Measurement uncertainty analysis of radiophotoluminescent glass dosimeter reader system based on GD-352M for estimation of protection quantity

1

조회

0

다운로드

4

인용

0

참고문헌

키워드

주제분류원자력공학

초록

At the Korea Institute of Radiological and Medical Sciences, physical human phantoms were developed toevaluate various radiation protection quantities, based on the mesh-type reference computationalphantoms of the International Commission on Radiological Protection. The physical human phantomswere fabricated such that a radiophotoluminescent glass dosimeter (RPLGD) with a Tin filter, namely GD352M, could be inserted into them. A Tin filter is used to eliminate the overestimated signals in lowenergy photons below 100 keV. The measurement uncertainty of the RPLGD reader system based onGD-352M should be analyzed for obtaining reliable protection quantities before using it for practicalapplications. Generally, the measurement uncertainty of RPLGD systems without Tin filters is analyzedfor quality assurance of radiotherapy units using a high-energy photon beam. However, in this study, themeasurement uncertainty of GD-352M was analyzed for evaluating the protection quantities. Themeasurement uncertainty factors in the RPLGD include the reference irradiation, regression curve,reproducibility, uniformity, energy dependence, and angular dependence, as described by the International Organization for Standardization (ISO). These factors were calculated using the Guide to theExpression of Uncertainty in Measurement method, applying ISO/ASTM standards 51261(2013),51707(2015), and SS-ISO 22127(2019). The measurement uncertainties of the RPLGD reader system witha coverage factor of k ¼ 2 were calculated to be 9.26% from 0.005 to 1 Gy and 8.16% from 1 to 10 Gy. Ablind test was conducted to validate the RPLGD reader system, which demonstrated that the readoutdoses included blind doses of 0.1, 1, 2, and 5 Gy. Overall, the En values were considered satisfactory

Abstract

At the Korea Institute of Radiological and Medical Sciences, physical human phantoms were developed toevaluate various radiation protection quantities, based on the mesh-type reference computationalphantoms of the International Commission on Radiological Protection. The physical human phantomswere fabricated such that a radiophotoluminescent glass dosimeter (RPLGD) with a Tin filter, namely GD352M, could be inserted into them. A Tin filter is used to eliminate the overestimated signals in lowenergy photons below 100 keV. The measurement uncertainty of the RPLGD reader system based onGD-352M should be analyzed for obtaining reliable protection quantities before using it for practicalapplications. Generally, the measurement uncertainty of RPLGD systems without Tin filters is analyzedfor quality assurance of radiotherapy units using a high-energy photon beam. However, in this study, themeasurement uncertainty of GD-352M was analyzed for evaluating the protection quantities. Themeasurement uncertainty factors in the RPLGD include the reference irradiation, regression curve,reproducibility, uniformity, energy dependence, and angular dependence, as described by the International Organization for Standardization (ISO). These factors were calculated using the Guide to theExpression of Uncertainty in Measurement method, applying ISO/ASTM standards 51261(2013),51707(2015), and SS-ISO 22127(2019). The measurement uncertainties of the RPLGD reader system witha coverage factor of k ¼ 2 were calculated to be 9.26% from 0.005 to 1 Gy and 8.16% from 1 to 10 Gy. Ablind test was conducted to validate the RPLGD reader system, which demonstrated that the readoutdoses included blind doses of 0.1, 1, 2, and 5 Gy. Overall, the En values were considered satisfactory

참고문헌 (28)

