Establishment of ISO 4037-1 Narrow-Spectrum Series X-rays for an Eye Dosimeter Performance Test in KHNP CRI

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

The International Commission on Radiological Protection (ICRP) recommended loweringoccupational dose limit for the eyes from 150 mSv in a year to average 20 mSv over five years (or 50mSv maximum in a year). The revised eye dose limit is same with the effective dose limit of workers. Assessment of eye dose could be conducted by wearing an eye dosimeter near the eyes. To obtainthe reliability of the eye dose assessment result evaluated by eye dosimeters, the dosimeters mustmeet the performance criteria. In this study, we established five X-rays according to the InternationalOrganization for Standardization to conduct performance tests of eye dosimeters. The half value layer(HVL) measurement method was applied to verify the equivalent filtration effect of the establishedradiation fields. Although the measured HVLs of the five X-rays had deviations, from 0.004 to 0.11 mm,with reference X-rays but satisfied the criteria.

Abstract

The International Commission on Radiological Protection (ICRP) recommended loweringoccupational dose limit for the eyes from 150 mSv in a year to average 20 mSv over five years (or 50mSv maximum in a year). The revised eye dose limit is same with the effective dose limit of workers. Assessment of eye dose could be conducted by wearing an eye dosimeter near the eyes. To obtainthe reliability of the eye dose assessment result evaluated by eye dosimeters, the dosimeters mustmeet the performance criteria. In this study, we established five X-rays according to the InternationalOrganization for Standardization to conduct performance tests of eye dosimeters. The half value layer(HVL) measurement method was applied to verify the equivalent filtration effect of the establishedradiation fields. Although the measured HVLs of the five X-rays had deviations, from 0.004 to 0.11 mm,with reference X-rays but satisfied the criteria.

참고문헌 (9)

  1. ICRP, "ICRP statement of tissue reactions and early and late effects of radiation in normal tissues and organs - Threshold doses for tissue reactions in a radiation protection context", pp.118~ (2012), ICRP
  2. ICRP, "1990 Recommendations of the International Commission on Radiological Protection", pp.60~ (1991), ICRP
  3. 공태영, "A preliminary evaluation of the implementation of a radiation protection program for the lens of the eye in Korean nuclear power plants", Nuclear Engineering and Technology, 53(9), pp.3035~3043 (2021), 한국원자력학회[DOI]
  4. IEC, "Radiation protection instrumentation - Dosimetry systems with integrating passive detectors for individual, workplace and environmental monitoring of photon and beta radiation", pp.62387~ (2020), IEC
  5. Development of Beta and X reference radiation field and calibration procedure for personal dosimeter calibration, "KHNP", ~, pp.2014-50003339-DAN-0273TR (2014)
  6. ISO, "Radiological protection-X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy - Part 1: Radiation characteristics and production methods, ISO 4037-1", ~ (2019)
  7. YXLON, "Y.MG325", 452 Universal bipolar constant potential X-ray system. YXLON. Products
  8. YXLON, "YXLON.TU 320-D03 Bipolar metal-ceramic X-ray tube. YXLON. Technical data", ~
  9. Standard Imaging, Inc, "User manual" (2005)

관련 논문AI추천

이 간행물의 다른 논문 (12건)