Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: EDTA-induced pseudothrombocytopenia (PTCP) during whole blood collection requires significant laboratory resources to obtain accurate results. We evaluated platelet-deaggregation function in EDTA-induced PTCP and platelet-clump flagging by the BC-6800Plus hematology analyzer using integrated digital image analysis. Methods: We prospectively collected 132 whole blood samples suspected of platelet clumping (102 in EDTA and 30 in sodium citrate) from 88 individuals. We compared platelet counts determined using the platelet count by impedance (PLT-I) function of the DxH 900 hematology analyzer and the PLT-I or optical platelet count (PLT-O) function of the BC- 6800Plus. Platelet clumping was verified through manual inspection and the MC-80 digital image analyzer. Results: Among the 132 whole blood samples, 43 EDTA samples showed platelet clumping. The DxH 900 PLT-I and BC-6800Plus PLT-I results demonstrated a strong correlation (r=0.711) for the EDTA samples but only a moderate correlation with the BC-6800Plus PLT-O results (r=0.506 and 0.545, respectively). The BC-6800Plus PLT-O results were consistent with the sodium citrate platelet counts, with a median dissociation rate of 102.5% (range, 74.9%–123.1%). The DxH 900 and BC-6800Plus analyzers had sensitivity values of 0.79 and 0.72, respectively, for platelet-clump flagging. When integrating the MC-80 digital image analysis results, the sensitivity of BC-6800Plus improved to 0.89 (standard mode) or 1.0 (PLT-Pro mode). Conclusions: BC-6800Plus PLT-O measurement results are close to the actual values obtained by platelet deaggregation with PTCP samples. Integrating the BC-6800Plus with a digital imaging analyzer effectively improved the diagnosis of PTCP and reduced the requirement for additional laboratory procedures.

Abstract

Background: EDTA-induced pseudothrombocytopenia (PTCP) during whole blood collection requires significant laboratory resources to obtain accurate results. We evaluated platelet-deaggregation function in EDTA-induced PTCP and platelet-clump flagging by the BC-6800Plus hematology analyzer using integrated digital image analysis. Methods: We prospectively collected 132 whole blood samples suspected of platelet clumping (102 in EDTA and 30 in sodium citrate) from 88 individuals. We compared platelet counts determined using the platelet count by impedance (PLT-I) function of the DxH 900 hematology analyzer and the PLT-I or optical platelet count (PLT-O) function of the BC- 6800Plus. Platelet clumping was verified through manual inspection and the MC-80 digital image analyzer. Results: Among the 132 whole blood samples, 43 EDTA samples showed platelet clumping. The DxH 900 PLT-I and BC-6800Plus PLT-I results demonstrated a strong correlation (r=0.711) for the EDTA samples but only a moderate correlation with the BC-6800Plus PLT-O results (r=0.506 and 0.545, respectively). The BC-6800Plus PLT-O results were consistent with the sodium citrate platelet counts, with a median dissociation rate of 102.5% (range, 74.9%–123.1%). The DxH 900 and BC-6800Plus analyzers had sensitivity values of 0.79 and 0.72, respectively, for platelet-clump flagging. When integrating the MC-80 digital image analysis results, the sensitivity of BC-6800Plus improved to 0.89 (standard mode) or 1.0 (PLT-Pro mode). Conclusions: BC-6800Plus PLT-O measurement results are close to the actual values obtained by platelet deaggregation with PTCP samples. Integrating the BC-6800Plus with a digital imaging analyzer effectively improved the diagnosis of PTCP and reduced the requirement for additional laboratory procedures.

참고문헌 (27)

