초록

Electrostatic capacitance measurement method in a fine hose was proposed, in which two ring-type electrodes were disposed on the hose in the direction of fluid flow instead of the conventional face-to-face electrodes. With the proposed electrode structure, we realized a Ringer’s solution exhaustion detector for an IV(invasive vein) injection set. On a 4 mm- diameter hose of IV set, we disposed two ring-type electrodes of 10 mm width at a distance of 5 mm each other and obtained 0.72 pF and 2.51 pF for air and 10 % dextrose Ringer’s solution in the hose, respectively. The capacitance between the two electrodes varied with the hose-wraparound coverage of electrode as well as the width of electrode and the distance between the electrodes. For hose-wraparound electrode coverage of 75 %, the capacitance varied from 0.62 pF to 1.98 pF with the Ringer’s solution level between the two electrodes. A charge amplifier converted the capacitance variation into electric signal and a comparator was used to detect whether Ringer's solution was exhausted or not. The result was delivered to a host using a RF transmitter with 320 MHz carrier frequency.

Abstract

Electrostatic capacitance measurement method in a fine hose was proposed, in which two ring-type electrodes were disposed on the hose in the direction of fluid flow instead of the conventional face-to-face electrodes. With the proposed electrode structure, we realized a Ringer’s solution exhaustion detector for an IV(invasive vein) injection set. On a 4 mm- diameter hose of IV set, we disposed two ring-type electrodes of 10 mm width at a distance of 5 mm each other and obtained 0.72 pF and 2.51 pF for air and 10 % dextrose Ringer’s solution in the hose, respectively. The capacitance between the two electrodes varied with the hose-wraparound coverage of electrode as well as the width of electrode and the distance between the electrodes. For hose-wraparound electrode coverage of 75 %, the capacitance varied from 0.62 pF to 1.98 pF with the Ringer’s solution level between the two electrodes. A charge amplifier converted the capacitance variation into electric signal and a comparator was used to detect whether Ringer's solution was exhausted or not. The result was delivered to a host using a RF transmitter with 320 MHz carrier frequency.

참고문헌 (12)

  1. 김창욱, "링거 자동 교체를 위한 잔량 검출 알고리즘", 한국산업정보학회논문지, 13(1), pp.30~36 (2008), 한국산업정보학회
  2. 하관용, "마이크로프로세 서를 이용한 병원용 환자 링거액 잔류유량 원격 실 시간 검사 시스템 개발", 정보및제어학술대회 논문 집, pp.279~282 (2005)
  3. イプ フザンデイエ-ル, "液流制御裝置" (1977)
  4. 강필식, "링거액 감지기", ~ (2003)
  5. 강상욱, "정맥 주사용 링거액 자동 유량 조절장치", ~ (2006)
  6. 강주완, "영상신호처리를 이용한 링 거액 유량측정시스템", ~ (2006)
  7. 김성길, "로드셀을 이용한 링거액 잔량 감지 및 통 보장치", ~ (2006)
  8. 이혁수, "비유전율을 이용한 공기감지장치에 관한 연구", 센서학회지, 16(5), pp.384~388 (2007), 한국센서학회
  9. 정도운, "로드셀을 이용한 요류검사기의 구현 및 평가", 센서학회지, 13(6), pp.436~445 (2004), 한국센서학회
  10. 김종명, "광센서를 이용한 실시간 중환자 요량감시 장치", 센서학회지, 17(1), pp.81~85 (2008), 한국센서학회
  11. 김경아, "전립선 진단을 위한 수압 측정 방식의 새로운 요 유량 계측기법", 센서학회지, 16(3), pp.179~186 (2007), 한국센서학회

관련 논문AI추천

이 간행물의 다른 논문 (9건)