Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants
1
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
PURPOSE. The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants. MATERIALS AND METHODS. Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group): EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests. RESULTS. The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the postloading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05). CONCLUSION. Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.
Abstract
PURPOSE. The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants. MATERIALS AND METHODS. Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group): EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests. RESULTS. The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the postloading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05). CONCLUSION. Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.
참고문헌 (30)
- Cibirka RM, "Examination of the implant-abutment interface after fatigue testing", J Prosthet Dent, 85, pp.268~275 (2001)
- Jörnéus L, "Loads and design of screw joints for single crowns supported by osseointegrated implants", Int J Oral Maxillofac Implants, 7, pp.353~359 (1992)
- Bianco G, "Osseointegrated implant for single tooth replacement: A retrospective multicenter study on routine use in private practice", Clin lmplant Dent Relat Res, 2, pp.152~158 (2000)
- Theoharidou A, "Abutment screw loosening in single-implant restorations: a systematic review", Int J Oral Maxillofac Implants, 23, pp.681~690 (2008)
- Richter EJ, "In vivo horizontal bending moments on implants", Int J Oral Maxillofac Implants, 13, pp.232~244 (1998)
- Bickford JH, "An introduction to the design and behavior of bolted joints", pp.515~564 (1995), Marcel Dekker
- Tsuge T, "Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants", Dent Mater J, 28, pp.373~381 (2009)
- Delben JA, "Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws", Int J Oral Maxillofac Implants, 26, pp.251~256 (2011)
- Stüker RA, "Preload and torque removal evaluation of three different abutment screws for single standing implant restorations", J Appl Oral Sci, 16, pp.55~58 (2008)
- Khraisat A, "Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system", J Prosthet Dent, 91, pp.326~334 (2004)
- Shigley JE, "Mechanical engineering design", pp.240~245 (1977), McGraw Hill
- Weiss EI, "Effect of repeated closures on opening torque values in seven abutment-implant systems", J Prosthet Dent, 84, pp.194~199 (2000)
- Delben JA, "Influence of abutment type and esthetic veneering on preload maintenance of abutment screw of implant-supported crowns", J Prosthodont, 23, pp.134~139 (2014)
- Khraisat A, "Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading", Clin Implant Dent Relat Res, 6, pp.157~164 (2004)
- Al Jabbari YS, "Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws", J Prosthodont, 17, pp.168~180 (2008)
- 신현모, "Influence of the implant-abutment connection design and diameter on the screw joint stability", The Journal of Advanced of Prosthodontics, 6(2), pp.126~132 (2014), 대한치과보철학회[DOI]
- Gracis S, "Internal vs. external connections for abutments", reconstructions: A systematic review, 2012, pp.23, 202~216
- Feitosa PC, "Stability of external and internal implant connections after a fatigue test", Eur J Dent, 7, pp.267~271 (2013)
- Kim KS, "Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading", Int J Oral Maxillofac Implants, 29, pp.1079~1084 (2014)
- Binon PP, "Implants and components: entering the new millennium", Int J Oral Maxillofac Implants, 15, pp.76~94 (2000)
- Sopwith DG, "The distribution of load in screw threads", Proc Inst Mech Eng, 159, pp.373~383 (1948)
- Kim BJ, "The effect of screw length on fracture load and abutment strain in dental implants with external abutment connections", Int J Oral Maxillofac Implants, 27, pp.820~823 (2012)
- 최진호, "치과용 임플랜트에서 지대주 나사의 길이 및 반복 조임 횟수가 지내주 나사의 풀림에 미치는 영향", 대한치과보철학회지, 44(4), pp.432~442 (2006), 대한치과보철학회
- Yeo IS, "The effect of abutment screw length on screw loosening in dental implants with external abutment connections after thermocycling", Int J Oral Maxillofac Implants, 29, pp.59~62 (2014)
- Park SD, "Microleakage of different sealing materials in access holes of internal connection implant systems", J Prosthet Dent, 108, pp.173~180 (2012)
- Winkler S, "screw mechanics and the settling effect: Overview", J Oral Implantol, 29, pp.242~245 (2003)
- Graf H., "Bruxism", Dent Clin North Am, 13, pp.659~665 (1969)
- Haack JE, "Elongation and preload stress in dental implant abutment screws", Int J Oral Maxillofac Implants, 10, pp.529~536 (1995)
- Maeda, Y, "In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication", J Oral Rehabil, 33, pp.75~78 (2006)
- 김은숙, "Influence of the implant abutment types and the dynamic loading on initial screw loosening", The Journal of Advanced of Prosthodontics, 5(1), pp.21~28 (2013), 대한치과보철학회[DOI]