  1. Banfi G, "The role of ethylenediamine tetraacetic acid(EDTA)as in vitro anticoagulant for diagnostic purposes", Clin Chem Lab Med, 45, pp.565~576 (2007)
  2. Gowland E, "Agglutination of platelets by a serum factor in the presence of EDTA", J Clin Pathol, 22, pp.460~464 (1969)
  3. Bartels PC, "Screening for EDTA-dependent deviations in platelet counts and abnormalities in platelet distribution histograms in pseudothrombocytopenia", Scand J Clin Lab Invest, 57, pp.629~636 (1997)
  4. Vicari A, "EDTA-dependent pseudothrombocytopaenia : a 12-month epidemiological study", Scand J Clin Lab Invest, 48, pp.537~542 (1988)
  5. Silvestri F, "Incidence and diagnosis of EDTA-dependent pseudothrombocytopenia in a consecutive outpatient population referred for isolated thrombocytopenia", Vox Sang, 68, pp.35~39 (1995)
  6. Baccini V, "Platelet counting: ugly traps and good advice. Proposals from the Frenchspeaking Cellular Hematology Group (GFHC)", J Clin Med, 9, pp.808~ (2020)
  7. Deng J, "Mindray SF-Cube technology : an effective way for correcting platelet count in individuals with EDTA dependent pseudo thrombocytopenia", Clin Chim Acta, 502, pp.99~101 (2020)
  8. Guo P, "Performance evaluation of the new platelet measurement channel on the BC-6800 Plus automated hematology analyzer", Int J Lab Hematol, 44, pp.281~287 (2022)
  9. Mukaka MM, "Statistics corner : a guide to appropriate use of correlation coefficient in medical research", Malawi Med J, 24, pp.69~71 (2012)
  10. Landis JR, "The measurement of observer agreement for categorical data", Biometrics, 33, pp.159~174 (1977)
  11. Jaeschke R, "Users’ guides to the medical literature. III. How to use an article about a diagnostic test. B. What are the results and will they help me in caring for my patients? The Evidence-Based Medicine Working Group", JAMA, 271, pp.703~707 (1994)
  12. Bao Y, "Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients", Transl Cancer Res, 9, pp.166~172 (2020)
  13. Lardinois B, "Pseudothrombocytopenia—a review on causes, occurrence and clinical implications", J Clin Med, 10, pp.594~ (2021)
  14. Wahed A, "Hematology and Coagulation", pp.277~291 (2020), Academic Press
  15. Zandecki M, "Spurious counts and spurious results on haematology analysers: a review. Part I: platelets", Int J Lab Hematol, 29, pp.4~20 (2007)
  16. Gulati G, "Unreliable automated complete blood count results:causes, recognition, and resolution", Ann Lab Med, 42, pp.515~530 (2022)
  17. Xu P, "The flagging features of the Sysmex XN-10 analyser for detecting platelet clumps and the impacts of platelet clumps on complete blood count parameters", Clin Chem Lab Med, 60, pp.748~755 (2022)
  18. Sandhaus LM, "Platelet counting by the Coulter LH 750, Sysmex XE 2100, and Advia 120: a comparative analysis using the RBC", platelet ratio reference method, 2002, pp.118, 235~241
  19. Fuster Ó, "Performance evaluation of low platelet count and platelet clumps detection on Mindray BC-6800 hematology analyzer", Clin Chem Lab Med, 54, pp.e49~e51 (2016)
  20. Sakurai S, "Aminoglycosides prevent and dissociate the aggregation of platelets in patients with EDTA-dependent pseudothrombocytopenia", Br J Haematol, 99, pp.817~823 (1997)
  21. Bizzaro N, "EDTA-dependent pseudothrombocytopenia: a clinical and epidemiological study of 112 cases, with 10-year follow-up", Am J Hematol, 50, pp.103~109 (1995)
  22. Casonato A, "EDTA dependent pseudothrombocytopenia caused by antibodies against the cytoadhesive receptor of platelet gpIIB-IIIA", J Clin Pathol, 47, pp.625~630 (1994)
  23. Pan LL, "Enhanced accuracy of optical platelet counts in microcytic anemia", Lab Med, 45, pp.32~36 (2014)
  24. Schoorl M, "New fluorescent method(PLT-F)on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting", Am J Clin Pathol, 140, pp.495~499 (2013)
  25. Schuff-Werner P, "Effective estimation of correct platelet counts in pseudothrombocytopenia using an alternative anticoagulant based on magnesium salt", Br J Haematol, 162, pp.684~692 (2013)
  26. Schuff-Werner P, "Pseudo-thrombocytopenia (PTCP). A challenge in the daily laboratory routine?", J Lab Med, 44, pp.295~304 (2020)
  27. Shreiner DP, "Pseudothrombocytopenia : manifestation of a new type of platelet agglutinin", Blood, 42, pp.541~549 (1973)

관련 논문AI추천

Reliability of the DI-60 Digital Image Analyzer for Detecting Platelet Clumping and Obtaining Accurate Platelet Counts김민선, 김보현, 윤영안, Choi Young-Jin, Heo Nam Hun, Woo Suhyeon대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호Maintenance of Platelet Counts with Low Level QC Materials and the Change in P-LCR according to Hemolysis with XN-9000심문정, Hyun-A Lee대한임상검사과학회 · 대한임상검사과학회지2018 · 제50권 제4호The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry김미정, 박상혁, 박찬정, 이보라, 장성수, 조영욱, 한민영대한진단검사의학회 · Annals of Laboratory Medicine2014 · 제34권 제6호Unreliable Automated Complete Blood Count Results: Causes, Recognition, and ResolutionGong Jerald, Gulati Gene, Uppal Guldeep대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제5호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호혈소판응집에 의한 위혈소판감소증의 검사실적 교정조덕, 기승정, 서순팔, 신명근, 신종희, 양동욱, 오봉준대한진단검사의학회 · Annals of Laboratory Medicine2005 · 제25권 제6호

이 간행물의 다른 논문 (25건)