Mechanical analysis of conventional and small diameter conical implant abutments

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

PURPOSE. The aim of the present study was to evaluate if a smaller morse taper abutment has a negative effect on the fracture resistance of implant-abutment connections under oblique compressive loads compared to a conventional abutment. MATERIALS AND METHODS. Twenty morse taper conventional abutments (4.8 mm diameter) and smaller abutments (3.8 mm diameter) were tightened (20 Ncm) to their respective implants (3.5 * 11 mm) and after a 10 minute interval, implant/abutment assemblies were subjected to static compressive test, performed in a universal test machine with 1 mm/min displacement, at 45 degree inclination. The maximum deformation force was determined. Data were statistically analyzed by student t test. RESULTS. Maximum deformation force of 4.8 mm and 3.8 mm abutments was approximately 95.33 kgf and 95.25 kgf, respectively, but no fractures were noted after mechanical test. Statistical analysis demonstrated that the evaluated abutments were statistically similar (P=.230). CONCLUSION. Abutment measuring 3.8 mm in diameter (reduced) presented mechanical properties similar to 4.8 mm (conventional) abutments, enabling its clinical use as indicated.

Abstract

PURPOSE. The aim of the present study was to evaluate if a smaller morse taper abutment has a negative effect on the fracture resistance of implant-abutment connections under oblique compressive loads compared to a conventional abutment. MATERIALS AND METHODS. Twenty morse taper conventional abutments (4.8 mm diameter) and smaller abutments (3.8 mm diameter) were tightened (20 Ncm) to their respective implants (3.5 * 11 mm) and after a 10 minute interval, implant/abutment assemblies were subjected to static compressive test, performed in a universal test machine with 1 mm/min displacement, at 45 degree inclination. The maximum deformation force was determined. Data were statistically analyzed by student t test. RESULTS. Maximum deformation force of 4.8 mm and 3.8 mm abutments was approximately 95.33 kgf and 95.25 kgf, respectively, but no fractures were noted after mechanical test. Statistical analysis demonstrated that the evaluated abutments were statistically similar (P=.230). CONCLUSION. Abutment measuring 3.8 mm in diameter (reduced) presented mechanical properties similar to 4.8 mm (conventional) abutments, enabling its clinical use as indicated.

참고문헌 (9)

  1. Steinebrunner L, "Implant-abutment interface design affects fatigue and fracture strength of implants", Clin Oral Implants Res, 19, pp.1276~1284 (2008)
  2. Bozkaya D, "Mechanics of the taper integrated screwedin(TIS) abutments used in dental implants", J Biomech, 38, pp.87~97 (2005)
  3. Coppede∧ AR, "Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: an in vitro study", Int J Prosthodont, 22, pp.283~286 (2009)
  4. Norton MR, "An in vitro evaluation of the strength of a 1-piece and 2-piece conical abutment joint in implant design", Clin Oral Implants Res, 11, pp.458~464 (2000)
  5. Reddy MS, "Initial clinical efficacy of 3-mm implants immediately placed into function in conditions of limited spacing", Int J Oral Maxillofac Implants, 23, pp.281~288 (2008)
  6. Khraisat A, "Fatigue resistance of two implant", abutment joint designs, 2002, pp.88, 604~610
  7. Weigl P., "New prosthetic restorative features of Ankylos implant system", J Oral Implantol, 30, pp.178~188 (2004)
  8. Perriard J, "Fatigue resistance of ITI implant-abutment connectors - a comparison of the standard cone with a novel internally keyed design", Clin Oral Implants Res, 13, pp.542~549 (2002)
  9. Okeson JP, "Management of temporomandibular disorders and occlusion. 4thed", pp.44~ (1998), Mosby

관련 논문AI추천

Influence of the implant-abutment connection design and diameter on the screw joint stability신현모, 윤미정, 전영찬, 정창모, 허중보, Brian Myung Chang대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제2호Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems양동석, 윤미정, 정창모, 조재영, 허재찬, 허중보대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제6호Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connectionGiuseppe Troiano, Khrystyna Zhurakivska, Luigi Laino, Rosario Prisco대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제6호Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading이주희, 차현석대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제6호Effect of cyclic loading on axial displacement of abutment into implant with internal tapered connection: a pilot study곽재영, 김성균, 설현우, 한종현, 허성주대한치과보철학회 · 대한치과보철학회지2013 · 제51권 제4호Influence of the connection design and titanium grades of the implant complex on resistance under static loading박수정, 안수진, 이석원, 이성복대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제5호