  1. D. Borrego, "Occupational doses to medical staffperforming or assisting with fluoroscopically guided interventional procedures", Radiology, 294, pp.353~359 (2020)[DOI]
  2. T. Kairn, "In vivo monitoring of total skin electron dose using optically stimulated luminescence dosimeters", Rep. Practical Oncol. Radiother, 25, pp.35~40 (2020)[DOI]
  3. J. D. Cho, "Flexible film dosimeter for in vivo dosimetry", Med. Phys, 47, pp.3204~3213 (2020)[DOI]
  4. J. F. Winslow, "Construction of anthrompomorphic phantoms for use in dosimetry studies", J. Appl. Clin. Med. Phys, 10, pp.195~204 (2009)[DOI]
  5. S. Hashim, "Evaluation of organ doses and specific k effective dose of 64-slice CT thorax examination using an adult anthropomorphic phantom", Radiat. Phys. Chem, 126, pp.14~20 (2016)[DOI]
  6. S. Tongkum, "Determination of radiation dose and low-dose protocol for digital chest tomosynthesis using radiophotoluminescent(RPL)glass dosimeters", Phys. Med, 73, pp.13~21 (2020)[DOI]
  7. P. Wołowiec, "The analysis of the measurement uncertainty with application of small detectors made of Gafchromic EBT3 films for the range of doses typical for in vivo dosimetry in teleradiotherapy", Radiat. Meas, 92, pp.72~79 (2016)[DOI]
  8. S. T. Chiu-Tsao, "Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies", Med. Phys, 32, pp.3350~3354 (2005)[DOI]
  9. Guerda Massillon-JL, "Energy Dependence of the New Gafchromic EBT3 Film:Dose Response Curves for 50 KV, 6 and 15 MV X-Ray Beams", International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 01(02), pp.60~65 (2012), Scientific Research Publishing, Inc.[DOI]
  10. E. Y. Marroquin, "Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density", J. Appl. Clin. Med. Phys, 17, pp.466~481 (2016)[DOI]
  11. K. Venables, "The use of in vivo thermoluminescent dosimeters in the quality assurance programme for the START breast fractionation trial", Radiother. Oncol, 71, pp.303~310 (2004)[DOI]
  12. A. M. Costa, "In vivo dosimetry with thermoluminescent dosimeters in external photon beam radiotherapy", Appl. Radiat. Isot, 68, pp.760~762 (2010)[DOI]
  13. P. E. Wesolowska, "Characterization of three solid state dosimetry systems for use in high energy photon dosimetry audits in radiotherapy", Radiat. Meas, 106, pp.556~562 (2017)[DOI]
  14. L. Duggan, "The light sensitivity of thermoluminescent materials: LiF:Mg,Cu,P, LiF:Mg,Ti and Al2O3:C", Radiation Measurements, 32(4), pp.335~342 (2000), Elsevier BV[DOI]
  15. K. Son, "Evaluation of the dosimetric characteristics of a radiophotoluminescent glass dosimeter for high-energy photon and electron beams in the field of radiotherapy", Radiat. Meas, 46, pp.1117~1122 (2011)[DOI]
  16. J. E. Rah, "Feasibility study of radiophotoluminescent glass rod dosimeter postal dose intercomparison for high energy photon beam", Appl. Radiat. Isot, 67, pp.324~328 (2009)[DOI]
  17. F. Araki, "Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife", Med. Phys, 31, pp.1980~1986 (2004)[DOI]
  18. P. Oonsiri, "Investigation of the dosimetric characteristics of radiophotoluminescent glass dosimeter for high-energy photon beams", J. Radiat. Res. Appl. Sci, 12, pp.65~71 (2019)[DOI]
  19. S. -M. Hsu, "Synthesis and physical characteristics of radiophotoluminescent glass dosimeters", Radiat. Meas, 42, pp.621~624 (2007)[DOI]
  20. International Organization for Standardization, "Uncertainty of Measurement. Guide Expression of Uncertainty in Measurement (GUM:1995)", ISO, 2008
  21. S. M. Hsu, "Clinical application of radiophotoluminescent glass dosimeter for dose verification of prostate HDR procedure", Med. Phys, 35, pp.5558~5564 (2008)[DOI]
  22. International Organization for Standardization(ISO), "Absorbed Dose Determination in External Beam Radiotherapy: an International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, vol. 398", IAEA TRS, ~ (2006)
  23. International Organization for Standardization(ISO), "Practice for Calibration of Routine Dosimetry Systems for Radiation Processing, vol. 51261", ISO, 2013
  24. International Organization for Standardization(ISO), "Guide for Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing, vol. 51707", ISO, 2015
  25. International Organization for Standardization(ISO), "Dosimetry with Radiophotoluminescent Glass Dosimeters for Dosimetry Audit in MV X-Ray Radiotherapy", SS, 2019
  26. Y. Hashimoto, "Energy dependence of a radiophotoluminescent glass dosimeter for HDR 192Ir brachytherapy source", Med. Phys, 46, pp.964~972 (2019)[DOI]
  27. Seung Kyu Lee, "Reference X-ray Irradiation System for Personal Dosimeter Testing and Calibration of Radiation Detector", 방사선방어학회지, 44(2), pp.72~78 (2019), 테스트학회[DOI]
  28. Conformity Assessment -General Requirements for Proficiency Testing, "International Organization for Standardization(ISO)", ISO, 2010, ISO[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (21건)