Effect of repetitive pecking at working length for glide path preparation using G-file
2
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Objectives: Glide path preparation is recommended to reduce torsional failure of nickeltitanium (NiTi) rotary instruments and to prevent root canal transportation. This study evaluated whether the repetitive insertions of G-files to the working length maintain the apical size as well as provide sufficient lumen as a glide path for subsequent instrumentation. Materials and Methods: The G-file system (Micro-Mega) composed of G1 and G2 files for glide path preparation was used with the J-shaped, simulated resin canals. After inserting a G1 file twice, a G2 file was inserted to the working length 1, 4, 7, or 10 times for four each experimental group, respectively (n = 10). Then the canals were cleaned by copious irrigation, and lubricated with a separating gel medium. Canal replicas were made using silicone impression material, and the diameter of the replicas was measured at working length (D0) and 1 mm level (D1) under a scanning electron microscope. Data was analysed by one-way ANOVA and post-hoc tests (p = 0.05). Results: The diameter at D0 level did not show any significant difference between the 1, 2, 4, and 10 times of repetitive pecking insertions of G2 files at working length. However, 10 times of pecking motion with G2 file resulted in significantly larger canal diameter at D1 (p < 0.05). Conclusions: Under the limitations of this study, the repetitive insertion of a G2 file up to 10 times at working length created an adequate lumen for subsequent apical shaping with other rotary files bigger than International Organization for Standardization (ISO) size 20, without apical transportation at D0 level.
Abstract
Objectives: Glide path preparation is recommended to reduce torsional failure of nickeltitanium (NiTi) rotary instruments and to prevent root canal transportation. This study evaluated whether the repetitive insertions of G-files to the working length maintain the apical size as well as provide sufficient lumen as a glide path for subsequent instrumentation. Materials and Methods: The G-file system (Micro-Mega) composed of G1 and G2 files for glide path preparation was used with the J-shaped, simulated resin canals. After inserting a G1 file twice, a G2 file was inserted to the working length 1, 4, 7, or 10 times for four each experimental group, respectively (n = 10). Then the canals were cleaned by copious irrigation, and lubricated with a separating gel medium. Canal replicas were made using silicone impression material, and the diameter of the replicas was measured at working length (D0) and 1 mm level (D1) under a scanning electron microscope. Data was analysed by one-way ANOVA and post-hoc tests (p = 0.05). Results: The diameter at D0 level did not show any significant difference between the 1, 2, 4, and 10 times of repetitive pecking insertions of G2 files at working length. However, 10 times of pecking motion with G2 file resulted in significantly larger canal diameter at D1 (p < 0.05). Conclusions: Under the limitations of this study, the repetitive insertion of a G2 file up to 10 times at working length created an adequate lumen for subsequent apical shaping with other rotary files bigger than International Organization for Standardization (ISO) size 20, without apical transportation at D0 level.
참고문헌 (21)
- Sonntag D, "Root canal shaping with manual stainless steel files and rotary Ni-Ti files performed by students", Int Endod J, 36, pp.246~255 (2003)
- Sattapan B, "Defects in rotary nickel-titanium files after clinical use", J Endod, 26, pp.161~165 (2000)
- Alapati SB, "SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical Use", J Endod, 31, pp.40~43 (2005)
- Berutti E, "Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments", J Endod, 30, pp.228~230 (2004)
- Patiño PV, "The influence of a manual glide path on the separation rate of NiTi rotary instruments", J Endod, 31, pp.114~116 (2005)
- 하정홍, "Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals", Restorative Dentistry and Endodontics, 37(4), pp.215~219 (2012), 대한치과보존학회[DOI]
- Ruddle CJ, "The ProTaper technique", Endod Top, 10, pp.187~190 (2005)
- West J, "Endodontic update 2006", J Esthet Restor Dent, 18, pp.280~300 (2006)
- Ruddle CJ, "Pathway of the pulp", pp.231~292 (2002), Mosby Elsevier
- Bahcall JK, "The causes, prevention, and clinical management of broken endodontic rotary files", Dent Today, 24, pp.74~80 (2005)
- Lopes HP, "Mechanical behavior of pathfinding endodontic instruments", J Endod, 38, pp.1417~1421 (2012)
- Yun HH, "A comparison of the shaping abilities of 4 nickel-titanium rotary instruments in simulated root canals", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 95, pp.228~233 (2003)
- Park H, "A comparison of Greater Taper files, ProFiles, and stainless steel files to shape curved root canals", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 91, pp.715~718 (2001)
- Arens FC, "Evaluation of single-use rotary nickel titanium instruments", J Endod, 29, pp.664~666 (2003)
- Cheung GSP, "Instrument fracture: mechanisms, removal of fragments, and clinical outcomes", Endod Top, 16, pp.1~26 (2007)
- West JD, "The endodontic Glidepath: “Secrets to rotary safety”", Dent Today, 29, pp.86~93 (2010)
- Blum JY, "Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient", J Endod, 29, pp.567~575 (2003)
- Peters OA, "ProTaper rotary root canal preparation: effects of canal anatomy on final shape analysed by micro CT", Int Endod J, 36, pp.86~92 (2003)
- Berutti E, "Use of nickel-titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals", J Endod, 35, pp.408~412 (2009)
- Wu MK, "Prevalence and extent of long oval canals in the apical third", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 89, pp.739~743 (2000)
- Bürklein S, "Incidence of dentinal defects after root canal preparation: reciprocating versus rotary instrumentation", J Endod, 39, pp.501~504 (2013)