 23(1), pp.15~25 (2012), 한국의학물리학회[DOI]
  5. Darby SC, "Risk of ischemic heart disease in women after radiotherapy for breast cancer", N Engl J Med, 368, pp.987~998 (2013)
  6. Kim CY, "Evaluation of planning dose accuracy in case of radiation treatment on inhomogeneous organ structure", The Journal of Korean Society for Radiation Therapy, 25(2), pp.137~143 (2013)
  7. Park JM, "The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT", The British Instituet of Radiology, 88(1049), pp.698~703 (2015)
  8. 오세안, "Comparison of IMRT and VMAT Techniques in Spine Stereotactic Radiosurgery with International Spine Radiosurgery Consortium Consensus Guidelines", 의학물리, 24(3), pp.145~153 (2013), 한국의학물리학회[DOI]
  9. Masahide S, "Comparison of MLC error sensitivity of various commercial devices for VMAT pre-treatment quality assurance", J Appl Clin Med Phys, 19(3), pp.87~93 (2018)
  10. Oliver M, "Clinical significance of multileaf collimator positional errors for volumetric modulated arc therapy", Radiother Oncol, 97, pp.554~560 (2010)
  11. Todsaporn F, "A novel and independent method for time-rosolved gantry angle quality assurance for VMAT", Appl Clin Med Phys, 18(5), pp.134~142 (2017)
  12. Quan EM, "A comparison of IMRT and VMAT plan quality for prostate cancer treatment", Int J Radiat Oncol Biol Phys, 83, pp.1169~1178 (2012)
  13. Wolff D, "Volumetric modulated arc therapy vs. serial tomotherapy step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer", Radiother Oncol, 93, pp.226~233 (2009)
  14. Manikandan A, "Quality assurance of dynamic parameters in volumetric modulated arc therapy", Br J Radiol, 85, pp.1002~1007 (2012)
  15. Tyagi N, "A real time dose monitoring and dose reconstruction tool for patient specific VMAT QA and delivery", Med Phys, 39, pp.7194~7204 (2012)
  16. Kim JI, "The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT", Radiat Oncol, 9, pp.167~ (2014)

관련 논문AI추천

용적세기조절회전방사선치료 (VMAT) 시 피부선량 감소를 위한 Jaw Tracking기법의 유용성 평가공예진, 김수인, 김호성, 백금문, 신나영, 이충환(사)한국방사선산업학회 · 방사선산업학회지2020 · 제14권 제4호소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가강동진, 권경태, 김연래, 민정환, 신영주, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제1호용적변조회전 방사선치료 시 Set-up 오차에 따른선량 전달 정확성의 평가김상현, 나권우, 박성아, 이주훈, 하광은(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제1호두경부 IMRT 및 VMAT 시 체적 감소가 전산화치료계획에 미치는 영향김가중, 백금문, 엄기천대한방사선과학회 · Journal of Radiological Science and Technology2023 · 제46권 제3호AAPM TG-119 보고서를 통한 세기조절방사선치료(IMRT)와부피적세기조절회전치료(VMAT)의 치료 전 환자별 정도관리박승, 강동진, 김연래, 김종하, 손승창, 신영주, 이근섭, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2012 · 제35권 제3호Evaluation of Beam-Matching Accuracy for 8 MV Photon Beam between the Same Model Linear Accelerator강성희, 김연래, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2020 · 제43권 제2호

이 간행물의 다른 논문 (9건)