초록

In this study, the optimal driving frequency was derived through finite element analysis (FEA) to optimize the developed piezoelectric ultrasonic medical devices(PUMD) for bone surgery. The core of the PUMD is the piezo- electric ceramic (PZT), which is a vibrator that generates vibration energy. The piezoelectric ceramic shows the max- imum current value with respect to the input voltage at the resonance frequency, which generates the maximum mechanical vibration. In the past, various studies have been conducted related to the analysis of PUMD, but most of the research so far has been limited to free vibration analysis. However, in order to derive the accurate resonant fre- quency, the initial stress generated by bolt tightening in the bolt-clamped Langevin type transducer (BLT) must be considered. In this study, after designing a PUMD, the driving performance according to the bolt tightening value was analyzed through FEA, and this was experimentally verified. First, the resonance mode and frequency response were confirmed through modal and harmonic analysis at 20-40 kHz, which is known as the optimal driving frequency band of PUMD for bone surgery. In addition, the design of the PUMD was confirmed by checking the mechanical behavior of the tip and the piezoelectric ceramic at the resonant frequency. Consequentially, the characteristic evaluation was performed, and it was confirmed that the resonant frequency result derived through the FEA was reasonable. Through this study, we presented a more rational FEA method than before for BLT transducers. We expect that this will shorten the time and cost of developing a PUMD, and will enable the development of more stable and high-quality products.

Abstract

In this study, the optimal driving frequency was derived through finite element analysis (FEA) to optimize the developed piezoelectric ultrasonic medical devices(PUMD) for bone surgery. The core of the PUMD is the piezo- electric ceramic (PZT), which is a vibrator that generates vibration energy. The piezoelectric ceramic shows the max- imum current value with respect to the input voltage at the resonance frequency, which generates the maximum mechanical vibration. In the past, various studies have been conducted related to the analysis of PUMD, but most of the research so far has been limited to free vibration analysis. However, in order to derive the accurate resonant fre- quency, the initial stress generated by bolt tightening in the bolt-clamped Langevin type transducer (BLT) must be considered. In this study, after designing a PUMD, the driving performance according to the bolt tightening value was analyzed through FEA, and this was experimentally verified. First, the resonance mode and frequency response were confirmed through modal and harmonic analysis at 20-40 kHz, which is known as the optimal driving frequency band of PUMD for bone surgery. In addition, the design of the PUMD was confirmed by checking the mechanical behavior of the tip and the piezoelectric ceramic at the resonant frequency. Consequentially, the characteristic evaluation was performed, and it was confirmed that the resonant frequency result derived through the FEA was reasonable. Through this study, we presented a more rational FEA method than before for BLT transducers. We expect that this will shorten the time and cost of developing a PUMD, and will enable the development of more stable and high-quality products.

참고문헌 (23)

  1. Mitragotri S, "Healing sound: the use of ultrasound in drug delivery and other therapeutic applications", Nat. Rev. Drug Discov, 4(3), pp.255~260 (2005)
  2. Felver B, "Cavi-tation occurrence around ultrasonic dental scalers", Ultrason Sonochem, 16(5), pp.692~697 (2009)
  3. Park R, "Root canal irrigation system using remotely generated high-power ultrasound", Ultrason Sonochem, pp.106168~ (2022)
  4. Seo HG, "Performance Evaluation of MR Compatible Transducer Actuator Proto-type for High Intensity Focused Ultrasound(HIFU) System for Prostate Treatment", Proceedings of the KIEE Conference, pp.413~414 (2021)
  5. Kang MG, "Recent Progress on PZT Based Piezoelectric Energy", Actuators, 5(1), pp.5~ (2016)
  6. Li X, "Full and half-wavelength ultra-sonic percussive drills", IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 65(11), pp.2150~2159 (2018)
  7. 이주희, "치과용 초음파 골수술기 설계 및 평가", 전기전자재료학회논문지, 25(12), pp.990~995 (2012), 한국전기전자재료학회[DOI]
  8. 박기문, "치과용 골 수술기의 초음파 진동자 설계 및 평가", 한국기계가공학회지, 15(3), pp.102~108 (2016), 한국기계가공학회[DOI]
  9. 유중돈, "초음파 용접용 혼의 설계", 대한용접접합학회지, 26(1), pp.9~11 (2008), 대한용접접합학회
  10. Jeong AM, "Optimization Method for Ultrasonic Transducer Design to the Bonded Dissimilar Materi-als", Proceeding of KSPE 2010 spring Conference, pp.157~158 (2010)
  11. 김상우, "초음파 진동 딥 드로잉 공정에서의 마찰감소효과 분석을 위한 유한요소해석 및 실험", 소성∙가공, 23(7), pp.413~418 (2014), 한국소성∙가공학회
  12. 김경태, "하이브리드 연삭시스템 초음파 공구 개발", 한국정밀공학회지, 30(11), pp.1121~1128 (2013), 한국정밀공학회
  13. 이수호, "초음파 융착기용 압전 세라믹스의 전기적 특성", 전기전자학회논문지, 22(1), pp.201~204 (2018), 한국전기전자학회[DOI]
  14. 서용석, "Ti-6Al-4V 합금의 동적 변형 거동에 관한 연구", 한국군사과학기술학회지, 20(2), pp.206~216 (2017), 한국군사과학기술학회[DOI]
  15. 이봉구, "와이어 본딩용 초음파 공구혼 설계에 관한 연구", 한국생산제조학회지, 22(4), pp.717~722 (2013), 한국생산제조학회[DOI]
  16. 김선락, "초음파 용접용 실린더 혼의 설계", 대한용접접합학회지, 27(4), pp.60~66 (2009), 대한용접접합학회
  17. Park MH, "Changes of Vibra-tional characteristics due to the spaces of the Langevin type vibrators", Proceedings of 2002 Spring Symposium of KIEEME, pp.97~102 (2002)
  18. Lenka Č, "The Effect of Waveguide Shapes on the Modal Properties and Amplification Factors", Ved. pr. Materiálo-votechnol. fak. Slov. tech. univ. Bratisl. sídlom Trnave, 21(Special Issue), pp.167~172 (2013)
  19. 김지선, "20kHz 급 금속 초음파 융착용 스텝형 바 혼의 최적설계", 한국산학기술학회논문지, 20(12), pp.94~101 (2019), 한국산학기술학회[DOI]
  20. 김영우, "와이어 본딩용 초음파 혼의 진동 특성", 대한기계학회논문집 A, 38(2), pp.227~233 (2014), 대한기계학회
  21. He T, "Optimization design for ultrasonic horn with large amplitude based on genetic algorithm", J. Vibroen-gineering, 17(3), pp.1157~1168 (2015)
  22. Kim DS, "Variation In the Vibration Characteristics of Piezoelectric Transducers due to Pre-Pressure", Proceed-ings of the KSNVE Annual Spring Conference, pp.484~487 (2005)

관련 논문AI추천