초음파 수술기의 수술 효율성 향상을 위한 진동자 임피던스 측정에 따른 조직 분류 연구
Classification of Organs Using Impedance of Ultrasonic Surgical Knife to improve Surgical Efficiency
0
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Ultrasonic shears is currently in wide use as an energy device for minimal invasive surgery. There is an advantage of minimizing the carbonization behavior of the tissue due to the vibrational energy transfer system of the transducer by applying a piezoelectric ceramic. However, the vibrational energy transfer system has a pitfall in energy consumption. When the movement of the forceps is interrupted by the tissue, the horn which transfers the vibrational energy of the transducer will be affected. A study was performed to recognize different tissues by measuring the impedance of the transducer of the ultrasonic shears in order to find the factor of energy consumption according to the tissue. In the first stage of the study, the voltage and current of the transducer connecting portion were measured, along with the phase changes. Subsequently, in the second stage, the impedance of the transducer was directly measured. In the final stage, using the handpiece, we grasped the tissue and observed the impedance differences appeared in the transducer To verify the proposed tissue distinguishing method, we used the handpiece to apply a force between 5N and 10N to pork while increasing the value of the impedance of the transducer from 400 Ω.. It was found that fat and skin tissue, tendon, liver and protein all have different impedance values of 420 Ω,490 Ω, 530 Ω, and 580 Ω, respectively. Thus, the impedance value can be used to distinguish the type of tissues grasped by the forceps. In the future study, this relationship will be used to improve the energy efficiency of ultrasonic shears.
Abstract
Ultrasonic shears is currently in wide use as an energy device for minimal invasive surgery. There is an advantage of minimizing the carbonization behavior of the tissue due to the vibrational energy transfer system of the transducer by applying a piezoelectric ceramic. However, the vibrational energy transfer system has a pitfall in energy consumption. When the movement of the forceps is interrupted by the tissue, the horn which transfers the vibrational energy of the transducer will be affected. A study was performed to recognize different tissues by measuring the impedance of the transducer of the ultrasonic shears in order to find the factor of energy consumption according to the tissue. In the first stage of the study, the voltage and current of the transducer connecting portion were measured, along with the phase changes. Subsequently, in the second stage, the impedance of the transducer was directly measured. In the final stage, using the handpiece, we grasped the tissue and observed the impedance differences appeared in the transducer To verify the proposed tissue distinguishing method, we used the handpiece to apply a force between 5N and 10N to pork while increasing the value of the impedance of the transducer from 400 Ω.. It was found that fat and skin tissue, tendon, liver and protein all have different impedance values of 420 Ω,490 Ω, 530 Ω, and 580 Ω, respectively. Thus, the impedance value can be used to distinguish the type of tissues grasped by the forceps. In the future study, this relationship will be used to improve the energy efficiency of ultrasonic shears.
참고문헌 (20)
- C.C. Chung, "Doubleblind, randomized trialcomparing Harmonic Scalpel hemorrhoidectomy, bipolar scissors hemorrhoidectomy, and scissors excision : ligation technique", Diseases of the colon and rectum, 45, pp.789~2002 (2002)
- J.H. Jo, "Thema : Piezoelectric Ceramic; Development trend of the piezoelectric ceramic material", 25, pp.10~ (2012), The Korean Institute of Electrical and Electronic Material Engineers
- D. C. Lee, "Development trend of the ultrasonic electric motor by piezoelectric ceramics", The Korea Institute of Electrical Engineers, 46, pp.23~ (1997)
- G. M. Lee, "Technical Description : Development trend with the introduction of the piezoelectric ceramic oscillator", Journal of KIEEME, 10, pp.172~ (1997)
- S. Dong, "Research Trends of Lead-Free Piezoelectric ceramic material", Journal of the Korean Ceramic Society, 8, pp.14~ (2005)
- J.E. Fox, "Practical Electrotherapy: A Guide to Safe Application" (2007), Churchill Livingstone Elsevier
- H.R. Kim, "Impedance analyzer development and energy control system research for surgical ultrasonic cutter's improvement", Journal of the Korea Society of Imageing Informatics in Medicine, 18(69), pp.12~ (2012)
- C.N. Jin, "Electric Properties of LB Films using Impedance Analysis of Quartz Crystal", The Korean Instiute of Electirical Engineers, 48, pp.503~ (1999)
- 차정옥, "Chemical Solution Deposition 방법을 이용한 BiFeO3", Pb(Zr0.52Ti0.48)O3 다층박막의 전기적 특성에 대한 연구, 2010, pp.19(1), 52~57[DOI]
- H. S. Potdar, "Dielectric Properties of BaTiO3 Ceramics Prepared from Powders with Bimodal Distribution", Journal of Industrial and Engineering Chemistry, 12(4), pp.584~588 (2006), 한국공업화학회
- G. Haertling, "Ceramic materials for electronics", pp.12~ (1991)
- K.W. Lee, "Technology and market trends of the present and future of the piezoelectric ceramics", Journal of the Korean Ceramic Society, 8, pp.33~ (2005)
- 박재성, "유한요소해석법을 이용한 2-2형 압전복합재료 초음파 트랜스듀서의 설계 및 압전특성", 전자공학회논문지 - IE, 48(2), pp.40~46 (2011), 대한전자공학회
- 김종민, "고분자의 극성에 따른 수정진동자 공진저항의 변화 특성", 공업화학, 18(1), pp.71~76 (2007), 한국공업화학회
- T. Morita, "Piezoelectric Materials Synthesized by the Hydrothermal Method and Their Applications", Materials, 3(52), pp.36~ (2010)
- D. Robertson, "Comparison of the frequency and physical nature of the lowest order parasitic mode in single crystal and ceramic 2-2 and 1-3 piezoelectric composite transducers", Ultrasonics, Ferroelectrics and Frequency Control, IEEE Transactions on, 53(15), pp.03~ (2006)
- S.H. Lee, "Piezoelectric and Dielectric Properties of NKN-(1-x)BNT-xBT Ceramics", Journal of KIEEME, 23, pp.771~ (2010)
- A. Turo, "Ultra-low noise front-end electronics for aircoupled ultrasonic non-destructive evaluation", Ndt & E International, 36, pp.93~ (2003)
- Y. Xu, "Ferroelectric materials and their applications" (1991)
- Y. Yanez, "Designing amplifiers with very low output noise for high impedance piezoelectric transducers", NDT & E International, 38, pp.491~ (2005)