In this study, the effect of the shape of the specimen and the feedrate of the dental ultrasonic surgical instrument on the cutting force and surface roughness of the specimen is analyzed. Experimental specimens were made of SAWBONES artificial bone materials in square and spherical specimens. In addition, the cutting feedrate of the surgical instrument was controlled through the developed moving system. The cutting force generated when cutting the specimen was measured through a force sensor. After the experiment, the cutting surface of the specimen was observed through a three-dimensional optical microscope and the surface roughness was measured. Through one-way ANOVA, the effect of each specimen shape and feed rate on surface roughness was analyzed. As a result of the experiment, the cutting force increased proportionally in the initial feed rate increase stage, but the increase in cutting force decreased as the feed rate continued to increase. Also, the cutting force showed a difference according to the shape of the specimen. The spherical specimen with a relatively small cutting surface area had less cutting force than the square specimen. However, as a result of one-way ANOVA, it was found that the specimen shape and feed rate did not affect the surface roughness. In future studies, it is expected to be used for comparative analysis of ultrasonic surgical instruments and correlation analysis between cutting factors.
Abstract
In this study, the effect of the shape of the specimen and the feedrate of the dental ultrasonic surgical instrument on the cutting force and surface roughness of the specimen is analyzed. Experimental specimens were made of SAWBONES artificial bone materials in square and spherical specimens. In addition, the cutting feedrate of the surgical instrument was controlled through the developed moving system. The cutting force generated when cutting the specimen was measured through a force sensor. After the experiment, the cutting surface of the specimen was observed through a three-dimensional optical microscope and the surface roughness was measured. Through one-way ANOVA, the effect of each specimen shape and feed rate on surface roughness was analyzed. As a result of the experiment, the cutting force increased proportionally in the initial feed rate increase stage, but the increase in cutting force decreased as the feed rate continued to increase. Also, the cutting force showed a difference according to the shape of the specimen. The spherical specimen with a relatively small cutting surface area had less cutting force than the square specimen. However, as a result of one-way ANOVA, it was found that the specimen shape and feed rate did not affect the surface roughness. In future studies, it is expected to be used for comparative analysis of ultrasonic surgical instruments and correlation analysis between cutting factors.
참고문헌 (19)
- 김나리, "치과 임플란트 수술용 고출력 초음파 수술기 설계 및 제작", 전기전자재료학회논문지, 30(10), pp.656~664 (2017), 한국전기전자재료학회[DOI]
- 사민우, "치과용 초음파 수술기의 절삭성능 평가를 위한 핸드피스 이송 시스템 구축", 한국정밀공학회지, 33(5), pp.377~383 (2016), 한국정밀공학회
- 사민우, "치과 치료용 초음파 스케일러의 성능 및 안정성에 대한 실험 평가", 대한기계학회논문집 A, 41(1), pp.13~19 (2017), 대한기계학회
- Elfar J, "Composite Bone Models in Orthopaedic Surgery Research and Education", J Am Acad Orthop Surg, 22(2), pp.111~120 (2014)
- Dentkos TR, "Evaluation of Cutting Efficiency of Orthograde Ultrasonic Tips by Using a Nonstatic Model", J Endod, 34(7), pp.863~865 (2008)
- Kim DY, "Evaluation of High-Speed Handpiece Cutting Efficiency According to Bur Eccentricity: An In Vitro Study", Applied Sciences, 9(16), pp.3395~ (2019)
- Al-Abdulah KI, "A modelbased bone milling state identification method via force sensing for a robotic surgical system", int J Med Robot, 15(3), pp.e1989~ (2019)
- Mae T, "Characteristics of ultrasound device : a new techn and excavation", J Exp Orthop, 6(1), pp.35~ (2019)
- Sinjari B, "Comparison of a Novel Ultrasonic Scaler Tip vs Conventional Design on a Titanium Surface", Materials, 11(12), pp.2345~ (2018)
- 이주희, "치과용 초음파 골수술기 설계 및 평가", 전기전자재료학회논문지, 25(12), pp.990~995 (2012), 한국전기전자재료학회[DOI]
- Chun KA, "Evaluation of the Safety and efficiency of novel metallic implant scaler tips manufactured by the powder injection molding techique", BMC Oral Health, 17(1), pp.110~ (2017)
- Baek SH, "Evaluation of the Safety and efficiency of novel metallic implant scaler tip on titanium surfaces", Clin Oral Implants Res, 23(11), pp.1269~1274 (2012)
- Harder S, "Performance of ultrasonic devices for bone surgery and associated intraosseous temperature development", int J Oral Maxillofac Implants, 24(3), pp.484~490 (2009)
- Misch CE, "Mechanical properties of trabecular bone in the human mandible: Implications for dental implant treatment planning and surgical placement", J Oral Maxillofac Surg, 57(6), pp.700~706 (1999)
- Rasagopal P, "A study surface integrity of aluminum hybrid composites during milling operation", J Mater Res Technol, 9(3), pp.4884~4893 (2020)
- Shah D, "Effect of Tool Nose Radius and Machining Parameters on Cutting Force, Cutting Temperature and Surface Roughness-An Experimental Study of Ti-6Al-4V(ELI)", Mater. Today Proc, 22(Pt 4), pp.1977~1986 (2020)
- 사민우, "A Study on Temperature Changes during Bone Scaling and Cutting of Dental Ultrasonic Scaling", Surgery System, 2020, pp.19(2), 1~8[DOI]