Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: Complete blood count (CBC) and white blood cell (WBC) differential are essential tests for various diseases. Related to this, the Mindray BC-6200 automated hematology analyzer (BC-6200, Mindray Bio-Medical Electronics Co., Ltd, China) was recently launched in clinical laboratories. This study aimed to evaluate the analytical and flagging performance of BC-6200. Methods: Using 688 whole blood samples, the precision and carryover of 12 CBC parameters were evaluated with BC-6200 according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP15-A3 and H26-A2, respectively. 11 CBC parameters of BC-6200 were compared with Sysmex XE-2100 (XE-2100, Sysmex Corporation, Japan) according to the CLSI guideline EP09c. To evaluate the flagging performance for blasts, immature granulocytes (IG), atypical lymphocytes (AL), and nucleated red blood cells (NRBC) of BC-6200, sensitivity, specificity, and efficiency to manual counts were estimated according to the CLSI guideline H20-A2. Results: Precisions of WBC, red blood cells (RBC), hemoglobin (Hb), hematocrit (Hct), and platelets (PLT) were acceptable. Carryover was less than 1% in WBC, RBC, Hb, Hct, and PLT. In WBC differentials, BC-6200 and XE-2100 showed very high correlations, except for basophils. Flagging performances of BC-6200 showed excellent results in efficiency; 91.4% for blasts, 79.4% for IG, 75.5% for AL, and 98.6% for NRBC. Conclusions: The analytical performances of BC-6200 were acceptable and comparable with those of XE-2100. The flagging performance was also comparable with that of manual counts. BC-6200 would be a competent instrument in clinical laboratories

Abstract

Background: Complete blood count (CBC) and white blood cell (WBC) differential are essential tests for various diseases. Related to this, the Mindray BC-6200 automated hematology analyzer (BC-6200, Mindray Bio-Medical Electronics Co., Ltd, China) was recently launched in clinical laboratories. This study aimed to evaluate the analytical and flagging performance of BC-6200. Methods: Using 688 whole blood samples, the precision and carryover of 12 CBC parameters were evaluated with BC-6200 according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP15-A3 and H26-A2, respectively. 11 CBC parameters of BC-6200 were compared with Sysmex XE-2100 (XE-2100, Sysmex Corporation, Japan) according to the CLSI guideline EP09c. To evaluate the flagging performance for blasts, immature granulocytes (IG), atypical lymphocytes (AL), and nucleated red blood cells (NRBC) of BC-6200, sensitivity, specificity, and efficiency to manual counts were estimated according to the CLSI guideline H20-A2. Results: Precisions of WBC, red blood cells (RBC), hemoglobin (Hb), hematocrit (Hct), and platelets (PLT) were acceptable. Carryover was less than 1% in WBC, RBC, Hb, Hct, and PLT. In WBC differentials, BC-6200 and XE-2100 showed very high correlations, except for basophils. Flagging performances of BC-6200 showed excellent results in efficiency; 91.4% for blasts, 79.4% for IG, 75.5% for AL, and 98.6% for NRBC. Conclusions: The analytical performances of BC-6200 were acceptable and comparable with those of XE-2100. The flagging performance was also comparable with that of manual counts. BC-6200 would be a competent instrument in clinical laboratories

참고문헌 (24)

