Performance of the cobas u 701 Analyzer in Urinary Tract Infection Screening

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Negative urine cultures to rule out urinary tract infections (UTI) generate a considerable laboratory workload; thus, a rapid screening test is desirable. We evaluated the performance of a new automated microscopy analyzer, cobas u 701 (Roche Diagnostics International, Rotkreuz, Switzerland) for the screening of UTI, and developed a rule-out strategy to reduce the number of samples requiring culture. We also assessed squamous epithelial cell (SEC) count as a predictor of culture contamination. Methods: In total, 1,604 urine samples from outpatients were analyzed with cobas u 701 and culture. Bacterial (BAC) and white blood cell (WBC) counts were used for sample interpretation. To determine a useful cut-off point to predict negative cultures, we selected the highest sensitivity and specificity values obtained from ROC curves. Diagnostic accuracy by age and gender was evaluated. Results: Urine culture showed growth of ≥104 colony forming units (CFU)/mL in 256 samples (16.0%). The highest sensitivity (91.8%) and specificity (68.4%) were obtained for cut-off points of 119 BAC/μL and 22 WBC/μL. The combination of BAC and WBC improved the performance of the rule-out strategy with a low rate of false-negative results (1.5%) and a high negative predictive value (NPV, 97.3%). Fifty-seven percent of the samples would not have required culture. SEC count was a poor predictor of culture contamination. Conclusions: cobas u 701 can substantially reduce the number of urine samples requiring culture, with a low false-negative rate and a high NPV.

Abstract

Background: Negative urine cultures to rule out urinary tract infections (UTI) generate a considerable laboratory workload; thus, a rapid screening test is desirable. We evaluated the performance of a new automated microscopy analyzer, cobas u 701 (Roche Diagnostics International, Rotkreuz, Switzerland) for the screening of UTI, and developed a rule-out strategy to reduce the number of samples requiring culture. We also assessed squamous epithelial cell (SEC) count as a predictor of culture contamination. Methods: In total, 1,604 urine samples from outpatients were analyzed with cobas u 701 and culture. Bacterial (BAC) and white blood cell (WBC) counts were used for sample interpretation. To determine a useful cut-off point to predict negative cultures, we selected the highest sensitivity and specificity values obtained from ROC curves. Diagnostic accuracy by age and gender was evaluated. Results: Urine culture showed growth of ≥104 colony forming units (CFU)/mL in 256 samples (16.0%). The highest sensitivity (91.8%) and specificity (68.4%) were obtained for cut-off points of 119 BAC/μL and 22 WBC/μL. The combination of BAC and WBC improved the performance of the rule-out strategy with a low rate of false-negative results (1.5%) and a high negative predictive value (NPV, 97.3%). Fifty-seven percent of the samples would not have required culture. SEC count was a poor predictor of culture contamination. Conclusions: cobas u 701 can substantially reduce the number of urine samples requiring culture, with a low false-negative rate and a high NPV.

참고문헌 (20)

  1. Stalenhoef JE, "Febrile urinary tract infection in the emergency room", Curr Opin Infect Dis, 28, pp.106~111 (2015)
  2. Okada H, "Detection of significant bacteriuria by automated urinalysis using flow cytometry", J Clin Microbiol, 38, pp.2870~2872 (2000)
  3. Tessari A, "Screening of presumptive urinary tract infections by the automated urine sediment analyser sediMAX", Clin Chem Lab Med, 53(S2), pp.s1503~s1508 (2015)
  4. Brilha S, "Use of flow cytometry(Sysmex UF-100)to screen for positive urine cultures : in search for the ideal cut-off", Clin Chem Lab Med, 48, pp.289~292 (2010)
  5. Evans R, "Testing by Sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory : a comparison", J Clin Pathol, 59, pp.661~662 (2006)
  6. Zaman Z, "Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures", J Clin Microbiol, 39, pp.4169~4171 (2001)
  7. Jolkkonen S, "Screening of urine samples by flow cytometry reduces the need for culture", J Clin Microbiol, 48, pp.3117~3121 (2010)
  8. Matthijs Oyaert, M.Sc.,, "Progress in Automated Urinalysis", Annals of Laboratory Medicine, 39(1), pp.15~22 (2019), 대한진단검사의학회[DOI]
  9. Falbo R, "Bacteriuria screening by automated whole-field-image-based microscopy reduces the number of urine cultures", J Clin Microbiol, 50, pp.1427~1429 (2012)
  10. Martinez MH, "UriSed as a screening tool for presumptive diagnosis of urinary tract infection", Clin Chim Acta, 425, pp.77~79 (2013)
  11. Lee W, "Comparison of the automated cobas u 701 urine microscopy and UF-1000i flow cytometry systems and manual microscopy in the examination of urine sediments", J Clin Lab Anal, 30, pp.663~671 (2016)
  12. Kim SH, "Evaluation of the Cobas u 701 microscopy analyser compared with urine culture in screening for urinary tract infection", J Med Microbiol, 66, pp.1110~1113 (2017)
  13. Broeren MA, "Screening for urinary tract infection with the SysmexUF-1000i urine flow cytometer", J Clin Microbiol, 49, pp.1025~1029 (2011)
  14. Stürenburg E, "Detection of significant bacteriuria by use of the iQ200 automated urine microscope", J Clin Microbiol, 52, pp.2855~2860 (2014)
  15. 김현지, "Age-Specific Cutoffs of the Sysmex UF-1000i Automated Urine Analyzer for Rapid Screening of Urinary Tract Infections in Outpatients", Annals of Laboratory Medicine, 39(3), pp.322~326 (2019), 대한진단검사의학회[DOI]
  16. Íñigo M, "Evaluation of the SediMax automated microscopy sediment analyzer and the Sysmex UF-1000i flow cytometer as screening tools to rule out negative urinary tract infections", Clin Chim Acta, 456, pp.31~35 (2016)
  17. De Rosa R, "Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection", Clin Chim Acta, 411, pp.1137~1142 (2010)
  18. Manoni F, "Urine particle evaluation : a comparison between the UF-1000i and quantitative microscopy", Clin Chem Lab Med, 48, pp.1107~1111 (2010)
  19. Mohr NM, "Urinary squamous epithelial cells do not accurately predict urine culture contamination, but may predict urinalysis performance in predicting bacteriuria", Acad Emerg Med, 23, pp.323~330 (2016)
  20. Walter FG, "Squamous cells as predictors of bacterial contamination in urine samples", Ann Emerg Med, 31, pp.455~458 (1998)

관련 논문AI추천

이 간행물의 다른 논문 (14건)