본문 바로가기
KERD
논문KCI 등재

The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry

1

조회

0

다운로드

0

인용

0

참고문헌

주제분류병리학

초록

Dear Editor The recently launched hematology autoanalyzer Sysmex XN- 2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA stain- ing dye named oxazine [1, 2]. The main advantage of the XN- 2000’s PLT-F method over the former instrument XE-2100’s op- tical method (PLT-O method) is the ability to identify apoptotic white cells and not count them as platelets [2]. The XN-2000’s PLT-F method was reported to estimate platelets in thrombocy- topenic samples more accurately than does the PLT-O method [1]. However, the performance of XN-2000’s PLT-F method still needs to be evaluated, including samples with broad range of platelet counts. We performed correlation and bias analyses comparing the platelet counts measured by the XN-2000 and XE-2100 autoanalyzers with those calculated by the immuno- platelet reference method (IRM) of flow cytometry.

Abstract

Dear Editor The recently launched hematology autoanalyzer Sysmex XN- 2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA stain- ing dye named oxazine [1, 2]. The main advantage of the XN- 2000’s PLT-F method over the former instrument XE-2100’s op- tical method (PLT-O method) is the ability to identify apoptotic white cells and not count them as platelets [2]. The XN-2000’s PLT-F method was reported to estimate platelets in thrombocy- topenic samples more accurately than does the PLT-O method [1]. However, the performance of XN-2000’s PLT-F method still needs to be evaluated, including samples with broad range of platelet counts. We performed correlation and bias analyses comparing the platelet counts measured by the XN-2000 and XE-2100 autoanalyzers with those calculated by the immuno- platelet reference method (IRM) of flow cytometry.

참고문헌 (4)

  1. Schoorl M, "New fluorescent method (PLT-F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting", Am J Clin Pathol, 140, pp.495~499 (2013)
  2. Briggs C, "Performance evaluation of the Sysmex haematology XN modular system", J Clin Pathol, 65, pp.1024~1030 (2012)
  3. International Council for Standardization in Haematology Expert Panel on Cytometry; International Society of Laboratory Hematology Task Force on Platelet Counting, "Platelet counting by the RBC", platelet ratio method. A reference method, 2001, pp.115, 460~464
  4. Arroyo JL, "Evaluation of a CD61 MoAb method for enumeration of platelets in thrombocytopenic patients and its impact on the transfusion decision-making process", Transfusion, 41, pp.1212~1216 (2001)

관련 논문AI추천

Maintenance of Platelet Counts with Low Level QC Materials and the Change in P-LCR according to Hemolysis with XN-9000심문정, Hyun-A Lee대한임상검사과학회 · 대한임상검사과학회지2018 · 제50권 제4호자동혈구분석기 Sysmex XN-2000의 성능 및 Low WBC Mode를 이용한 백혈구 감별계산 정확성 평가김혜란, 송새암, 신정환, 오승환, 이자영, 이정녀, 전경란대한진단검사의학회 · Laboratory Medicine Online2015 · 제5권 제4호Reliability of the DI-60 Digital Image Analyzer for Detecting Platelet Clumping and Obtaining Accurate Platelet Counts김민선, 김보현, 윤영안, Choi Young-Jin, Heo Nam Hun, Woo Suhyeon대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia허정원, Kim Seunghwan, Park Sholhui, So Min-Kyung대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호Establishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000김미정, 박상혁, 박찬정, 이보라, 장성수, 조영욱, 한민영대한진단검사의학회 · Annals of Laboratory Medicine2016 · 제36권 제3호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호