Maintenance of Platelet Counts with Low Level QC Materials and the Change in P-LCR according to Hemolysis with XN-9000

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류기타자연과학

초록

The platelet count in clinical laboratories is essential for the diagnosis and treatment of hemostasis abnormalities, and accurate platelet counting in the low count range is of prime importance for deciding if a platelet transfusion is needed and for monitoring after chemotherapy. Quality control is designed to reduce and correct any deficiencies in the internal analytical process of a clinical laboratory prior to the release of patient results. Fragmented erythrocytes are the major confusing factors for platelet counting because of their similar size to platelets. The authors found that the low range QC values were out of 2SD with a Sysmex automatic analyzer in internal quality control process. Thus far, there has been little discussion on the relationship between hemolysis and the platelet parameters. Therefore, this study focused on the performance of automated platelet counts, including the PLT-F, the PLT-I, and PLT-O methods at the low platelet range using the low level QC materials and compared the 5 platelet parameters with the hemolyzed samples. The results showed that the CV was the smallest with PLT-F and P-LCR increased from 18.4 to 31.9% in the hemolysis samples. These results indicate that a more accurate estimation of the platelet counts can be achieved using the PLT-F method than the PLT-I method at the low platelet range. The use of the PLT-F system improves the confidence of results in low platelets samples in a routine hematology laboratory. The results suggest that P-LCR is a new parameter in assessing samples when the specimen is suspected of hemolysis and deterioration. Nevertheless, further studies will be needed to establish the relationship with P-LCR and hemolysis using human blood specimens.

Abstract

The platelet count in clinical laboratories is essential for the diagnosis and treatment of hemostasis abnormalities, and accurate platelet counting in the low count range is of prime importance for deciding if a platelet transfusion is needed and for monitoring after chemotherapy. Quality control is designed to reduce and correct any deficiencies in the internal analytical process of a clinical laboratory prior to the release of patient results. Fragmented erythrocytes are the major confusing factors for platelet counting because of their similar size to platelets. The authors found that the low range QC values were out of 2SD with a Sysmex automatic analyzer in internal quality control process. Thus far, there has been little discussion on the relationship between hemolysis and the platelet parameters. Therefore, this study focused on the performance of automated platelet counts, including the PLT-F, the PLT-I, and PLT-O methods at the low platelet range using the low level QC materials and compared the 5 platelet parameters with the hemolyzed samples. The results showed that the CV was the smallest with PLT-F and P-LCR increased from 18.4 to 31.9% in the hemolysis samples. These results indicate that a more accurate estimation of the platelet counts can be achieved using the PLT-F method than the PLT-I method at the low platelet range. The use of the PLT-F system improves the confidence of results in low platelets samples in a routine hematology laboratory. The results suggest that P-LCR is a new parameter in assessing samples when the specimen is suspected of hemolysis and deterioration. Nevertheless, further studies will be needed to establish the relationship with P-LCR and hemolysis using human blood specimens.

참고문헌 (16)

  1. Segal HC, "Accuracy of platelet counting haematology analyzers in severe thrombocytopenia and potential impact on platelet transfusion", Br J Haematol, 128, pp.520~525 (2005)
  2. Saliba RM, "Leukemia burden delays lymphocyte and platelet recovery after allo-SCT for AML", Bone Marrow Transplant, 43, pp.685~692 (2009)
  3. Briggs C, "Continuing developments with the automated platelet count", Int J Lab Hematol, 29, pp.77~91 (2007)
  4. Briggs C, "Performance evaluation of Sysmex haematology XN modular system", J Clin Pathol, 65, pp.1024~1030 (2012)
  5. Briggs C, "The most accurate platelet count on the Sysmex XE-2100: optical or impedance?", Clin Lab Haematol, 26, pp.157~158 (2004)
  6. Schoorl M, "New fluorescent method(PLT-F)on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting", Am J Clin Pathol, 140, pp.495~499 (2013)
  7. Niethammer AG, "Use of the platelet histogram maximum in evaluating thrombocytopenia", Am J Hematol, 60, pp.19~23 (1999)
  8. Park Y, "Mean platelet volume as an indicator of platelet activation : methodological issues", Platelets, 13, pp.301~306 (2002)
  9. Wada A, "Accuracy of a new platelet count system(PLT0F)depends on the staining property of its reagents", PLoS One, 10, pp.1~13 (2015)
  10. Grotto H, "Platelet larger cell ratio(P-LCR)in patients with dyslipidemia", Clin Lab Haematol, 26, pp.347~349 (2004)
  11. Briggs C, "New quantitative parameters on a recently introduced automated blood cell counter : the XE2100", Clin Lab Haematol, 22, pp.345~350 (2000)
  12. Tanaka Y, "Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers", J Clin Pathol, 28, pp.341~348 (2014)
  13. Tantanate C, "Performance evaluation of automated impedance and optical fluorescence platelet counts compared with international reference method in patients with thalassemia", Arch Pathol Lab Med, 141, pp.830~836 (2017)
  14. Zandecki M, "Spurious counts and sprious results on haematology analyzers", Int J Lab Hematol, 29, pp.4~20 (2007)
  15. Siegenthaler M, "Images in clinical medicine : spurious platelet count", N Engl J Med, 354, pp.1936~ (2006)
  16. Tantanate C, "Spuriously high platelet counts by various automated hematology analyzers in patient with disseminated intravascular coagulation", Clin Chem Lab Med, 53, pp.257~259 (2015)

관련 논문AI추천

Reliability of the DI-60 Digital Image Analyzer for Detecting Platelet Clumping and Obtaining Accurate Platelet Counts김민선, 김보현, 윤영안, Choi Young-Jin, Heo Nam Hun, Woo Suhyeon대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry김미정, 박상혁, 박찬정, 이보라, 장성수, 조영욱, 한민영대한진단검사의학회 · Annals of Laboratory Medicine2014 · 제34권 제6호Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia허정원, Kim Seunghwan, Park Sholhui, So Min-Kyung대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호자동혈구분석기 Sysmex XN-2000의 성능 및 Low WBC Mode를 이용한 백혈구 감별계산 정확성 평가김혜란, 송새암, 신정환, 오승환, 이자영, 이정녀, 전경란대한진단검사의학회 · Laboratory Medicine Online2015 · 제5권 제4호Unreliable Automated Complete Blood Count Results: Causes, Recognition, and ResolutionGong Jerald, Gulati Gene, Uppal Guldeep대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제5호Establishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000김미정, 박상혁, 박찬정, 이보라, 장성수, 조영욱, 한민영대한진단검사의학회 · Annals of Laboratory Medicine2016 · 제36권 제3호

이 간행물의 다른 논문 (19건)