링거액 소진 감지를 위한 정전용량방식의 차동센서 설계 및 제작

Design & implementation of differential sensor using electrostatic capacitance method for detecting Ringer's solution exhaustion

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류공학일반

초록

This paper proposes a differential structure sensor for detecting Ringer's solution exhaustion, in which three C-type electrodes of 10 mm width are disposed on a ringer hose at a distance of 5 mm each other in the direction of Ringer's solution flow. In the center of middle electrode, two capacitances are formed at the proposed sensor. When ringer hose is filled with Ringer's solution, there is no difference between two capacitances. But capacitance difference exist under the Ringer's solution shortage, because the shortage causes the hose filled with air from the top position electrode. The capacitance difference got to maximum 1.81 pF, when air was filled between top and middle electrode and the last of hose was filled with 10 % dextrose injection Ringer's solution. The capacitance difference varied with hose-wraparound coverage of electrodes as well as the width of them. For hose-wraparound electrode coverage of 90 % and 70 %, the maximum capacitance difference was 1.81 pF and 1.56 pF, respectively. A differential charge amplifier converted the capacitance difference to electric signal, and minimized electrodes' adhering problem and external noise coupling problem.

Abstract

This paper proposes a differential structure sensor for detecting Ringer's solution exhaustion, in which three C-type electrodes of 10 mm width are disposed on a ringer hose at a distance of 5 mm each other in the direction of Ringer's solution flow. In the center of middle electrode, two capacitances are formed at the proposed sensor. When ringer hose is filled with Ringer's solution, there is no difference between two capacitances. But capacitance difference exist under the Ringer's solution shortage, because the shortage causes the hose filled with air from the top position electrode. The capacitance difference got to maximum 1.81 pF, when air was filled between top and middle electrode and the last of hose was filled with 10 % dextrose injection Ringer's solution. The capacitance difference varied with hose-wraparound coverage of electrodes as well as the width of them. For hose-wraparound electrode coverage of 90 % and 70 %, the maximum capacitance difference was 1.81 pF and 1.56 pF, respectively. A differential charge amplifier converted the capacitance difference to electric signal, and minimized electrodes' adhering problem and external noise coupling problem.

참고문헌 (13)

  1. 이혁수, "비유전율을 이용한 공기감지장치에 관한 연구", 센서학회지, 16(5), pp.384~388 (2007), 한국센서학회
  2. イプ フザンデイエ-ル, "液流制御裝置" (1977)
  3. 김청월, "정전용량변화를 이용한 링거액소진감지장치의 구현", 센서학회지, 19(1), pp.1~7 (2010), 한국센서학회
  4. 강필식, "링거액 감지기", ~ (2004)
  5. 김종명, "광센서를 이용한 실시간 중환자 요량감시 장치", 센서학회지, 17(1), pp.81~85 (2008), 한국센서학회
  6. 정도운, "로드셀을 이용한 요류검사기의 구현 및 평가", 센서학회지, 13(6), pp.436~445 (2004), 한국센서학회
  7. 강주완, "영상신호처리를 이용한 링 거액 유량측정시스템", ~ (2006)
  8. 김창욱, "링거 자동 교체를 위한 잔량 검출 알고리즘", 한국산업정보학회논문지, 13(1), pp.30~36 (2008), 한국산업정보학회
  9. 김성길, "로드셀을 이용한 링거액 잔량 감지 및 통 보장치", ~ (2005)
  10. 전영탁, "링거액 경보 장치", ~ (2006)
  11. 강상욱, "링거액 소진 경보 장치", ~ (2008)
  12. 장종범, "링거액의 잔류여부 감지 장치", ~ (2002)
  13. Qun Wei, "Capacitive sensor for the detection of residual quantity of intravenous drip solution in a plastic intravenous bag", 센서학회지, 19(4), pp.271~277 (2010), 한국센서학회