The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry
1
조회
0
다운로드
0
인용
0
참고문헌
초록
Dear Editor The recently launched hematology autoanalyzer Sysmex XN- 2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA stain- ing dye named oxazine [1, 2]. The main advantage of the XN- 2000’s PLT-F method over the former instrument XE-2100’s op- tical method (PLT-O method) is the ability to identify apoptotic white cells and not count them as platelets [2]. The XN-2000’s PLT-F method was reported to estimate platelets in thrombocy- topenic samples more accurately than does the PLT-O method [1]. However, the performance of XN-2000’s PLT-F method still needs to be evaluated, including samples with broad range of platelet counts. We performed correlation and bias analyses comparing the platelet counts measured by the XN-2000 and XE-2100 autoanalyzers with those calculated by the immuno- platelet reference method (IRM) of flow cytometry.
Abstract
Dear Editor The recently launched hematology autoanalyzer Sysmex XN- 2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA stain- ing dye named oxazine [1, 2]. The main advantage of the XN- 2000’s PLT-F method over the former instrument XE-2100’s op- tical method (PLT-O method) is the ability to identify apoptotic white cells and not count them as platelets [2]. The XN-2000’s PLT-F method was reported to estimate platelets in thrombocy- topenic samples more accurately than does the PLT-O method [1]. However, the performance of XN-2000’s PLT-F method still needs to be evaluated, including samples with broad range of platelet counts. We performed correlation and bias analyses comparing the platelet counts measured by the XN-2000 and XE-2100 autoanalyzers with those calculated by the immuno- platelet reference method (IRM) of flow cytometry.
참고문헌 (4)
- Schoorl M, "New fluorescent method (PLT-F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting", Am J Clin Pathol, 140, pp.495~499 (2013)
- Briggs C, "Performance evaluation of the Sysmex haematology XN modular system", J Clin Pathol, 65, pp.1024~1030 (2012)
- International Council for Standardization in Haematology Expert Panel on Cytometry; International Society of Laboratory Hematology Task Force on Platelet Counting, "Platelet counting by the RBC", platelet ratio method. A reference method, 2001, pp.115, 460~464
- Arroyo JL, "Evaluation of a CD61 MoAb method for enumeration of platelets in thrombocytopenic patients and its impact on the transfusion decision-making process", Transfusion, 41, pp.1212~1216 (2001)