STATEMENT OF PROBLEM. Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink. PURPOSE. In this study, Bicon. Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured. MATERIAL AND METHODS. 10 Bicon. implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load’s step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutmentfixture under each loading condition from reference length. RESULTS. It was evident, that the amount of abutment sinking in Bicon. Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm. CONCLUSION. Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.
Abstract
STATEMENT OF PROBLEM. Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink. PURPOSE. In this study, Bicon. Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured. MATERIAL AND METHODS. 10 Bicon. implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load’s step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutmentfixture under each loading condition from reference length. RESULTS. It was evident, that the amount of abutment sinking in Bicon. Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm. CONCLUSION. Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.
참고문헌 (16)
- Binon PP, "Implants and components: entering the new millennium", Int J Oral Maxillofac Implants, 15, pp.76~94 (2000)
- Goodacre CJ, "Clinical complications of osseointegrated implants", J Prosthet Dent, 81, pp.537~552 (1999)
- Schwarz MS, "Mechanical complications of dental implants", Clin Oral Implants Res, 11, pp.156~158 (2000)
- Jemt T, "Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study", Int J Oral Maxillofac Implants, 6, pp.29~36 (1991)
- Taylor TD, "Prosthodontic problems and limitations associated with osseointegration", J Prosthet Dent, 79, pp.74~78 (1998)
- Albrektsson T, "A multicenter report on osseointegrated oral implants", J Prosthet Dent, 60, pp.75~84 (1988)
- Keating K, "Connecting abutments to Dental implants “An Engineers perspective”", Irish Dentist, pp.43~46 (2001)
- Chapman RJ, "The locking taper attachment for implant abutments: use and reliability", Implant Dent, 5, pp.257~261 (1996)
- Gibbs CH, "Occlusal forces during chewing and swallowing as measured by sound transmission", J Prosthet Dent, 46, pp.443~449 (1981)
- Robert G. Craig, "Restorative Dental Materials, 6th ed", pp.60~62 (1980), C.V. Mosby Co
- Anderson DJ, "Measurement of stress in mastication. I", J Dent Res, 35, pp.664~670 (1956)
- Anderson DJ, "Measurement of stress in mastication. II", J Dent Res, 35, pp.671~673 (1956)
- Richter EJ, "In vivo vertical forces on implants", Int J Oral Maxillofac Implants, 10, pp.99~108 (1995)
- Van Eijden TM, "Three dimensional analysis of human biteforce magnitude and moment", Arch Oral Biol, 36, pp.535~539 (1991)
- 이한라, "하중조건에 따른 원추형 내측연결 임플랜트 시스템에서 지대주 침하 및 적합에 관한 연구", 구강회복응용과학지, 24(1), pp.77~90 (2008), 대한턱관절교합학회
- Dailey B, "Axial displacement of abutments into implants and implant replicas, with the tapered cone-screw internal connection, as a function of tightening atorque", Int J Oral Maxillofac Implants, 24, pp.251~256 (2009)