Evaluation of Beam-Matching Accuracy for 8 MV Photon Beam between the Same Model Linear Accelerator

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study aimed to assess of beam-matching accuracy for an 8 MV beam between the same model linear accelerators( Linac) commissioned over two years. Two models were got the customer acceptance procedure(CAP) criteria. For commissioning data for beam-matched linacs, the percentage depth doses(PDDs), beam profiles, output factors, multi- leaf collimator(MLC) leaf transmission factors, and the dosimetric leaf gap(DLG) were compared. In addition, the accuracy of beam matching was verified at phantom and patient levels. At phantom level, the point doses specified in TG-53 and TG-119 were compared to evaluate the accuracy of beam modelling. At patient level, the dose volume histogram( DVH) parameters and the delivery accuracy are evaluated on volumetric modulated arc therapy(VMAT) plan for 40 patients that included 20 lung and 20 brain cases. Ionization depth curve and dose profiles obtained in CAP showed a good level for beam matching between both Linacs. The variations in commissioning beam data, such as PDDs, beam profiles, output factors, TF, and DLG were all less than 1%. For the treatment plans of brain tumor and lung cancer, the average and maximum differences in evaluated DVH parameters for the planning target volume(PTV) and the organs at risk(OARs) were within 0.30% and 1.30%. Furthermore, all gamma passing rates for both beam-matched Linacs were higher than 98% for the 2%/2 ㎜ criteria and 99% for the 2%/3 ㎜ criteria. The overall variations in the beam data, as well as tests at phantom and patient levels remains all within the tolerance (1% difference) of clinical acceptability between beam-matched Linacs. Thus, we found an excellent dosimetric agreement to 8 MV beam characteristics for the same model Linacs.

Abstract

This study aimed to assess of beam-matching accuracy for an 8 MV beam between the same model linear accelerators( Linac) commissioned over two years. Two models were got the customer acceptance procedure(CAP) criteria. For commissioning data for beam-matched linacs, the percentage depth doses(PDDs), beam profiles, output factors, multi- leaf collimator(MLC) leaf transmission factors, and the dosimetric leaf gap(DLG) were compared. In addition, the accuracy of beam matching was verified at phantom and patient levels. At phantom level, the point doses specified in TG-53 and TG-119 were compared to evaluate the accuracy of beam modelling. At patient level, the dose volume histogram( DVH) parameters and the delivery accuracy are evaluated on volumetric modulated arc therapy(VMAT) plan for 40 patients that included 20 lung and 20 brain cases. Ionization depth curve and dose profiles obtained in CAP showed a good level for beam matching between both Linacs. The variations in commissioning beam data, such as PDDs, beam profiles, output factors, TF, and DLG were all less than 1%. For the treatment plans of brain tumor and lung cancer, the average and maximum differences in evaluated DVH parameters for the planning target volume(PTV) and the organs at risk(OARs) were within 0.30% and 1.30%. Furthermore, all gamma passing rates for both beam-matched Linacs were higher than 98% for the 2%/2 ㎜ criteria and 99% for the 2%/3 ㎜ criteria. The overall variations in the beam data, as well as tests at phantom and patient levels remains all within the tolerance (1% difference) of clinical acceptability between beam-matched Linacs. Thus, we found an excellent dosimetric agreement to 8 MV beam characteristics for the same model Linacs.

참고문헌 (17)

  1. Hrbacek J, "Quantitative evaluation of a beam-mathcing procedure using one-dimensional gamma analysis", Med Phys, 34, pp.2917~2927 (2007)
  2. Beyer GP, "Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators", J Appl Clin Med Phys, 14, pp.273~288 (2013)
  3. Glide-Hurst C, "Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study", Med Phys, 40(3) (2013)
  4. SjSstrSm D, "A beam-matching concept for medical linear accelerators", Acta Oncol, 48, pp.192~200 (2009)
  5. 강성희, "Possibility of Interchanging Patients for Beam-Matched Linear Accelerators from the Same Vendor", Journal of the Korean Physical Society, 75(8), pp.628~635 (2019), 한국물리학회[DOI]
  6. Fraass B, "American Association of Physicist in medicine radiation therapy committee task group 53 : quality assurance for clinical treatment planning", Med Phys, 25, pp.1773~1829 (1998)
  7. Ezzell GA, "IMRT commissioning : multiple institution planning and dosimetry comparison", Med Phys, 36, pp.5359~5383 (2009)
  8. Nithya L, "Analyzing the performance of the planning system by u se o f AAPM T G119 t est cases", Radiol Phys Technol, 9, pp.22~29 (2016)
  9. 김연래, "Comparison of the Performance between Portal Dosimetry and a Commercial Two-dimensional Array System on Pretreatment Quality Assurance for Volumetric-modulated Arc and Intensity-modulated Radiation Therapy", Journal of the Korean Physical Society, 64(8), pp.1202~1206 (2014), 한국물리학회
  10. Ashokkumar S, "Dosimetric validation of volumetric modulated arc therapy with three 6 MV beam-matched linear accelerators", Asian Pacific Journal of Cancer Prevention, 18, pp.3439~3444 (2017)
  11. Xu Z, "Assessment of beammatched Linacs quality", accuracy for interchanging SBRT or SRT patient using VMAT without replanning, 2019, pp.20(1), 68~75
  12. Watts RJ, "Comparative measurements on a series of accelerators by the same vendor", Med Phys, 26, pp.2581~2585 (1999)
  13. Attalla EM, "Dosimetric evaluation of a beam matching procedure", Chinese-German J Clin Oncol, 13, pp.89~93 (2014)
  14. Bhangle JR, "Dosimetric analysis of beam-matching procedure of two similar linear accelerators", J Med Phys, 36, pp.176~180 (2011)
  15. Krishnappan C, "Evaluation of beam matching accuracy among six Linacs from the same vendor", Radiol Phys Technol, 11, pp.423~433 (2018)
  16. 백태성, "Feasibility Study of Patient Specific Quality Assurance Using Transit Dosimetry Based on Measurement with an Electronic Portal Imaging Device", 의학물리, 28(2), pp.54~60 (2017), 한국의학물리학회[DOI]
  17. 손순룡, "Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy", 방사선기술과학, 38(4), pp.451~461 (2015), 대한방사선과학회[DOI]

관련 논문AI추천

Performance Evaluation of Stealth Chamber as a Novel Reference Chamber for Measuring Percentage Depth Dose and Profile of VitalBeam Linear Accelerator강상원, 강성희, 김경현, 김연래, 서태석, 신영주, 이정우, 정재용, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제3호용적변조회전 방사선치료 시 Set-up 오차에 따른선량 전달 정확성의 평가김상현, 나권우, 박성아, 이주훈, 하광은(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제1호Dose Volume Histogram Analysis for Comparison of Usability of Linear Accelerator Flattening Filter김미현, 동경래, 지연상, 최지원, Jae-Kwang Ryu(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제4호소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가강동진, 권경태, 김연래, 민정환, 신영주, 정재용대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제1호체적조절호형방사선치료 시 갠트리 회전과 다엽콜리메이터의 이동 속도에 따른 선량분포 평가강동진, 김연래, 신영주, 이정우, 정재용, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2019 · 제42권 제3호A Comparison of Patient-specific Delivery Quality Assurance (DQA) Devices in Radiation Therapy장경환대한방사선과학회 · Journal of Radiological Science and Technology2023 · 제46권 제3호

이 간행물의 다른 논문 (7건)