V 형상을 가지는 원자현미경 Cantilever의 정량적 마찰력 교정
Quantitative Lateral Force Calibration of V-shaped AFM Cantilever
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Atomic force microscopy (AFM) has been used as a tool, not only for imaging surfaces, but also for measuring surface forces and mechanical properties at the nano-scale. Force calibration is crucial for quantitatively measuring the forces that act between the AFM probe of a force sensing cantilever and a sample. In this work, the lateral force calibrations of a V-shaped cantilever were performed using the finite element method, multiple pivot loading, and thermal noise methods. As a result, it was shown that the multiple pivot loading method was appropriate for the lateral force calibration of a V-shaped cantilever. Further, through crosschecking of the abovementioned methods, it was concluded that the thermal noise method could be used for determining the lateral spring constants as long as the lateral deflection sensitivity was accurately determined. To obtain the lateral deflection sensitivity from the sticking portion of the friction loop, the contact stiffness should be taken into account.
Abstract
Atomic force microscopy (AFM) has been used as a tool, not only for imaging surfaces, but also for measuring surface forces and mechanical properties at the nano-scale. Force calibration is crucial for quantitatively measuring the forces that act between the AFM probe of a force sensing cantilever and a sample. In this work, the lateral force calibrations of a V-shaped cantilever were performed using the finite element method, multiple pivot loading, and thermal noise methods. As a result, it was shown that the multiple pivot loading method was appropriate for the lateral force calibration of a V-shaped cantilever. Further, through crosschecking of the abovementioned methods, it was concluded that the thermal noise method could be used for determining the lateral spring constants as long as the lateral deflection sensitivity was accurately determined. To obtain the lateral deflection sensitivity from the sticking portion of the friction loop, the contact stiffness should be taken into account.
참고문헌 (20)
- Binnig, G., "Atomic Force Microscope", Phys Rev Lett, 56, pp.930~933 (1986)
- 정구현, "Nanomechanical Properties of Thin Films of Type I Collagen Fibrils", LANGMUIR, 26(5), pp.3629~3636, AMER CHEMICAL SOC
- Giesbers, A. J. M., "Nanolithography and Manipulation of Graphene using an Atomic Force Microscope", Solid State Commun, 147, pp.366~369 (2008)
- Bordag, M., "Shear Stress Measurements on InAs Nanowires by AFM Manipulation", Small, 3, pp.1398~1401 (2007)
- Bharat Bhushan, "Nanotribology and Nanomechanics of AFM Probe-Based Data Recording Technology", J Phys: Condens Matter, 20, pp.365207~ (2008)
- Hutter, J. L., "Calibration of Atomic-Force Microscope Tips", Rev. Sci. Instrum., 64, pp.1868~1873 (1993)
- Cleveland, J. P., "A Nondestructive Method for Determining the Spring Constant of Cantilevers for Scanning Force Microscopy", Rev. Sci. Instrum., 64, pp.403~405 (1993)
- Sader, J. E., "Calibration of Rectangular Atomic Force Microscope Cantilevers", Rev. Sci. Instrum., 70, pp.3967~3969 (1999)
- Torii, A., "A Method for Determining the Spring Constant of Cantilevers for Atomic Force Microscopy", Meas Sci Technol, 7, pp.179~184 (1996)
- Chung, K. H., "Accurate Noncontact Calibration of Colloidal Probe Sensitivities in Atomic Force Microscopy", Rev. Sci. Instrum., 80, pp.065107~ (2009)
- Ogletree, D. F., "Calibration of Frictional Forces in Atomic ForceMicroscopy", Rev. Sci. Instrum., 67, pp.3298~3306 (1996)
- Varenberg, M., "An Improved Wedge Calibration Method for Lateral Force in Atomic Force Microscopy", Rev. Sci. Instrum., 74, pp.3362~3367 (2003)
- Wang, F., "Effect of Contact Stiffness on Wedge Calibration of Lateral Force in Atomic Force Microscopy", Rev. Sci. Instrum., 78, pp.043701~ (2007)
- Li, Q., "Lateral Force Calibration of an Atomic Force Microscope with a Diamagnetic Levitation Spring System", Rev. Sci. Instrum., 77, pp.065105~ (2006)
- Bogdanovic, G., "Tip Friction — Torsional Spring Constant Determination", Colloid Surf B-Biointerfaces, 19, pp.397~405 (2000)
- Chung, K. H., "Note: Lateral Force Microscope Calibration using Multiple Location Pivot Loading of Rectangular Cantilevers", Rev. Sci. Instrum., 81, pp.026104~ (2010)
- Neumeister, J. M., "Lateral, Normal, and Longitudinal Spring Constants of Atomic Force Microscopy Cantilevers", Rev. Sci. Instrum., 65, pp.2527~2531 (1994)
- Higgins, M. J., "Noninvasive Determination of Optical Lever Sensitivity in Atomic Force Microscopy", Rev. Sci. Instrum., 77, pp.013701~ (2006)
- Wagner, K., "Noncontact Method for Calibration of Lateral Forces in Scanning Force Microscopy", Langmuir, 27, pp.4635~4644 (2011)
- Cain, R. G., "Force Calibration in Lateral Force Microscopy", J. Colloid Interface Sci., 227, pp.55~65 (2000)