  1. Libby P, "Clonal hematopoiesis : crossroads of aging, cardiovascular disease, and cancer : JACC review topic of the week", J Am Coll Cardiol, 74, pp.567~577 (2019)
  2. Orwoll ES, "Hematologic abnormalities in patients with endocrine and metabolic disorders", Hematol Oncol Clin North Am, 1, pp.261~279 (1987)
  3. Briggs C, "ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting", Int J Lab Hematol, 36, pp.613~627 (2014)
  4. Sullivan E, "Hematology analyzer : from workhorse to thoroughbred", LABMEDICINE, 37, pp.273~278 (2006)
  5. Park S, "False-positive fiag of WBC and change of mean platelet volume(MPV)caused by K3-EDTA on the DxH 900 hematology analyzer", Scand J Clin Lab Invest, 80, pp.644~648 (2020)
  6. Kulik K, "Evaluation and comparison of the new Mindray BC-6200 hematology analyzer with ADVIA 2120i", Int J Lab Hematol, 43, pp.395~402 (2021)
  7. Delsol G, "Leukoerythroblastosis and cancer frequency, prognosis, and physiopathologic signiflcance", Cancer, 44, pp.1009~1013 (1979)
  8. Stachon A, "Nucleated red blood cells in the blood of medical intensive care patients indicate increased mortality risk : a prospective cohort study", Crit Care, 11, pp.R62~ (2007)
  9. Kim H, "Reference intervals for clinically reportable platelet parameters on the Mindray BC-6800Plus hematology analyzer", Clin Chem Lab Med, 58, pp.e213~e215 (2020)
  10. Deng J, "Mindray SF-Cube technology : an effective way for correcting platelet count in individuals with EDTA dependent pseudothrombocytopenia", Clin Chim Acta, 502, pp.99~101 (2020)
  11. 조윤아, "Mindray BC-6800 자동혈구분석기의 유용성 평가", Laboratory Medicine Online, 3(3), pp.131~137 (2013), 대한진단검사의학회
  12. Briggs C, "New quantitative parameters on a recently introduced automated blood cell counter – the XE 2100", Clin Lab Hematol, 22, pp.345~350 (2000)
  13. User veriflcation of precision and estimation of bias; Approved guideline–Third edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document EP15-A3 (2014)
  14. Validation, veriflcation, and quality assurance of automated hematology analyzers; Approved standard–Second edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document H26-A2 (2010)
  15. Measurement procedure comparison and bias estimation using patient samples, Third edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI guideline EP09c (2018)
  16. Reference leukocyte (WBC) differential count (proportional) and evaluation of instrumental method; Approved standard–Second edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document H20-A2 (2007)
  17. Vis JY, "Veriflcation and quality control of routine hematology analyzers", Int J Lab Hematol, 38(S1), pp.100~109 (2016)
  18. Mukaka MM, "Statistics corner : a guide to appropriate use of correlation coefflcient in medical research", Malawi Med J, 24, pp.69~71 (2012)
  19. Nakul-Aquaronne D, "Evaluation of the Sysmex Xe-2100 hematology analyzer in hospital use", J Clin Lab Anal, 17, pp.113~123 (2003)
  20. Takemura H, "Evaluation of cell count and classiflcation capabilities in body fiuids using a fully automated Sysmex XN equipped with high-sensitive Analysis(hsA)mode and DI-60 hematology analyzer system", PLoS One, 13, pp.e0195923~ (2018)
  21. Lesesve JF, "Accurate basophils counting: not an easy goal!", Clin Lab Hematol, 27, pp.143~144 (2005)
  22. Amundsen EK, "Is the blood basophil count sufflciently precise, accurate, and speciflc? : three automated hematology instruments and fiow cytometry compared", Am J Clin Pathol, 137, pp.86~92 (2012)
  23. Kim H, "Evaluation of ABX Pentra DX 120 and Sysmex XE-2100 in umbilical cord blood", Int J Lab Hematol, 35, pp.658~665 (2013)
  24. Shen Y, "Clinical performance evaluation of the new hematology analyzer Mindray BC-6000", Int J Lab Hematol, 41, pp.622~634 (2019)

관련 논문AI추천

Mindray BC-6800 자동혈구분석기의 유용성 평가강희정, 김미영, 김한성, 이영경, 조윤아대한진단검사의학회 · Laboratory Medicine Online2013 · 제3권 제3호Validation of Hematology Rapid Reporting System for Complete Blood Cell Count with Differential김지용, 박지영, 윤수영, 이정민대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제1호자동혈구분석기 Sysmex XN-2000의 성능 및 Low WBC Mode를 이용한 백혈구 감별계산 정확성 평가김혜란, 송새암, 신정환, 오승환, 이자영, 이정녀, 전경란대한진단검사의학회 · Laboratory Medicine Online2015 · 제5권 제4호Current Status and Proposal of a Guideline for Manual Slide Review of Automated Complete Blood Cell Count and White Blood Cell Dfferential권계철, 김영재, 김희진, 박효순, 신상용, 우희연, 이영경, 장윤환, 조성란, 채석래, 최종락, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제6호Differential Blast Counts Obtained by Automated Blood Cell Analyzers김용구, 박연준, 오은지, 임지향, 정승원, 채효진, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제6호Effective and Practical Complete Blood Count Delta Check Method and Criteria for the Quality Control of Automated Hematology AnalyzersCho Young-Uk, Jang Seongsoo, Kim Min-Sun, Kim Seung-Il, Namgoong Seung, Park Chan-Jeoung대한진단검사의학회 · Annals of Laboratory Medicine2023 · 제43권 제5호

이 간행물의 다른 논문 (11건)